MANUAL ADDENDUM MODEL T201 AMMONIA ANALYZER (For use with the T200 Operators Manual, P/N 06858) TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TELEDYNE-API) 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 TOLL-FREE: FAX: TEL: E-MAIL: WEB SITE: Copyright 2011-2013 Teledyne Advanced Pollution Instrumentation 800-324-5190 858-657-9816 858-657-9800 api-sales@teledyne.com www.teledyne-api.
ABOUT THIS MANUAL This addendum is to be used in conjunction with the T200 NOx Analyzer manual, p/n 06858; for operating the T201 analyzer; it includes document(s) as listed below. Part No. 07356 Note Name/Description T201 Spare Parts List (in Section 7.0) Note: Please refer to our Website or Sales for more recent updates. We recommend that this manual and the T200 manual be read in their entirety before any attempt is made to operate the instrument.
This page intentionally left blank.
Teledyne API Model T201 NH3 Analyzer Operator Manual T201 Ammonia Analyzer SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a safety alert symbol, and are placed throughout this manual; the safety symbols are also located inside the instrument.
CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement ces consignes. Chaque consigne de sécurité est représentée par un pictogramme d’alerte de sécurité; ces pictogrammes se retrouvent dans ce manuel et à l’intérieur des instruments.
TABLE OF CONTENTS MODEL T201 AMMONIA ANALYZER .............................................................I CONSIGNES DE SÉCURITÉ ......................................................................... IV 1.0 T201 AMMONIA ANALYZER.................................................................7 1.1 Theory of Operation...................................................................................................... 10 1.1.1 Minimizing PMT Drift.........................................................
Table 2-4. NO Calibration Procedure - NO Gas through the SPAN Port....................... 19 Table 2-5. Confirming Ammonia Converter Efficiency................................................... 21 Table 2-6. NH3 Calibration Procedure – NH3 Gas through the SAMPLE Port ............... 22 Table 2-7. NH3 Calibration Procedure – NH3 Gas through the SPAN Port.................... 23 Table 3-1. Analog Output Data Default Settings ........................................................... 25 Table 3-2.
1.0 T201 AMMONIA ANALYZER This manual is to be used in conjunction with the T200 NOx Analyzer manual, p/n 06858 supplied with this instrument. It is important that you familiarize yourself with the workings of the NOx analyzer before proceeding with NH3 measurements. The T201 ammonia analyzer consists of a modified T200 nitrogen oxide analyzer and an M501 NH3 ammonia converter. The analyzer measures the following individual gas concentrations: TNx (NH3 and NOx), NOx (NO and NO2) and NO in the sample gas.
T201 Ammonia Analyzer Teledyne API Model T201 NH3 Analyzer Operator Manual Figure 1-1.
Teledyne API Model T201 NH3 Analyzer Operator Manual T201 Ammonia Analyzer Figure 1-2.
T201 Ammonia Analyzer Teledyne API Model T201 NH3 Analyzer Operator Manual 1.1 THEORY OF OPERATION The Teledyne-API Model T201 analyzer measures ammonia by oxidizing it to nitric oxide by the following reaction: 4NH3 + 5O2 4NO + 6H2O The resulting nitric oxide is then measured by the chemiluminescent reaction with ozone. Consult Section 10.1 in the T200 manual (provided with this instrument) for more details on the nitric oxide measurement.
Teledyne API Model T201 NH3 Analyzer Operator Manual T201 Ammonia Analyzer The ammonia converter efficiency factor (NH3_CE_FACTOR1) is discussed in more detail in Section 2.3 of this manual. The conversion efficiency of the M501 NH3 should be checked prior to starting long term tests. Both efficiency factors can be accessed through the analyzer VARS MENU.
T201 Ammonia Analyzer Teledyne API Model T201 NH3 Analyzer Operator Manual As can be seen from the above reactions, the calculation of the ammonia concentration and overall calibration of the instrument must be done carefully, if accurate ammonia concentrations are to be measured. 1.2 SPECIAL CONSIDERATIONS FOR AMMONIA MEASUREMENT Ammonia is a difficult gas to measure due to its chemical characteristics. The gas tends to adsorb onto surfaces and diffuse into many materials.
