Datasheet
30
testo Saveris
The alkali manganese batteries AA (0515 0414) are included in these ordering data (Analog Coupler excluded). testo Saveris probes are delivered with a calibration protocol of the factory
adjustment data. Calibration certificates must be ordered separately.
Measuring range
Accuracy
-35 to +50 °C
±0.4 °C (-25 to +50 °C)
±0.8 °C (remaining range)
Measuring range
(Instrument)
Accuracy
(Instrument)
Resolution (Instrument)
Dimensions (housing):
Operating temperature
Probe type
Probe type
NTC
Weight
Storage temperature
Measuring rate
Protection class IP68
Conformity with standards
Radio frequency
NTC via mini-DIN socket,
door contact connection
cable included in delivery
(1.80 m)
Connection
Battery life
(Type: 4 AA batteries)
Display (optional)
Transmission distance
-35 to +50 °C
±0.4 °C (-25 to +50 °C)
±0.8 °C (remaining range)
NTC
0.1 °C
-50 to +150 °C
±0.2 °C (-25 to +70 °C)
±0.4 °C (remaining range)
NTC
2 TCs via TC socket, max.
difference in potential 2 V
-200 to +400 °C
TC type T
0.1 °C/TC type S 1 °C
-195 to +1350 °C
±0.5 °C or 0.5% of m.v.
TC type K
0 to +1760 °C
TC type S
-100 to +750 °C
TC type J
IP54
1 Pt100 via mini-DIN socket
0.01 °C
-200 to +600 °C
at +25 °C
±0.1 °C (0 to +60 °C)
±0.2 °C (-100 to +200 °C)
±0.5 °C (remaining range)
Pt100
IP68
Housing material
-35 to +50 °C
Approx. 240 g
Standard 15 min, 1 min to 24 h can be set
Memory
6,000 measurement values per channel
-40 to +55 °C (incl. batteries)
DIN EN 12830
Battery life at +25 °C, 3 years; for freezer applications, 3 years with L91 Photo lithium Energizer batteries
-20 to +50 °C
LCD, 2 lines; 7-segment with symbols
approx. 100 m w/o obstruction at frequency 2.4 GHz
Plastic
2,4 GHz
Wall bracket
included
* The testo Saveris Converter V 2.0 (order no. 0572 0258) is required for integration of testo Saveris wireless probes into systems with Base Firmware V 1.X . For more information please
contact our customer hotline or your Testo partner.
Radio probes
–
– –
––
– –
0.1 °CResolution 0.1 °C – –
–
–
–
–
–
80 x 85 x 38 mm
testo Saveris T1 testo Saveris T2 testo Saveris T3 testo Saveris Pt
Version without display
0572 1250*testo Saveris T1 Radio probe with internal NTC, radio frequency 2.4 GHz, without display
0572 1251*
testo Saveris T2 Radio probe with external probe connection and internal NTC, door contact, radio frequency 2.4 GHz, without display
0572 9252*testo Saveris T3 2-channel radio probe with 2 external TC probe connections (choice of TC characteristics), radio fre-
quency 2.4 GHz, without display
0572 7251*testo Saveris Pt Radio probe with 1 external Pt100 probe connection, radio frequency 2.4 GHz, without display
Version with display
0572 1260*testo Saveris T1 D Radio probe with internal NTC, radio frequency 2.4 GHz, with display
0572 1261*testo Saveris T2 D Radio probe with external probe connection and internal NTC, radio frequency 2.4 GHz, with display
0572 9262*testo Saveris T3 D 2-channel radio probe with 2 external TC probe connections (choice of TC characteristics), radio
frequency 2.4 GHz, with display
0572 7261*testo Saveris Pt D Radio probe with 1 external Pt100 probe connection, radio frequency 2.4 GHz, with display
Part no.
Internal sensorExternal probe
Technical data