INSTRUCTION MANUAL MODEL GFC7000E CARBON DIOXIDE ANALYZER © TELEDYNE ANALYTICAL INSTRUMENTS 16830 Chestnut St. City of Industry, Ca. 91748 USA Phone: Phone: Fax: Fax: Copyright 2003 Teledyne Instruments Incorporated 626-961-9221 626-934-1500 626-961-2538 626-934-1651 04584 REV.
Model GFC7000E Instruction Manual GFC7000E Documentation SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument.
Model GFC7000E Instruction Manual GFC7000E Documentation TABLE OF CONTENTS SAFETY MESSAGES I TABLE OF CONTENTS ...................................................................................................................... II LIST OF APPENDICES.................................................................................................................... VII LIST OF FIGURES VII LIST OF TABLES VIII 1. MGFC7000E DOCUMENTATION........................................................................
Model GFC7000E Instruction Manual GFC7000E Documentation 6.7. SETUP – RNGE: Analog Output Reporting Range Configuration ..................................................... 56 6.7.1. Physical Range versus Analog Output Reporting Ranges........................................................ 56 6.7.2. Reporting Range Modes ...................................................................................................57 6.7.3. Single Range mode (SNGL) ...................................................
Model GFC7000E Instruction Manual GFC7000E Documentation 6.13.1.2. Control Inputs ...................................................................................................... 116 6.13.2. Remote Operation Using the External Serial I/O ............................................................... 117 6.13.2.1. Terminal Operating Modes...................................................................................... 117 6.13.2.2. Help Commands in Terminal Mode ................................
Model GFC7000E Instruction Manual GFC7000E Documentation 10.3. Electronic Operation ........................................................................................................... 161 10.3.1. Overview ................................................................................................................... 161 10.3.2. CPU........................................................................................................................... 163 10.3.3. Optical Bench & GFC Wheel.....
Model GFC7000E Instruction Manual GFC7000E Documentation 11.2.2. Non-Repeatable Zero and Span ..................................................................................... 195 11.2.3. Inability to Span – No SPAN Key.................................................................................... 196 11.2.4. Inability to Zero – No ZERO Key .................................................................................... 196 11.3. Other Performance Problems ....................................
Model GFC7000E Instruction Manual GFC7000E Documentation LIST OF APPENDICES APPENDIX A - VERSION SPECIFIC SOFTWARE DOCUMENTATION APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0 APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0 APPENDIX A-3: Warnings and Test Functions, Revision E.0 APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0 APPENDIX A-5: MGFC7000E iDAS Functions, Revision E.0 APPENDIX A-6: Terminal Command Designators, Revision E.
Model GFC7000E Instruction Manual GFC7000E Documentation Figure 7-6: Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Pneumatic Connections–MGFC7000E with Zero/Span Valves with External Zero air Scrubber (OPT 53)................................................................................
Model GFC7000E Instruction Manual Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table 6-2: 6-3: 6-4: 6-5: 6-6: 6-7: 6-8: 6-9: 6-10: 6-11: 6-12: 6-13: 6-14: 6-15: 6-16: 6-17: 6-18: 6-19: 6-20: 6-21: 6-22: 6-23: 6-24: 6-25: 6-26: 6-27: 6-28:
Model GFC7000E Instruction Manual Table Table Table Table Table Table Table A-2: A-3: A-4: A-5: A-6: A-7: D-1: GFC7000E Documentation GFC7000E Warning Messages, Revision E.0 .................................................... GFC7000E Test Functions, Revision E.0 .......................................................... GFC7000E Signal I/O Definitions, Revision E.0 ................................................ GFC7000E DAS Trigger Events, Revision E.0 ................................................
Model GFC7000E Instruction Manual MGFC7000E Documentation 1. MGFC7000E DOCUMENTATION Thank you for purchasing the Model GFC7000E Gas Filter Correlation Carbon Dioxide Analyzer! The documentation for this instrument is available in several different formats: • Printed format, part number 045840100 • Electronic format on a CD-ROM, part number 045840200 The electronic manual is in Adobe® Systems Inc. “Portable Document Format”.
Model GFC7000E Instruction Manual MGFC7000E Documentation 2.0 Specifications and Warranty This section contains a list of the analyzer’s performance specifications, a description of the conditions and configuration under which EPA equivalency was approved and Teledyne Instruments Incorporated’s warranty statement. 3.0 Getting Started: A concise set of instructions for setting up, installing and running your analyzer for the first time. 4.
Model GFC7000E Instruction Manual MGFC7000E Documentation Appendices: For easier access and better updating, some information has been separated out of the manual and placed in a series of appendices at the end of this manual. These include: software menu trees, warning messages, definitions of iDAS & serial I/O variables, spare parts list, repair questionnaire, interconnect listing and drawings, and electronic schematics.
Model GFC7000E Instruction Manual Specifications, Approvals and Warranty 2. SPECIFICATIONS, APPROVALS AND WARRANTY 2.1. Specifications Table 2-1: Model GFC7000E Basic Unit Specifications Min/Max Range (Physical Analog Output) In 1ppb increments from 50ppb to 2 000ppm, dual ranges or auto ranging Measurement Units ppb, ppm, µg/m3, mg/m3, %(user selectable) Zero Noise < 0.1 ppm (RMS) Span Noise < 1% of reading (RMS) Lower Detectable Limit1 < 0.2 ppm1 Zero Drift (24 hours) <0.
Model GFC7000E Instruction Manual Specifications, Approvals and Warranty 2.2. CE Mark Compliance Emissions Compliance The Teledyne Instruments Model GFC7000E Gas Filter Correlation CO2 Analyzer was tested and found to be fully compliant with: EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A. Tested on 11-29-2001 at CKC Laboratories, Inc., Report Number CE01-249.
Model GFC7000E Instruction Manual Specifications, Approvals and Warranty The warranties set forth in this section shall be of no force or effect with respect to any product: (i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used in any manner other than in accordance with the instruction provided by TELEDYNE ANALYTICAL INSTRUMENTS or (iii) not properly maintained.
Model GFC7000E Instruction Manual Getting Started 3. GETTING STARTED 3.1. Unpacking and Initial Set Up CAUTION To avoid personal injury, always use two persons to lift and carry the Model GFC7000E. 1. Verify that there is no apparent external shipping damage. If damage has occurred, please advise the shipper first, then Teledyne Analytical Instruments. 2. Included with your analyzer is a printed record of the final performance characterization performed on your instrument at the factory.
Model GFC7000E Instruction Manual Getting Started 5. Check the connectors of the various internal wiring harnesses and pneumatic hoses to make sure they are firmly and properly seated. 6. Verify that all of the optional hardware ordered with the unit has been installed. These are listed on the paperwork accompanying the analyzer. 7.
Model GFC7000E Instruction Manual Getting Started 8. VENTILATION CLEARANCE: Whether the analyzer is set up on a bench or installed into an instrument rack, be sure to leave sufficient ventilation clearance. • AREA MINIMUM REQUIRED CLEARANCE Back of the instrument 4 in. Sides of the instrument 1 in. Above and below the instrument 1 in. Various rack mount kits are available for this analyzer. See Section 5.1 of this manual for more information. 3.1.1.
Model GFC7000E Instruction Manual Getting Started Figure 3-2: Rear Panel Layout 9. Attach a strip chart recorder and/or data-logger to the appropriate analog output connections on the rear panel of the analyzer. ANALOG 1 A1 2 A2 3 4 5 A3 6 A4 7 8 When the instrument is in its default configuration, the A1 and A2 channels output a signal that is proportional to the CO2 concentration of the sample gas.