Teledyne API Model T201 NH3 Analyzer Operator Manual T201 Ammonia Analyzer 1.4 T201 ANALYZER SPECIFICATIONS Table 1-1.
This page intentionally left blank.
2.0 CALIBRATION PROCEDURE This section begins with a high-level overview of the calibration procedure for the T201 analyzer. Details are provided starting in Section 2.1. First, take the time to read the T200 manual to familiarize yourself with the Chemiluminescence process. It is important to remember that the analyzer is merely measuring different levels of nitric oxide sample gas on three separate channels (TNx, NOx, NO).
Calibration Procedure Teledyne API Model T201 NH3 Analyzer Operator Manual 2.1 ZEROING THE ANALYZER The analyzer can be zeroed by either applying zero air straight to the sample port or through the optional Zero/Span valves, if installed. It is important to remember that if the analyzer was previously sampling ammonia gas, prior to conducting a Zero calibration, it will take some period of time before the zero baseline is reached.
Teledyne API Model T201 NH3 Analyzer Operator Manual Calibration Procedure Table 2-2. Zero Calibration Procedure - Zero Gas through ZERO Port Step No. Action 1. Press CALZ 2 Table 2.1 Notes Comment The T201 enters the calibrate mode from sample mode. The zero gas is supplied through the ZERO inlet port on the rear panel of the M501 NH3. Follow Steps 2 to 8 in Table 2-1. Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected concentration values.
Calibration Procedure Teledyne API Model T201 NH3 Analyzer Operator Manual Table 2-3. NO Calibration Procedure – NO Gas through the SAMPLE Port Important: Each channel (TNx and NOx) must be Spanned individually Step No. Action 1. Press CAL 2. Channel Selection 3. Range Selection Comment The T201 enters the calibrate mode from sample mode. Confirm that excess NO gas is flowing past the sample port. Press TNX or NOx then Press ENTR. Press LOW or HIGH range then Press ENTR.
Teledyne API Model T201 NH3 Analyzer Operator Manual Calibration Procedure Table 2-4. NO Calibration Procedure - NO Gas through the SPAN Port Step No. Action 1. Press CALS 2 Table 2.3 Notes Comment The T201 enters the calibrate mode from sample mode. The NO gas is supplied through the SPAN inlet port on the rear panel of the M501 NH3. Follow Steps 2 to 5 in Table 2-3. Both the TNx and NOx channels must be spanned individually with nitric oxide gas.
Calibration Procedure Teledyne API Model T201 NH3 Analyzer Operator Manual The most important criterion an individual can avail themselves of when spanning an analyzer with ammonia gas is “patience”. Ammonia is a very sticky gas and the response of the analyzer/calibration system depends on a number of factors.
Teledyne API Model T201 NH3 Analyzer Operator Manual Calibration Procedure Table 2-5. Confirming Ammonia Converter Efficiency Step No. 1. Action Press CAL 2. Channel Selection 3. Range Selection 4. Press CONC Comment The T201 enters the calibrate mode from Sample mode. Press TNX then Press ENTR. Press LOW range then Press ENTR. The following will be displayed: TNx CONV Exit Press CONV enter 1.
Calibration Procedure Teledyne API Model T201 NH3 Analyzer Operator Manual Table 2-6. NH3 Calibration Procedure – NH3 Gas through the SAMPLE Port Step No. Action Comment 1. Generate a known level of NH3 gas using a calibration gas source Monitor the TNx concentration until it has stabilized; this signifies that both the NOx and NH3 gases have stabilized.
Teledyne API Model T201 NH3 Analyzer Operator Manual Calibration Procedure Table 2-7. NH3 Calibration Procedure – NH3 Gas through the SPAN Port Step No. Action 1. Press CALS Comment The T201 enters the calibrate mode from Sample mode. The ammonia gas is supplied through the SPAN inlet port on the rear panel of the M501 NH3. Calculate the Efficiency factor as outlined in Steps 2 in table 2.5. 2 Note 07271B DCN6646 Table 2.5 Follow Steps 4 to 6 in table 2.5.
