Datasheet
Attention!
Never dismantle the probe while it is combined with the voltage source and only connect it to a grounded oscilloscope.
correct
incorrect
incorrect
correct
incorrect
incorrect
Probe Compensation
Proper compensation of the probe is required to assure amplitude accuracy of the waveform being measured by
matching the probe to the oscilloscope’s input capacitance. Compensation should be adjusted whenever the probe is
connected to or transferred between oscilloscopes.
Low Frequency Adjustment
Apply a 1 kHz square wave to the probe or connect to the oscilloscope’s calibrator output.
Adjust the single LF trimmer located in the BNC Box until you achieve a flat-topped square wave (see figure below).
High Frequency Adjustment
Connect the probe to a 1 MHz square wave signal (rise time less than 0,7ns).
Remove the two plastic caps from the BNC compensation box.
Adjust left trimmer first then right trimmer until you achieve a flat-topped square wave (see figure below).
Derating Curve