Model GFC7000E Instruction Manual Table 3-1: Pin Getting Started GFC7000E Analog Output Pin Outs Analog Output VDC Signal A1 1 V Out 2 A2 3 4 A3 5 6 A4 (Spare) 7 mADC Signal 8 I Out + Ground I Out - V Out I Out + Ground I Out - V Out I Out + Ground I Out - V Out Not Available Ground Not Available • The default analog output voltage setting of the GFC7000E CO2 Analyzer is 0 – 5 VDC with a range of 0 – 500 ppm. • TO change these settings, see Sections 6.9.4 and 6.
Model GFC7000E Instruction Manual Getting Started The pin assignments for the status outputs can be found in the table below: Table 3-2: GFC7000E Status Output Pin Outs Output # Status Definition 1 SYSTEM OK On if no faults are present. CONC VALID On if CO2 concentration measurement is valid. If the CO2 concentration measurement is invalid, this bit is OFF. 3 HIGH RANGE On if unit is in high range of DUAL or AUTO range modes.
Model GFC7000E Instruction Manual Getting Started The pin assignments for the digital control inputs can be found in the table below: Table 3-3: GFC7000E Control Input Pin Outs Input # Status Definition A REMOTE ZERO CAL The Analyzer is placed in Zero Calibration mode. The mode field of the display will read ZERO CAL R. B REMOTE SPAN CAL The Analyzer is placed in Span Calibration mode. The mode field of the display will read SPAN CAL R.
Model GFC7000E Instruction Manual Getting Started CAUTION In order to prevent dust from getting into the gas flow channels of your analyzer, it was shipped with small plugs inserted into each of the pneumatic fittings on the back panel. Make sure that all of these dust plugs are removed before attaching exhaust and supply gas lines. Table 3-4: Model GFC7000E Rear Panel Pneumatic Connections Rear Panel Label SAMPLE EXHAUST PRESSURE SPAN Function Connect a gas line from the source of sample gas here.
Model GFC7000E Instruction Manual Calibrated CO2 Gas Indicating soda-lime Getting Started Source of SAMPLE Gas Removed during Calibration MODEL 700 Gas Dilution Calibrator VENT Sample Exhaust Vent Span MODEL GFC7000E Pressure Span MODEL 701 Zero Air Generator IZS Purge In Figure 3-4: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator 1. Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more than 2 meters long.
Model GFC7000E Instruction Manual Getting Started 2. Attach sources of zero air and span gas (see Figures 3-3 through 3-8 inclusive). • Span Gas is a gas specifically mixed to match the chemical composition of the type of gas being measured at near full scale of the desired measurement range.
Model GFC7000E Instruction Manual Getting Started 3.1.2.2. Connections with Internal Valve Options Installed Figures 3-5 through 3-8 show the proper pneumatic connections for MGFC7000E’s with various optional internal valve sets installed.
Model GFC7000E Instruction Manual Getting Started Source of SAMPLE Gas VENT if input is pressurized MODEL 700 Gas Dilution Calibrator Certified CO2 Gas Sample Needle valve to control flow Exhaust VENT Vent Span MODEL GFC7000E Pressure Span Indicating soda-lime VENT IZS Purge In MODEL 701 Zero Air Generator Figure 3-7: Pneumatic Connections–MGFC7000E with Zero/Span Valves (OPT 52) Source of SAMPLE Gas VENT if input is pressurized Certified CO Gas MODEL 700 Gas Dilution Calibrator Sample I
Model GFC7000E Instruction Manual Getting Started Some applications may require multipoint calibration checks where span gas of several different concentrations is needed. We recommend using high-concentration, certified, calibration gas supplied to the analyzer through a Gas Dilution Calibrator such as a Teledyne Instruments Model 700. This type of calibrator precisely mixes Span Gas and Zero Air to produce any concentration level between 0 ppm and the concentration of the calibrated gas.
Model GFC7000E Instruction Manual Getting Started 3.2.1. Startup After electrical and pneumatic connections are made, turn on the instrument and pump power. The exhaust and PMT cooler fans should start. The display should immediately display a single, horizontal dash in the upper left corner of the display. This will last approximately 30 seconds while the CPU loads the operating system. Once the CPU has completed this activity it will begin loading the analyzer firmware and configuration data.
Model GFC7000E Instruction Manual Getting Started The analyzer should automatically switch to SAMPLE mode after completing the boot-up sequence and start monitoring CO2 gas. Warm-Up 3.2.2. Warm Up The MGFC7000E requires about 30 minutes warm-up time before reliable CO2 measurements can be taken. During that time, various portions of the instrument’s front panel will behave as follows. See Figure 3-10 for locations.
Model GFC7000E Instruction Manual Getting Started 3.2.3. Warning Messages Because internal temperatures and other conditions may be outside be specified limits during the analyzer’s warm-up period, the software will suppress most warning conditions for 30 minutes after power up. If warning messages persist after the 30 minutes warm up period is over, investigate their cause using the troubleshooting guidelines in Chapter 11 of this manual.
Model GFC7000E Instruction Manual Getting Started To View and Clear the various warning messages press: SAMPLE TEST deactivates warning messages TEST HVPS WARNING CAL MSG RANGE=500.000 PPM SAMPLE MSG < TST TST > CAL SAMPLE TEST NOTE: If the warning message persists after several attempts to clear it, the message may indicate a real problem and not an artifact of the warm-up period CO2 = 0.00 HVPS WARNING CAL MSG CLR SETUP CO2 = 0.00 CLR SETUP CO2 = 0.
Model GFC7000E Instruction Manual Getting Started 3. If your analyzer has a Ethernet card (Option 63) installed and your network is running a dynamic host configuration protocol (DHCP) software package, the Ethernet option will automatically configure its interface with your LAN. However, it is a good idea to check these settings to make sure that the DHCP has successfully downloaded the appropriate network settings from your network server (See Section 6.10.9.2).
Model GFC7000E Instruction Manual SAMPLE Getting Started RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE 8 SETUP ENTER SETUP PASS : 818 1 8 ENTR EXIT SETUP X.X CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT RANGE CONTROL MENU MODE SET UNIT EXIT Press this button to select the concentration units of measure: Press this button to set the analyzer for SNGL DUAL or AUTO ranges PPB, PPM, UGM, MGM SETUP X.
Model GFC7000E Instruction Manual Getting Started 2. Set the expected CO2 span gas concentration SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL The CO2 span concentration values automatically default to 400.0 Conc. To change this value to the actual concentration of the span gas, enter the number by pressing the key under each digit until the expected value appears. SETUP M-P CAL RANGE = 500.000 PPM < TST TST > ZERO CO2 =X.XXX EXIT CONC M-P CAL CO2 SPAN CONC: 400.
Model GFC7000E Instruction Manual Getting Started 3. Perform the Zero/Span Calibration Procedure SAMPLE* RANGE = 500.000 PPM CO2 =XXX.X < TST TST > CAL SAMPLE SETUP STABIL=XXX.X PPM Set the Display to show the STABIL test function. This function calculates the stability of the COx measurement CO2 =XXX.X < TST TST > CAL SETUP ACTION: Allow zero gas to enter the sample port at the rear of the instrument. Wait until STABIL falls below 1.0 ppm. M-P CAL STABIL=XXX.X PPM CO2 =XXX.
Model GFC7000E Instruction Manual Getting Started NOTE Once you have completed the above set-up procedures, please fill out the Quality Questionnaire that was shipped with your unit and return it to Teledyne Instruments. This information is vital to our efforts in continuously improving our service and our products. THANK YOU.