Calibration Procedure Teledyne API Model T201 NH3 Analyzer Operator Manual This page intentionally left blank.
3.0 CONFIGURABLE ANALOG OUTPUT OVERVIEW There are three different methods to extract concentration data from the analyzer: internal data acquisition system (DAS), hyperlink, or strip chart. DAS The user can set up an internal DAS configuration either through the user display or using TAPI’s windows based software called APICOM. The analyzer stores data internally, which is available later for download through the RS-232 or Ethernet port.
Configurable Analog Output Overview Teledyne API Model T201 NH3 Analyzer Operator Manual Table 3-2. Analog Output Pin-Outs PIN 1 2 3 4 5 6 7 8 ANALOG OUTPUT A1 A2 A3 A4 VOLTAGE SIGNAL CURRENT SIGNAL V Out I Out + Ground I Out - V Out I Out + Ground I Out - V Out I Out + Ground I Out - V Out N/A Ground N/A Additionally A1, A2 and A3 may be equipped with optional 0-20 mA, or 4-20, current loop drivers. The 4-20 mA option is not available on A4.
4.0 MAINTENANCE SCHEDULE The maintenance requirements of the T201 are the same as a standard T200 NOx analyzer but with additional maintenance items (Table 4-1). Please refer to Section 13 in the T200 Operator Manual for the T200 maintenance schedule and for maintenance and repair procedures. The T201 requires the following additional maintenance items. Table 4-1.
Maintenance Schedule Teledyne API Model T201 NH3 Analyzer Operator Manual Table 4-2. M501 NH3 Converter Rebuild Parts List Part No. KIT000193 KIT000139 HE0000007 Description RETROFIT, M501 NH3 INNER TUBE w/SCREEN REBUILD, CERAMIC BUSHINGS, M501 NH3 CERAMIC HEATER, 220W@60V M501XX 4.2 REPLACING THE CATALYTIC CARTRIDGE 1. Turn off power to the converter and allow to cool. IT IS VERY IMPORTANT THAT THE CONVERTER BE COOL BEFORE ATTEMPTING ANY DISASSEMBLY OR REPAIRS.
Teledyne API Model T201 NH3 Analyzer Operator Manual Maintenance Schedule 4.3 REPLACING THE THERMOCOUPLE The following instructions provide the necessary information to replace the existing thermocouple. Figure 4-2 is provided for reference.
Maintenance Schedule Teledyne API Model T201 NH3 Analyzer Operator Manual Once you have the right tools and parts, replace the thermocouple as follows: 1. Ensure power is removed from the M501 NH3 Converter. If the Converter has been operational you will need to wait for at least 2 hours for the Converter oven to cool before continuing with the replacement of the thermocouple. 2. Remove the cover from the Converter chassis. 3. Unscrew the (4) nuts that secure the front panel to the chassis.
Teledyne API Model T201 NH3 Analyzer Operator Manual Maintenance Schedule 18. After the Auto-Tune process is completed, verify that the “process” temperature is indicating that the desired temperature is stable and being regulated. The converter is now ready for operation. 4.4 AUTOZERO FLOW CHECK Since the T201 is an ammonia analyzer, the flow through the AutoZero orifice is especially important. Check the AutoZero flow as follows: 1. This procedure should be performed with the sample pump running. 2.
Maintenance Schedule Teledyne API Model T201 NH3 Analyzer Operator Manual This page intentionally left blank.
5.0 ALARMS AND CAUTIONS Please refer to the T200 Operation Manual for a functional block diagram of the analyzer. 5.1 ALARM During initial warm up the internal Zero/Span factory option may exhibit a “Block Temperature” warning. This can be cleared once the block temperature stabilizes at 50 degrees Celsius. The block temperature is the actual temperature of the IZS block manifold inside the M501 NH3 converter.
Alarms and Cautions Teledyne API Model T201 NH3 Analyzer Operator Manual This page intentionally left blank.
