9 Performing the measurement 
56 
9.8  SiteRecognition 
With the PC software testo IRSoft QR Codes, markers can be set to clearly 
identify a measuring location. By setting a marker with the built-in digital imager, 
the images, which are then saved, are automatically allocated to the relevant 
site (allocation is saved with this image). For the creation of markers, 
transferring of measuring site data to the imager and copying of images to the 
PC software, see instruction manual for the PC software. 
1 
Open the Menu. 
2 
Select SiteRecognition (joystick or touchscreen). 
 
SiteRecognition opens. 
 
The real image is displayed and a position frame is shown. 
To quickly change the measuring locations, SiteRecognition can be 
assigned to the quick select button.  
Enter measuring site marker 
1 
Position the imager so that the marker is inside the position frame. 
2 
After recognition of the marker ID: confirm transfer of measuring site 
data. 
 
The thermographic image that is 
taken next is assigned to the 
measuring location. 
Several images can be saved for the same measuring location. 