Model GFC7000E Instruction Manual Getting Started Sample Gas Outlet fitting Sample Gas Flow Sensor Sample Chamber Sync/Demod PCA Housing Pressure Sensor(s) Bench Temperature Thermistor Shock Absorbing Mounting Bracket Opto-Pickup PCA Purge Gas Pressure Regulator IR Source GFC Wheel Heat Sync GFC Wheel Motor GFC Temperature Sensor Purge Gas Inlet GFC Heater Figure 3-12: Optical Bench Layout SAMPLE GAS INLET PUMP EXHAUST GAS OUTLET Purge Gas Pressure Control Assy Sample Gas Critical Flow Orifice
Model GFC7000E Instruction Manual Getting Started User Notes 04584 Rev A1 32
Model GFC7000E Instruction Manual Frequently Asked Questions 4. FREQUENTLY ASKED QUESTIONS 4.1. FAQ’s The following is a list from the Teledyne Instruments’ Customer Service Department of the most commonly asked questions relating to the Model CO2 Analyzer. Q: How do I get the instrument to zero / Why is the zero key not displayed? A: See Section 11.2.4 Inability to zero. Q: How do I get the instrument to span / Why is the span key not displayed? A: See Section 11.2.3 Inability to span.
Model GFC7000E Instruction Manual Frequently Asked Questions Q: How long does the IR source last? A: Typical lifetime is about 2-3 years. Q: Where is the sintered filter/sample flow control orifice? A: These components are located inside the flow control assembly that is attached to the inlet side of the sample pump, see Figure 3-13. See Section 11.5.1 for instructions on disassembly and replacement.
Model GFC7000E Instruction Manual Frequently Asked Questions iDAS - acronym for internal data acquisition system IP – acronym for internet protocol LAN - acronym for local area network LED - acronym for light emitting diode PCA - acronym for printed circuit assembly, the Æ PCB with electronic components, ready to use.
Model GFC7000E Instruction Manual Optional Hardware and Software 5. OPTIONAL HARDWARE AND SOFTWARE This section includes a brief description of the hardware and software options available for the Model GFC7000E Gas Filter Correlation Carbon Dioxide Analyzer. For assistance with ordering these options please contact the Sales department of Teledyne – Advanced Pollution Instruments at: TEL: TEL: FAX: WEB SITE: 626-961-9221 626-934-1500 626-961-2538 www.teledyne-ai.com 5.1.
Model GFC7000E Instruction Manual Optional Hardware and Software 5.3. Expendable Kits (Options 42C, 42D and 43) OPTION NUMBER DESCRIPTION OPT 42C 1 year’s supply of replacement of 47mm dia. particulate filters OPT 42D 1 full replacement’s volume of indicating soda-lime for the external CO2 scrubber included with options 51 & 53 ( approximate active lifetime: 1 year) OPT 43 Options 42 C & 42D 5.4. Calibration Valves Options There are four available options involving Zero/Span/Shutoff valves.
Model GFC7000E Instruction Manual Optional Hardware and Software The state of the zero/span valves can also be controlled: • Manually from the analyzer’s front panel by using the SIGNAL I/O controls located under the DIAG Menu (Section 6.9.2), • By activating the instrument’s AutoCal feature (Section 7.6), • Remotely by using the external digital control inputs (Section 6.13.1.2 and Section 7.5.
Model GFC7000E Instruction Manual Optional Hardware and Software 5.4.2. Zero/Span/Shutoff with External CO2 Scrubber (Option 51) Option 51 is operationally and pneumatically identical to Option 50 above (see section 5.4.1), except that the zero air is generated by an externally mounted zero air scrubber filled with indicating soda-lime that changes color from white to pink as it becomes saturated. 5.4.3.
Model GFC7000E Instruction Manual Optional Hardware and Software Sample Gas Critical Flow Orifice INSTRUMENT CHASSIS SAMPLE GAS INLET PUMP EXHAUST GAS OUTLET GFC Wheel Motor Purge Gas Flow Control Orifice GFC Motor Heat Sync Purge Gas Pressure Control Assy GFC Wheel Housing PURGE GAS INLET FLOW / PRESSURE SENSOR PCA External CO2 Scrubber ( Option 53 Only) VENT SPAN OUTLET SAMPLE PRESSURE SENSOR SAMPLE CHAMBER FLOW SENSOR PRESSURE SPAN INLET ZERO/SPAN VALVE SAMPLE /CAL VALVE 2 3 3 1 1 IZ
Model GFC7000E Instruction Manual Optional Hardware and Software 5.5.2. RS-232 Multidrop (Option 62) The multidrop option is used with any of the RS-232 serial ports to enable communications of several analyzers with the host computer over a chain of RS-232 cables.
Model GFC7000E Instruction Manual Optional Hardware and Software 5.6.2. Manual on CD (Part number 045840200) This operators manual is also available on CD. The electronic document is stored in Adobe Systems Inc. Portable Document Format (PDF) and is viewable with Adobe Acrobat Reader® software, downloadable for free at http://www.adobe.com/ The CD version of the manual has many advantages: • Fully searchable text.
Model GFC7000E Instruction Manual Optional Hardware and Software When in maintenance mode the instrument ignores all commands received via the COMM ports that alter the operation state of the instrument This includes all calibration commands, diagnostic menu commands and the reset instrument command. the instrument continues to measure concentration and send data when requested. This option is of particular use for instruments connected to multidrop or Hessen protocol networks. 5.10.
Model GFC7000E Instruction Manual Operating Instructions 6. OPERATING INSTRUCTIONS To assist in navigating the analyzer’s software, a series of menu trees can be found in Appendix A-1 of this manual. NOTES The flow charts appearing in this section contain typical representations of the analyzer’s display during the various operations being described. These representations may differ slightly from the actual display of your instrument.
Model GFC7000E Instruction Manual Table 6-1: Operating Instructions Analyzer Operating modes MODE MEANING DIAG One of the analyzer’s diagnostic modes is being utilized (see Section 6.9). M-P CAL This is the basic, multi-point calibration mode of the instrument and is activated by pressing the CAL key. SAMPLE Sampling normally, flashing indicates adaptive filter is on. SAMPLE A 1 SETUP 1 Indicates that unit is in SAMPLE Mode and AUTOCAL feature is activated.
Model GFC7000E Instruction Manual Operating Instructions Title PPB, PPM, UGM, MGM The full scale limit at which the reporting range of the analyzer is currently set. THIS IS NOT the Physical Range of the instrument. See Section 6.7 for more information. STABIL PPB, PPM UGM, MGM Standard deviation of CO2 concentration readings. Data points are recorded every ten second. The calculation uses the last 25 data points.
Model GFC7000E Instruction Manual Operating Instructions To view the TEST Functions press the following Key sequence: SAMPLE RANGE = 500.000 PPM < TST TST > CAL SETUP RANGE RANGE11 RANGE21 STABIL CO2 MEAS CO2 REF MR RATIO PRES Refer to SAMP FL Table 6-2 for SAMP TEMP BENCH TEMP definitions WHEEL TEMP of these test BOX TEMP functions.
Model GFC7000E Instruction Manual Table 6-3: MESSAGE Operating Instructions List of Warning Messages MEANING ANALOG CAL WARNING The instruments A/D circuitry or one of its analog outputs is not calibrated. BENCH TEMP WARNING The Temperature of the optical bench is outside the specified limits.
Model GFC7000E Instruction Manual Operating Instructions To view and clear warning messages SAMPLE TEST deactivates warning messages TEST HVPS WARNING CAL MSG RANGE=500.000 PPM SAMPLE MSG < TST TST > CAL HVPS WARNING SAMPLE NOTE: If the warning message persists after several attempts to clear it, the message may indicate a real problem and not an artifact of the warm-up period Figure 6-3 TEST CAL MSG CO2 = 0.00 CLR SETUP CO2 = 0.00 CLR SETUP CO2 = 0.
