User manual
5. Operation 31
3
Select the desired option with / and confirm
with :
· On: The flow measurement value is displayed in the
lower measurement value line
· OFF: The flow measurement value is not displayed
in the lower measurement value line
OFF is selected:
º Next handling objective 4 SET DAMPING.
On is selected:
- The set unit of flow blinks.
4 Select the desired unit with / and confirm
with :
- The set air-density and the corresponding unit are
displayed.
Entering the air-density is required to calculate the flow
correctly.
5 Set the value with / and confirm with :
- The set Pitot factor is displayed, Pitot factor lights
up.
The Pitot factor is dependant on the Pitot tube used.
6 Set the value with / and confirm with :
4
Setting ddamping:
When damping is activated, a moving mean value is
shown in the display, which is created from an
adjustable number of measurement values:
· 1 = Damping deactivated, current measurement value
is shown
· 20 = Maximum damping, mean value calculation over
the last 20 measurement values.
The configuration modus is open, Damping lights up.
i Set the value with / and confirm with :