Information about surface measurement:
The specified response times t
99
are measured on polished steel or aluminium plates at +60 °C.
The specified accuracies are sensor accuracies.
Their accuracy in your application depends on the surface properties (roughness), the material of the measurement
object (thermal capacity and heat transfer) and the sensor accuracy. Testo will produce a corresponding calibration
certificate for the deviations of your measurement system in your application. For this, Testo uses a surface test bed
developed in cooperation with the PTB (Physikalisch Technische Bundesanstalt - National Metrology Institute of
Germany).