Model GFC7000E Instruction Manual Operating Instructions The default status of the calibration password is OFF. To enable the calibration password press: SAMPLE RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE 8 SETUP ENTR accepts displayed password value ENTER SETUP PASS : 818 1 SETUP X.X 8 ENTR EXIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X CAL. PASSWORD default state is OFF EXIT CAL. PASSWORD ENABLE: OFF OFF SETUP X.
Model GFC7000E Instruction Manual Operating Instructions 6.4. SETUP Mode The SETUP mode contains a variety of choices that are used to configure the analyzer’s hardware and software features, perform diagnostic procedures, gather information on the instruments performance and configure or access data from the internal data acquisition system (iDAS). For a visual representation of the software menu trees, refer to Appendix A-1.
Model GFC7000E Instruction Manual Operating Instructions If the EXIT key is pressed before the ENTR key, the analyzer will beep alerting the user that the newly entered value has been lost. 6.4.1. SETUP Mode Password Security Whenever the Model GFC7000E’s SETUP mode is activated the instrument will prompt the user to enter a security password. The default password is 818.
Model GFC7000E Instruction Manual Operating Instructions 6.6. SETUP – CLK: Setting the Internal Time-of-Day Clock The MGFC7000E has a time of day clock that supports the AutoCal timer, time of day TEST function, and time stamps on most COM port messages. To set the time-of-day, press: SAMPLE RANGE = 500.000 PPM CO2 =X.XXX SETUP < TST TST > CAL ENTER SETUP PASS : 818 SAMPLE 8 1 SETUP X.X 8 ENTR EXIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.
Model GFC7000E Instruction Manual Operating Instructions In order to compensate for CPU clocks which run faster or slower, you can adjust a variable called CLOCK_ADJ to speed up or slow down the clock by a fixed amount every day. To change this variable, press: SAMPLE RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE SETUP SETUPX.X 1 ) CONC_PRECISION = 3 PREV NEXT JUMP EDIT PRNT EXIT ENTER SETUP PASS : 818 Continue to press NEXT until … 8 1 SETUP X.X 8 ENTR EXIT SETUP X.
Model GFC7000E Instruction Manual Operating Instructions 6.7. SETUP – RNGE: Analog Output Reporting Range Configuration The analyzer has three active analog output signals, accessible through a connector on the rear panel.
Model GFC7000E Instruction Manual Operating Instructions than 500 ppm, the full scale of expected values is only 25% of the instrument’s 2 000 ppm physical range. Unmodified, the corresponding output signal would also be recorded across only 25% of the range of the recording device. The MGFC7000E solves this problem by allowing the user to select a scaled reporting range for the analog outputs that only includes that portion of the physical range relevant to the specific application.
Model GFC7000E Instruction Manual SAMPLE* Operating Instructions RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE SETUP ENTER SETUP PASS : 818 8 1 ENTR EXIT 8 SETUP X.X CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT RANGE CONTROL MENU MODE SET UNIT SETUP X.X EXIT RANGE MODE: SNGL SNGL DUAL AUTO Only one of the range modes may be active at any time. Go To Section 6.7.3 ENTR EXIT Go To Section 6.7.4 EXIT Returns to the Main SAMPLE Display Go To Section 6.7.
Model GFC7000E Instruction Manual SAMPLE* RANGE = 500.000 PPM CO2 X.XXX < TST TST > CAL 8 SETUP 1 SETUP C.3 8 ENTR EXIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP C.3 SETUP C.3 EXIT RANGE MODE: SNGL 04584 Rev A1 MODE SET UNIT 0 0 EXIT RANGE: 500.0 Conc 5 SETUP C.3 ENTR EXIT ENTR EXIT RANGE CONTROL MENU SETUP C.3 EXIT SNGL DUAL AUTO SNGL DUAL AUTO SETUP C.3 RANGE CONTROL MENU MODE SET UNIT RANGE MODE: SNGL SETUP C.
Model GFC7000E Instruction Manual Operating Instructions 6.7.4. Dual Range Mode (DUAL) Selecting Dual Range mode allows the A1 and A2 outputs to be configured with different reporting ranges. The analyzer software calls these two ranges low and high. The low range setting corresponds with the analog output labeled A1 on the Rear Panel of the instrument. The high Range Setting corresponds with the A2 output. While the software names these two ranges low and high, they do not have to be configured that way.
Model GFC7000E Instruction Manual Operating Instructions 6.7.5. Auto Range Mode (AUTO) In AUTO range mode, the analyzer automatically switches the reporting range between two userdefined ranges (low and high). The unit will switch from low range to high range when the CO2 concentration exceeds 98% of the low range span. The unit will return from high range back to low range once both the CO2 concentration falls below 75% of the low range span.
Model GFC7000E Instruction Manual Operating Instructions 6.7.6. Range Units The MGFC7000E can display concentrations in parts per billion (109 mols per mol, PPB), parts per million (106 mols per mol, PPM), micrograms per cubic meter (µg/m3, UG), milligrams per cubic meter (mg/m3, MG) or percent (volume CO2/volume sample gas, %). Changing units affects all of the display, analog outputs, COM port and iDAS values for all reporting ranges regardless of the analyzer’s range mode.
Model GFC7000E Instruction Manual Operating Instructions 6.7.7. Dilution Ratio The dilution ratio is a software option that allows the user to compensate for any dilution of the sample gas before it enters the sample inlet. Using the dilution ratio option is a 4-step process: 1. Select reporting range units: Follow the procedure in Section 6.7.6. 2. Select the range: Use the procedures in Section 6.7.2 – 6.7.5.
Model GFC7000E Instruction Manual Operating Instructions 6.8. SETUP – VARS: Using the Internal Variables The MGFC7000E has several-user adjustable software variables, which define certain operational parameters. Usually, these variables are automatically set by the instrument’s firmware, but can be manually re-defined using the VARS menu. Table 6-6 lists all variables that are available within the 818 password protected level. Table 6-6: No. 0 1 2 Variable Names (VARS) Revision B.
Model GFC7000E Instruction Manual Operating Instructions To access and navigate the VARS menu, use the following key sequence. SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE SETUP ENTER SETUP PASS : 818 8 1 SETUP X.X 8 ENTR EXIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT EXIT ignores the new setting. SECONDARY SETUP MENU ENTR accepts the new setting. COMM VARS DIAG ALRM SETUP X.X EXIT 0 ) DAS_HOLD_OFF=15.0 Minutes SETUP X.X NEXT JUMP DAS_HOLD_OFF=15.
Model GFC7000E Instruction Manual Operating Instructions 6.9. SETUP – DIAG: Using the Diagnostics Functions A series of diagnostic tools is grouped together under the SETUPÆMOREÆDIAG menu. As these parameters are dependent on firmware revision (see Menu Tree A-6 in Appendix A). The individual parameters, however, are explained in more detail in the section indicated in Table 6-7.
Model GFC7000E Instruction Manual Operating Instructions 6.9.1. Accessing the Diagnostic Features To access the DIAG functions press the following keys: SAMPLE* RANGE = 500.00 PPM CO2 =X.
Model GFC7000E Instruction Manual Operating Instructions To enter the signal I/O test mode, press: SAMPLE RANGE = 500.000 PPM SETUP < TST TST > CAL 8 1 8 PREV NEXT JUMP DIAG I / O ENTER SETUP PASS : 818 SAMPLE SIGNAL I / O DIAG CO2 =X.XXX ENTR EXIT EXIT returns to the main SAMPLE display ENTR EXIT Test Signals Displayed Here PREV NEXT JUMP PRNT EXIT EXAMPLE SETUP X.X CFG DAS RNGE PASS CLK MORE SETUP X.
