User Manual
6 Technical data
17
6.1.2. General technical data
Characteristic
Values
Parameters
Pressure: kPa / MPa / bar / psi
Temperature: °C / °F / K
Sensors
2 Pressure: sensors, 2 Temperature (NTC
Thermistors)
Meas. cycle
0.5 s
Interfaces
Pressure connections: 3 x 1/4" SAE
2 Temperature NTC measurement channels
Measuring ranges
HP/LP pressure measuring range: -100 to
6000 kPa / -0.1 to 6 MPa / -1 to 60 bar (rel)
/ -14.7 to 870 psi
Temperature measuring range: -50 to
+150 °C / -58 to 302°F
Vacuum measuring range (rel): -1 to
0 bar / -14.7 to 0 psi
Overload
65 bar, 6500 kPa, 6.5 MPa, 940 psi
Resolution
Pressure resolution: 0.01 bar / 0.1 psi /
1 kPa / 0.001 MPa
Temperature resolution: 0.1 °C/0.1 °F
Accuracy (nominal
temperature
22 °C/71.6 °F)
Pressure: ± 0.5 % FS
Temperature 58 to 302°F (0.9°F ± 1 digit)
0 to 150°C (±0.5 °C ±1 digit)
No. of refrigerants
60
Selectable
refrigerants
No refrigerant, R11, R12, R22, R123,
R1234ze, R125, R13B1, R134a, R14,
R142B, R152a, R161, R23, R227, R290,
R32, R401A, R401B, R401C, R402A,
R402B, R404A, R406A, R407A, R407B,
R407C, R407D, R407F, R408A, R409A,
R410A, R411A, R412A, R413A, R414B,
R416A, R417A, R420A, R421A, R421B,
R422A, R422B, R422C, R422D, R424A,
R426A, R427A, R434A, R437A, R438A,
R502, R503, R507, R508A, R508B, R600,
R600a, R718 (H
2
O), R744 (CO
2
) (only in
measuring range up to 870 psi (60 bar),
R1234yf