6.0 CALCULATING MOLY CONVERTER EFFICIENCIES 6.1 PURPOSE To provide instructions on how to calculate the efficiency of a Moly converter when using a GPT method of testing converters, by using the US EPA method, where the actual concentration of ozone is not a factor in the accuracy of the calculation of the converter efficiency. This procedure is based on the Code of Federal Regulations, Title 40, Chapter I, subchapter C, Part 50, Appendix F. 6.
Calculating Moly Converter Efficiencies Teledyne API Model T201 NH3 Analyzer Operator Manual 3. The first gas check is to test to see how much the converter is eating NO gas or out gassing NO gas. Bypass the converter in the machine, by placing a short piece of tubing in place of the converter. Perform a straight dilution with NO gas and air as a diluent gas. Input this 450 PPB NO gas into the analyzer, allow the machine to stabilize, and write down the NOx value on your data sheet on line 3. 4.
MOLY TEST DATA SHEET Line # TEST RESULT 2 LEAK-CHECK (WHEN HOT) YES / NO 3 NOX RESPONSE (MOLY BYPASSED) __________ 4 NOX RESPONSE (MOLY IN-LINE) __________ 5 OUT-GASSING / EATING (NO – NOX) __________ (>-5 PPB, <5 PPB) 6 (NOx ORIG) (NOX mode, O3 off) __________ PPB 7 (NOx REM) (NOX mode, O3 on) __________ PPB 8 NOX LOSS (9A - 10B) (<4% of NOx ORIG; example: for 450PPB 9 (NO ORIG) (NO mode, O3 off) __________ PPB 10 (NO REM) (NO mode, O3 on) __________ PPB 11 NO2 (9B - 10A)
Calculating Moly Converter Efficiencies Teledyne API Model T201 NH3 Analyzer Operator Manual This page intentionally left blank.
7.0 T201 SPARE PARTS LIST This section lists the spare parts for the T201; however, due to the dynamic nature of spare parts numbers, it is recommended that you refer to our Website or contact Sales for recent updates. (Reference: 07356 29 March 2012, 12:17PM) PARTNUMBER DESCRIPTION 000940100 CD, ORIFICE, .003 GREEN 000940400 CD, ORIFICE, .004 BLUE 000940500 CD, ORIFICE, .007 ORANGE 000940600 CD, ORIFICE, .
T201 Spare Parts List Teledyne API Model T201 NH3 Analyzer Operator Manual (Reference: 07356 29 March 2012, 12:17PM) PARTNUMBER DESCRIPTION 049310100 PCA,TEC DRIVER,PMT,(KB) 051211000 ASSY, O3 DESTRUCT W/FTGS (MNFLD - EXH) * 052930200 ASSY, BAND HEATER TYPE K, NOX 055710000 ASSY, FLT, INLINE SS, 3/8” CONN, 0.5 MIC 058021100 PCA, MOTHERBD, GEN 5-ICOP 062390000 ASSY, MOLY GUTS w/WOOL 066430000 PCA, OZONE PULSE DRIVER, 100HZ, T200 066970000 PCA, INTRF.
Teledyne API Model T201 NH3 Analyzer Operator Manual (Reference: 07356 29 March 2012, 12:17PM) PARTNUMBER DESCRIPTION HW0000685 LATCH, MAGNETIC, FRONT PANEL KIT000095 AKIT, REPLACEMENT COOLER KIT000219 AKIT, 4-20MA CURRENT OUTPUT KIT000253 ASSY & TEST, SPARE PS37 KIT000254 ASSY & TEST, SPARE PS38 OR0000001 ORING, 2-006VT *(KB) OR0000002 ORING, 2-023V OR0000025 ORING, 2-133V OR0000027 ORING, 2-042V OR0000034 ORING, 2-011V FT10 OR0000039 ORING, 2-012V OR0000044 ORING, 2-125V OR000008
T201 Spare Parts List Teledyne API Model T201 NH3 Analyzer Operator Manual This page intentionally left blank.