Model GFC7000E Instruction Manual Operating Instructions 6.9.4. Analog I/O Configuration The analog I/O functions that are available in the MGFC7000E are: Table 6-8: DIAG - Analog I/O Functions Sub Menu Function AOUTS CALIBRATED: Shows the status of the analog output calibration (YES/NO) and initiates a calibration of all analog output channels. CONC_OUT_1 Sets the basic electronic configuration of the A1 analog output (CO2 ) .
Model GFC7000E Instruction Manual Operating Instructions The following DC current output limits apply to the current loop modules: Table 6-10: Analog Output Current Loop Range RANGE MINIMUM OUTPUT MAXIMUM OUTPUT 0-20 mA 0 mA 20 mA These are the physical limits of the current loop modules, typical applications use 2-20 or 4-20 mA for the lower and upper limits. Please specify desired range when ordering this option. The default offset for all ranges is 0 mA.
Model GFC7000E Instruction Manual Operating Instructions 6.9.4.1. Analog Output Signal Type and Range Span Selection To select an output signal type (DC Voltage or current) and level for one output channel, activate the ANALOG I/O CONFIGURATION MENU (see Section 6.9.1) then press: FROM ANALOG I/O CONFIGURATION MENU DIAG PREV ANALOG I / O CONFIGURATION NEXT DIAG AIO ENTR EXIT AOUTS CALIBRATED: NO < SET SET> DIAG AIO CAL EXIT Press SET> to select the analog output channel to be configured.
Model GFC7000E Instruction Manual Operating Instructions To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU (see Section 6.9.1), then press: STARTING FROM DIAGNOSTIC MENU (see Section 6.7.
Model GFC7000E Instruction Manual Operating Instructions To select manual output calibration for a particular channel, activate the ANALOG I/O CONFIGURATION MENU (see Section 6.9.1), then press: DIAG Exit to return to the main sample display PREV ANALOG I / O CONFIGURATION NEXT ENTR EXIT DIAG AIO < SET SET> DIAG AIO AOUTS CALIBRATED: NO CAL Press SET> to select the analog output channel to be configured.
Model GFC7000E Instruction Manual Operating Instructions See Table 3-1 for pin assignments of Analog Out connector on the rear panel V +DC Gnd V OUT + V IN + V OUT - V IN - Recording Device ANALYZER Figure 6-5: Setup for Calibrating Analog Voltage Outputs To make these adjustments, the AOUT auto-calibration feature must be turned off (Section 6.9.4.2). Activate the ANALOG I/O CONFIGURATION MENU (see Section 6.9.
Model GFC7000E Instruction Manual Operating Instructions 6.9.4.4. Current Loop Output Adjustment A current loop option is available and can be installed as a retrofit for each of the analog outputs of the analyzer (Sections 5.2). This option converts the DC voltage analog output to a current signal with 0-20 mA output current. The outputs can be scaled to any set of limits within that 0-20 mA range. However, most current loop applications call for either 2-20 mA or 4-20 mA range.
Model GFC7000E Instruction Manual Operating Instructions FROM ANALOG I/O CONFIGURATION MENU DIAG The instrument attempt to automatically calibrate the channel … then beep. ANALOG I / O CONFIGURATION PREV NEXT DIAG AIO SET> ENTR EXIT DIAG AIO CONC_OUT_2 D/A/ CAL ERROR EXIT AIN CALIBRATED: NO EDIT EXIT DIAG AIO CONC_OUT_2 CURR-Z: 0 mV U100 UP10 UP DOWN DN10 D100 ENTR EXIT Press SET> to select the analog output channel to be configured:.
Model GFC7000E Instruction Manual Operating Instructions 6.9.4.5. AIN Calibration This is the sub-menu to conduct the analog input calibration. This calibration should only be necessary after major repair such as a replacement of CPU, motherboard or power supplies. Activate the ANALOG I/O CONFIGURATION MENU (see Section 6.9.
Model GFC7000E Instruction Manual Operating Instructions 6.9.6. Dark Calibration Test The dark calibration test interrupts the signal path between the IR photo-detector and the remainder of the sync/demod board circuitry. This allows the instrument to compensate for any voltage levels inherent in the sync/demod circuitry that might effect the calculation of CO2 concentration.
Model GFC7000E Instruction Manual Operating Instructions This calibration must be performed when the pressure of the sample gas is equal to ambient atmospheric pressure. Before performing the following pressure calibration procedure, disconnect the sample gas pump and the sample gas-line vent from the sample gas inlet on the instrument’s rear panel.
Model GFC7000E Instruction Manual Operating Instructions To cause the analyzer to measure and record a value for PRES, press. SAMPLE RANGE = 500.000 PPM CO2 =X.XXX SIGNAL I / O DIAG < TST TST > CAL SETUP NEXT SAMPLE ENTR EXIT ENTR EXIT ENTER SETUP PASS : 818 SIGNAL I / O DIAG 8 1 8 SETUP X.X ENTR EXIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE EXIT NEXT Exit at any time to return to main the SETUP menu Repeat Pressing NEXT until . . . ACTUAL PRESS : 27.
Model GFC7000E Instruction Manual Operating Instructions 6.9.9. Test Channel Output When activated, output channel A3 can be used to report one of the test functions viewable from the SAMPLE mode display. To activate the A3 channel and select a test function, follow this key sequence: SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX Continue to press NEXT until … SETUP < TST TST > CAL SAMPLE 8 DIAG ENTER SETUP PASS : 818 1 SETUP X.
Model GFC7000E Instruction Manual Operating Instructions 6.10. SETUP – COMM: Using the Analyser’s Communication Ports The MGFC7000E is equipped with two serial communication ports located on the rear panel (Figure 3-2). Both ports operate similarly and give the user the ability to communicate with, issue commands to, and receive data from the analyzer through an external computer system or terminal. By default, both ports operate on the RS-232 protocol.
Model GFC7000E Instruction Manual Operating Instructions SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL ENTER SETUP PASS : 818 SAMPLE 8 SETUP 1 8 SETUP X.X ENTR EXIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT SECONDARY SETUP MENU COMM VARS DIAG ALRM Toggle these keys to cycle through the available character set: 0-9 EXIT SETUP X.X COMMUNICATIONS MENU ID COM2 COM1 SETUP X.
Model GFC7000E Instruction Manual Operating Instructions 6.10.3. COMM Port Cable Connections There are two DB-9 connectors on the MGFC7000E rear panel. COM1 is a male connector, COM2 a female connector (Table 6-15 lists pin assignments). Teledyne Instruments offers two mating cables, one of which should be applicable for your use. Part number WR000077, a DB-9 female to DB-9 female cable, 6 feet long, allows connection of COM1 with the serial port of most personal computers.
Model GFC7000E Instruction Manual Table 6-15: Operating Instructions COM1 and COM2 DB-9 Pin Assignments Pin # COM1 (RS-232) COM2 (RS-232) COM2 (RS-485) 1 Not used Not used Not used 2 Transmit Data* Receive Data DATA - 3 Receive Data* Transmit Data Data + 4 Not used Not used Not used 5 Signal Ground Signal Ground Signal Ground 6 Not used Not used Not used 7 DATA SET READY* DATA SET READY Not used 8 REQUEST TO SEND* (=DTE Ready) REQUEST TO SEND Not used Not used Not use
Model GFC7000E Instruction Manual Table 6-16: MODE1 Operating Instructions COMM Port Communication modes ID DESCRIPTION 1 Quiet mode suppresses any feedback from the analyzer (iDAS reports, and warning messages) to the remote device and is typically used when the port is communicating with a computer program such as APICOM. Such feedback is still available but a command must be issued to receive them.
