User Manual
Table Of Contents
9 Technical data
20
9.2.5. testo 805i
Feature Values
Measuring range -30 °C to 250 °C / -22 to 482 °F
Accuracy ± 1 digit ± 1.5 °C or ± 1.5% of m.v. (0 to 250 °C)
± 2.0 °C (-20.0 to -0.1 °C)
± 2.5 °C (-30.0 to -20.1 °C)
± 2.7 °F or ± 1.5% of m.v. (32 to 482 °F)
± 3.6 °F (-4 to 32 °F)
± 4.5 °F (-22 to -4 °F)
Resolution 0.1 °C / 0.1 °F
Measurement rate 2/sec
Available units of
measurement
°C, °F
Connection 7/16“ – UNF
Storage temperature -20 °C to 60 °C / -4 to 140 °F
Operating temperature -10 °C to +50 °C / 14 to 122 °F
Battery type 3 micro batteries AAA
Battery life 30 hrs
Optics 10:1
Laser marking Diffraction lens as laser marking (laser circle)
Dimensions 140 mm × 36 mm × 25 mm
Emission level Adjustable from 0.1 to 1.0
Directives, standards
and tests
EU guideline: 2014/30/EU
9.2.6. testo 605i
Feature Values
Measuring range -20 to 60 °C, -4 to 140 °F, 0 to 100% RH
Accuracy ± 1 digit ± 1.8% RH + 3% of m.v. at +25 °C (5 to 80% RH)
± 0.03% RH / K (0 to 60 °C)
± 0.8 °C (-20 to 0 °C) / ± 1.44 °F (-4 to 32 °F)
± 0.5 °C (0 to +60 °C) / ± 0.9 °F (32 to 140 °F)
Resolution 0.1 °F / 0.1 °C
0.1% RH