Data Sheet
Table Of Contents
- 1 About this document
- 2 Safety and disposal
- 3 Product-specific approvals
- 4 Product-specific information
- 5 Use
- 6 Product description
- 7 First steps
- 8 Using the product
- 9 Smart App
- 10 Maintenance
- 11 Technical data
- 12 Tips and assistance
- 13 Support
11 Technical data
70
10.6 Changing batteries/rechargeable
batteries
The instrument is switched off.
1
Fold out the suspension hook,
release the clip and remove the
battery compartment lid.
2
Remove the spent (rechargeable) batteries and insert new ones (4 x
AA / Mignon / LR6) into the battery compartment. Observe the
polarity!
3
Attach and close the battery compartment lid (the clip must click into
place).
4
Switch the instrument on.
11 Technical data
Feature Value
Measurement parameters
Pressure: kPa/MPa/bar/psi
Temperature: °C/°F/K
Vacuum: hPa / mbar/ Torr / mTorr / inH
2
O /
micron / inHg / Pa
Sensor
testo 550s
testo 557s
Connections: 3
Valves: 3
Connections: 4
Valves: 4
Pressure: 2 x pressure sensor
Temperature: 2 x NTC
Vacuum: via external probe
Up to 4 Smart Probes via Bluetooth
®
connection
Measuring cycle
0.5 s
Interfaces
Pressure ports: 3 x 7/16" UNF, 1 x 5/8" UNF
NTC measurement
External vacuum probe