Model GFC7000E Instruction Manual Operating Instructions Press the following keys to select a communication mode for a one of the COMM Ports, such as the following example where HESSEN PROTOCOL mode is enabled: SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL ENTER SETUP PASS : 818 SAMPLE 8 SETUP 1 8 ENTR EXIT PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE SECONDARY SETUP MENU SETUP X.X COMM VARS DIAG Select which COM port to configure SETUP X.
Model GFC7000E Instruction Manual Operating Instructions 6.10.7. COM Port Baud Rate To select the baud rate of one of the COM Ports, press: SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL ENTER SETUP PASS : 818 SAMPLE 8 SETUP 1 8 ENTR EXIT PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE EXIT SECONDARY SETUP MENU SETUP X.X COMM VARS DIAG ALRM SETUP X.X Select which COM port to configure. ID COMMUNICATIONS MENU COM1 COM2 SET> EXIT COM1 MODE:0 SETUP X.
Model GFC7000E Instruction Manual Operating Instructions 6.10.8. COM Port Testing The serial ports can be tested for correct connection and output in the COMM menu. This test sends a string of 256 ‘w’ characters to the selected COM port. While the test is running, the red LED on the rear panel of the analyzer should flicker. To initiate the test press the following key sequence. SAMPLE RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE 8 SETUP ENTER SETUP PASS : 818 1 SETUP X.
Model GFC7000E Instruction Manual Operating Instructions 6.10.9.1. Ethernet Card COM2 Communication Modes and Baud Rate The firmware on board the Ethernet card automatically sets the communication modes and baud rate (115 200 kBaud ) for the COM2 port. Once the Ethernet option is installed and activated, the COM2 submenu is replaced by a new submenu, INET. This submenu is used to manage and configure the Ethernet interface with your LAN or Internet Server(s). 6.10.9.2.
Model GFC7000E Instruction Manual Operating Instructions Customer Service personnel. NOTE It is a good idea to check these settings the first time you power up your analyzer after it has been physically connected to the LAN/Internet to make sure that the DHCP has successfully downloaded the appropriate information from you network server(s). If the gateway IP, instrument IP and the subnet mask are all zeroes (e.g. “0.0.0.0”), the DCHP was not successful.
Model GFC7000E Instruction Manual Operating Instructions 6.10.9.3. Manually Configuring the Network IP Addresses There are several circumstances when you may need to manually configure the interface settings of the analyzer’s Ethernet card.
Model GFC7000E Instruction Manual Operating Instructions STEP 2: Configure the INSTRUMENT IP, GATEWAY IP and SUBNET MASK addresses by pressing: Internet Configuration Keypad Functions From Step 1 above) DHCP: OFF SETUP X.X SET> EDIT SETUP X.X EXIT FUNCTION [0] Press this key to cycle through the range of numerals and available characters (“0 – 9” & “ . ”) Moves the cursor one character left or right. DEL Deletes a character at the cursor location.
Model GFC7000E Instruction Manual Operating Instructions 6.10.9.4. Changing the Analyzer’s HOSTNAME The HOSTNAME is the name by which the analyzer appears on your network. The default name for all Teledyne Instruments Model GFC7000E analyzers is MGFC7000E. To change this name (particularly if you have more than one Model GFC7000E analyzer on your network), press. SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE 8 SETUP SETUP X.X 8 SETUP X.
Model GFC7000E Instruction Manual Operating Instructions Some keys only appear as needed. 6.11. SETUP – ALRM: Using the Gas Concentration Alarms The Model GFC7000E includes two CO2 concentration alarms Each alarm has a user settable limit, and is associated with an opto-isolated TTL relay accessible via the status output connector on the instrument’s back panel (see Section 6.13.1.1). If the CO2 concentration measured by the instrument rises above that limit, the alarm‘s status output relay is closed.
Model GFC7000E Instruction Manual Operating Instructions 6.12. SETUP – DAS: Using the Data Acquisition System (iDAS) The MGFC7000E analyzer contains a flexible and powerful, internal data acquisition system (iDAS) that enables the analyzer to store concentration and calibration data as well as a host of diagnostic parameters. The iDAS of the MGFC7000E can store up to about one million data points, which can, depending on individual configurations, cover days, weeks or months of valuable measurements.
Model GFC7000E Instruction Manual Operating Instructions or downloaded, edited and uploaded to the instrument in form of a string of plain-text lines through the communication ports. iDAS data are defined by the PARAMETER type and are stored through different triggering EVENTS in data CHANNELS, which relate triggering events to data parameters and define certain operational functions related to the recording and reporting of the data. 6.12.1.1.
Model GFC7000E Instruction Manual Operating Instructions 6.12.1.2. iDAS Parameters Data parameters are types of data that may be measured and stored by the iDAS. For each Teledyne Instruments analyzer model, the list of available data parameters is different, fully defined and not customizable. Appendix A-5 lists firmware specific data parameters for the MGFC7000E.
Model GFC7000E Instruction Manual Operating Instructions • EXITZR, EXITSP, SLPCHG (exit zero, exit span, slope change): Sampling at the end of (irregularly occurring) calibrations or when the response slope changes. These triggering events create instantaneous data points, e.g., for the new slope and offset (concentration response) values at the end of a calibration. Zero and slope values are valuable to monitor response drift and to document when the instrument was calibrated.
Model GFC7000E Instruction Manual Operating Instructions The CALDAT, STBZRO and STBSPN channels collect data based on events (e.g. a calibration operation) rather than a timed interval. This does not represent any specific length of time since it is dependent on how often calibrations are performed.
Model GFC7000E Instruction Manual Operating Instructions Triggering Events and Data Parameters/Functions for these default channels are: LIST OF CHANNELS LIST OF PARAMETERS NAME: CONC EVENT: ATIMER REPORT PERIOD: 000:01:00 NO. OF RECORDS: 800 RS-232 REPORT: OFF CHANNEL ENABLED: ON PARAMETER: CONC1 MODE: AVG PRECISION: 1 STORE NUM SAMPLES OFF CAL HOLD OFF: ON NAME: PNUMTC EVENT: ATIMER REPORT PERIOD: 001:00:00 NO.
Model GFC7000E Instruction Manual Operating Instructions These default Data Channels can be used as they are, or they can be customized from the front panel to fit a specific application. They can also be deleted to make room for custom userprogrammed Data Channels. Appendix A-5 lists the firmware-specific iDAS configuration in plain-text format.
Model GFC7000E Instruction Manual Operating Instructions 6.12.2.1. Viewing iDAS Data and Settings iDAS data and settings can be viewed on the front panel through the following keystroke sequence. SAMPLE* RANGE = 500.000 PPM CO2 =X.
Model GFC7000E Instruction Manual Operating Instructions 6.12.2.2. Editing iDAS Data Channels IDAS configuration is most conveniently done through the APICOM remote control program. The following list of key strokes shows how to edit using the front panel. SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE EXIT will return to the previous SAMPLE display. 8 SETUP ENTER SETUP PASS : 818 1 SETUP X.
Model GFC7000E Instruction Manual Operating Instructions PARAMETERS: Four parameters are included in this channel EVENT: This channel is set up to record 800 data points. To edit the name of a data channel, follow the above key sequence and then press: From the end of the previous key sequence … SETUP X.X EDIT SETUP X.X C NAME:CONC O PRINT EXIT NAME:CONC N C - - ENTR EXIT ENTR accepts the new string and returns to the previous menu.
