User Manual

5 Using the manifold
10
WARNING
Dropping this instrument or any other comparable mechanical
shock can damage the refrigerant pipes and hoses. The valve
actuators may also suffer damage, which in turn could result in
further damage inside the instrument and may not be detectable
from outside.
> For your own safety you should return the manifold to the
Testo Service Department for technical inspection.
> You should therefore always replace the refrigerant hoses with
new ones after an accidental drop has occurred or after any
visible wear and tear.
Choosing the refrigerant
1. Press
[R, , ■].
- It activates the refrigerant menu and the currently selected
refrigerant flashes.
2. Setting the refrigerant:
Key functions
Description
[▲] or [▼]
Selecting another refrigerant
[R, ►, ■]
Confirm the selection and exit the refrigerant
menu
Available refrigerants
Representation
Description
R...
Refrigerant number of refrigerant acc. to
ISO 817
---
no refrigerant selected.
Example: Selecting refrigerant R401B
1. Press
[R, , ■] to activate refrigerant menu.
2. Press
[▲] or [▼] several times, until R401B flashes.
3. Press
[R, , ■] to confirm the selection.
Exiting the refrigerant selection
> Press [R, , ■] or automatically after 30 s, if no other key has
been pressed.