6 Commissioning
18
For longer breaks, you should discharge and recharge the batteries every 3-
4 months. Trickle charging should not exceed 2 days.
6.4 Switching the instrument on and off
Switching on the camera
1
Remove protective cap from the lens.
2
Press .
The camera starts.
The start screen appears on the
display.
To ensure measuring accuracy, the camera performs automatic zeroing
approx. every 60 s. A “click” is heard when this happens. The image
freezes briefly when this occurs. Zeroing is carried out more frequently
during the camera warm-up period (takes approx. 90 seconds).
During the warm-up period, the display shows and no measuring
accuracy is guaranteed. The image can already be used for indication
purposes and saved.
Switching off the camera
1
Press and hold down until the
progress bar on the display is
complete.