Model GFC7000E Instruction Manual Operating Instructions 6.12.2.4. Editing iDAS Parameters Data channels can be edited individually from the front panel without affecting other data channels. However, when editing a data channel, such as during adding, deleting or editing parameters, all data for that particular channel will be lost, because the iDAS can store only data of one format (number of parameter columns etc.) for any given channel.
Model GFC7000E Instruction Manual Operating Instructions To configure a specific data parameter, press: FROM THE EDIT DATA PARAMETER MENU (see previous section) SETUP X.X 0) PARAM=CONC1, MODE=AVG PREV NEXT SETUP X.X INS DEL EDIT EXIT PARAMETERS:CONC! SET> EDIT EXIT SETUP X.X PARAMETERS: PMTDET PREV NEXT ENTR EXIT If more than on parameter is active for this channel, these cycle through list of existing Parameters. SETUP X.X SAMPLE MODE:AVG EDIT EXIT SETUP X.
Model GFC7000E Instruction Manual • Operating Instructions REPORT PERIOD: Sets how often the sample readings stored in volatile memory are processed, (e.g. average, minimum or maximum are calculated) and the results stored permanently in the instruments Disk-on-Chip as well as transmitted via the analyzer’s communication ports. The REPORT PERIOD may be set from the front panel.
Model GFC7000E Instruction Manual Operating Instructions To define the REPORT PERIOD, follow the instruction shown in section 6.12.2.2 then press: From the DATA ACQUISITION menu (see Section 6.12.2.2) Edit Data Channel Menu Use the PREV and NEXT keys to scroll to the data channel to be edited. SETUP X.X 0) CONC: PREV NEXT SETUP X.X
Model GFC7000E Instruction Manual Operating Instructions From the DATA ACQUISITION menu (see Section 6.12.2.2) Edit Data Channel Menu SETUP X.X 0) CONC: PREV NEXT SETUP X.X EDIT PRINT EXIT Press SET> key until… SETUP X.X EDIT SETUP X.X YES will delete all data in this channel.
Model GFC7000E Instruction Manual Operating Instructions To enable automatic COM port reporting, follow the instruction shown in section 6.12.2.2 then press: From the DATA ACQUISITION menu (see Section 6.12.2.2) Edit Data Channel Menu SETUP X.X PREV NEXT SETUP X.X EDIT PRINT EXIT Press SET> key until… SETUP X.X EDIT SETUP X.
Model GFC7000E Instruction Manual Operating Instructions To disable a data channel, follow the instruction shown in section 6.12.2.2 then press: From the DATA ACQUISITION menu (see Section 6.12.2.2) Edit Data Channel Menu SETUP X.X 0) CONC: PREV NEXT SETUP X.X EDIT PRINT EXIT Press SET> key until… SETUP X.X EDIT SETUP X.
Model GFC7000E Instruction Manual Operating Instructions 6.12.3. Remote iDAS Configuration Editing channels, parameters and triggering events as described in this section can be performed via the APICOM remote control program using the graphic interface shown in Figure 6-8. Refer to Section 6.13 for details on remote access to the MGFC7000E analyzer. Figure 6-8: APICOM user interface for configuring the iDAS.
Model GFC7000E Instruction Manual Figure 6-9: Operating Instructions iDAS Configuration Through a Terminal Emulation Program. Both procedures are best started by downloading the default iDAS configuration, getting familiar with its command structure and syntax conventions, and then altering a copy of the original file offline before uploading the new configuration.
Model GFC7000E Instruction Manual Operating Instructions 6.13. Remote Operation of the Analyzer 6.13.1. Remote Operation Using the External Digital I/O 6.13.1.1. Status Outputs The status outputs report analyzer conditions via optically isolated NPN transistors, which sink up to 50 mA of DC current. These outputs can be used interface with devices that accept logic-level digital inputs, such as programmable logic controllers (PLC’s).
Model GFC7000E Instruction Manual Operating Instructions The pin assignments for the Status Outputs are: Table 6-24: Status Output Pin Assignments Output # Status Definition 1 SYSTEM OK On if no faults are present. CONC VALID On if CO2 concentration measurement is valid. If the CO2 concentration measurement is invalid, this bit is OFF. 3 HIGH RANGE On if unit is in high range of DUAL or AUTO range modes. 4 ZERO CAL On whenever the instruments zero point is being calibrated.
Model GFC7000E Instruction Manual Operating Instructions There are two methods to activate control inputs. The internal +5V available from the “+” pin is the most convenient method (Figure 6.11). However, to ensure that these inputs are truly isolated, a separate, external 5 VDC power supply should be used (Figure 6.11).
Model GFC7000E Instruction Manual Operating Instructions 6.13.2.2. Help Commands in Terminal Mode Table 6-26: Terminal Mode Software Commands COMMAND Function Control-T Switches the analyzer to terminal mode (echo, edit). If mode flags 1 & 2 are OFF, the interface can be used in interactive mode with a terminal emulation program. Control-C Switches the analyzer to computer mode (no echo, no edit).
Model GFC7000E Instruction Manual Table 6-27: Operating Instructions Command Types COMMAND COMMAND TYPE C Calibration D Diagnostic L Logon T Test measurement V Variable W Warning 6.13.2.4. Data Types Data types consist of integers, hexadecimal integers, floating-point numbers, Boolean expressions and text strings. • Integer data are used to indicate integral quantities such as a number of records, a filter length, etc.
Model GFC7000E Instruction Manual Operating Instructions 6.13.2.5. Status Reporting Reporting of status messages as an audit trail is one of the three principal uses for the RS-232 interface (the other two being the command line interface for controlling the instrument and the download of data in electronic format). You can effectively disable the reporting feature by setting the interface to quiet mode (Section 6.10.6., Table 6-16).
Model GFC7000E Instruction Manual Operating Instructions This string can be altered to match your modem’s initialization and can be up to 100 characters long. To change this setting press: SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SETUP SETUP X.X SET> SAMPLE 8 COM1 MODE:0 EDIT EXIT ENTER SETUP PASS : 818 1 8 ENTR EXIT SETUP X.X SETUP X.X COM1 BAUD RATE:19200 EDIT CFG DAS RNGE PASS CLK MORE EXIT SETUP X.X EXIT returns to the previous menu SETUP X.
Model GFC7000E Instruction Manual Operating Instructions 6.13.2.7. COM Port Password Security In order to provide security for remote access of the MGFC7000E, a LOGON feature can be enabled to require a password before the instrument will accept commands. This is done by turning on the SECURITY MODE (Mode 4, Section 6.10.6). Once the SECURITY MODE is enabled, the following items apply. • A password is required before the port will respond or pass on commands.
Model GFC7000E Instruction Manual • Operating Instructions Check on system parameters for trouble-shooting and quality control. APICOM is very helpful for initial setup, data analysis, maintenance and trouble-shooting. Figure 6-8 shows an example of APICOM being used to remotely configuration the instruments iDAS feature.
Model GFC7000E Instruction Manual Operating Instructions The Hessen protocol is designed to accomplish two things: to obtain the status of remote instruments, including the concentrations of all the gases measured; and to place remote instruments into zero or span calibration or measure mode. API’s implementation supports both of these principal features.
Model GFC7000E Instruction Manual SAMPLE* Repeat the entire process to set up the COM2 port RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE 8 Operating Instructions SETUP X.X SETUP NEXT OFF ENTER SETUP PASS : 818 1 Continue pressing next until … COM1 HESSEN PROTOCOL : OFF PRIMARY SETUP MENU PREV NEXT CFG DAS RNGE PASS CLK MORE SETUP X.X SETUP X.X ID The sum of the mode IDs of the selected modes is displayed here SETUP X.X COM2 SETUP X.
Model GFC7000E Instruction Manual Operating Instructions NOTE While Hessen Protocol Mode can be activated independently for COM1 and COM2, The TYPE selection affects both Ports. 6.13.4.5. Setting The Hessen Protocol Response Mode The Teledyne Instruments implementation of Hessen Protocol allows the user to choose one of several different modes of response for the analyzer. Table 6-30: Teledyne Instruments Hessen Protocol Response Modes MODE ID MODE DESCRIPTION CMD This is the Default Setting.
Model GFC7000E Instruction Manual Operating Instructions 6.13.4.7. Setting Hessen Protocol Status Flags Teledyne Instruments’ implementation of Hessen protocols includes a set of status bits that the instrument includes in responses to inform the host computer of its condition. Each bit can be assigned to one operational and warning message flag.
Model GFC7000E Instruction Manual Operating Instructions To assign or reset the status flag bit assignments, press: SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE 8 SETUP ENTER SETUP PASS : 818 1 ENTR EXIT 8 PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE SECONDARY SETUP MENU SETUP X.X COMM VARS DIAG SETUP X.X ID EXIT ALRM EXIT COMMUNICATIONS MENU COM1 HESN COM2 EXIT Repeat pressing SET> until … SETUP X. HESSEN STATUS FLAGS EDIT SETUP X.
Model GFC7000E Instruction Manual Calibration Procedures 7. CALIBRATION PROCEDURES This section contains a variety of information regarding the various methods for calibrating a Model GFC7000E CO2 Analyzer as well as other supporting information. NOTE The procedures in this section assume that the calibration password feature is disabled (the instruments default state). If it is enabled a password prompt screen (see Section 6.3.
Model GFC7000E Instruction Manual Calibration Procedures 7.1.2. Calibration Gas Traceability All equipment used to produce calibration gases should be verified against standards of the National Institute for Standards and Technology (NIST). To ensure NIST traceability, we recommend to acquire cylinders of working gas that are certified to be traceable to NIST Standard Reference Materials (SRM). These are available from a variety of commercial sources. 7.1.3.
Model GFC7000E Instruction Manual Calibration Procedures Calibrated CO2 Gas Source of SAMPLE Gas Removed during Calibration MODEL 700 Gas Dilution Calibrator Indicating soda-lime VENT Sample Exhaust Vent Span MODEL GFC7000E Pressure Span MODEL 701 Zero Air Generator IZS Purge In Figure 7-2: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator STEP TWO: Set the expected CO2 Span Gas concentration: SAMPLE* RANGE = 500.000 PPM CO2 =X.
Model GFC7000E Instruction Manual Calibration Procedures STEP THREE: Perform the Zero/Span Calibration Procedure: SAMPLE* RANGE = 500.000 PPM < TST TST > CAL SAMPLE CO2 =XXX.X SETUP STABIL=XXX.X PPM < TST TST > CAL Set the Display to show the STABIL test function. This function calculates the stability of the COx measurement CO2 =XXX.X SETUP ACTION: Allow zero gas to enter the sample port at the rear of the instrument. Wait until STABIL falls below 1.0 ppb. M-P CAL STABIL=XXX.
Model GFC7000E Instruction Manual Calibration Procedures 7.3. Manual Calibration Checks Informal calibration checks, which only evaluate but do not alter the analyzer’s response curve, are recommended as a regular maintenance item and in order to monitor the analyzer’s performance. To carry out a calibration check rather than a full calibration, follow these steps. STEP ONE: Connect the sources of zero air and span gas as shown in Figures 7.1 or 7.2.
Model GFC7000E Instruction Manual Calibration Procedures Source of SAMPLE Gas VENT if input is pressurized Certified CO2 Gas Sample Needle valve to control flow MODEL 701 Zero Air Generator Exhaust VENT Vent Span MODEL GFC7000E Pressure Span VENT IZS Purge In Indicating soda-lime Figure 7-3: Pneumatic Connections–MGFC7000E with Zero/Span/Shutoff Valves (OPT 50) Source of SAMPLE Gas VENT if input is pressurized Certified CO2 Gas Sample Exhaust MODEL 701 Zero Air Generator VENT Vent Span
Model GFC7000E Instruction Manual Figure 7-5: Calibration Procedures Pneumatic Connections–MGFC7000E with Zero/Span Valves (OPT 52) Source of SAMPLE Gas VENT if input is pressurized Certified CO Gas MODEL 700 Gas Dilution Calibrator Sample Indicating soda-lime MODEL Exhaust VENT Vent Span GFC7000E Pressure Span External Zero Air Scrubber MODEL 701 Zero Air Generator Figure 7-6: IZS Purge In Pneumatic Connections–MGFC7000E with Zero/Span Valves with External Zero air Scrubber (OPT 53) STEP TWO
Model GFC7000E Instruction Manual Calibration Procedures STEP THREE: Perform the zero/span calibration. Zero and span checks using the zero/span valve option are similar to that described in Section 3.3, except that zero air and span gas is supplied to the analyzer through the zero/span valves rather than through the sample inlet port. The zero and cal operations are initiated directly and independently with dedicated keys (CALZ & CALS). SAMPLE* SAMPLE RANGE = 500.
Model GFC7000E Instruction Manual Calibration Procedures 7.5. Manual Calibration Checks with Zero/Span Valves Zero and span checks using the VARIOUS zero/span valve options available for the MGFC7000E are similar to that described in Section 7.3, except that the zero and calibration operations are initiated directly and independently with dedicated keys CALZ and CALS. To perform a manual calibration check of an analyzer with a valve option installed, use the following method.
Model GFC7000E Instruction Manual SAMPLE* Calibration Procedures RANGE = 500.000 PPM CAL CALZ SAMPLE CO2 =XXX.X CALS SETUP CO2 =XXX.X STABIL=XXX.X PPM < TST TST > CAL CALZ Set the Display to show the STABIL test function. This function calculates the stability of the COx measurement CALS SETUP ACTION: Allow zero gas to enter the sample port at the rear of the instrument. Wait until STABIL falls below 1.0 ppb.
Model GFC7000E Instruction Manual Calibration Procedures 7.6. Automatic Zero/Span Cal/Check (AutoCal) The AutoCal system allows unattended periodic operation of the ZERO/SPAN valve options by using the MGFC7000E’s internal time of day clock. AutoCal operates by executing SEQUENCES programmed by the user to initiate the various calibration modes of the analyzer and open and close valves appropriately. It is possible to program and run up to 3 separate sequences (SEQ1, SEQ2 and SEQ3).
Model GFC7000E Instruction Manual Calibration Procedures The following example sets Sequence #2 to do a Zero-Span Calibration every other day starting at 1 am on September 4, 2001, lasting 15 minutes, without calibration. This will start ½ hour later each iteration. MODE AND ATTRIBUTE VALUE Sequence 2 Mode ZERO-SPAN Timer Enable ON Starting Date Sept. 4, 2001 Starting Time 01:00 Delta Days 2 Delta Time 00:30 Duration 15.
Model GFC7000E Instruction Manual Calibration Procedures To program the Sequence: SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX SETUP < TST TST > CAL CALZ CALS SETUP X.X STARTING TIME:14:15 EDIT EXIT ENTER SETUP PASS : 818 SAMPLE 8 1 8 ENTR EXIT DELTA DAYS: 1 SETUP X.X EDIT PRIMARY SETUP MENU SETUP X.X ACAL DAS RNGE PASS CLK MORE CFG EXIT EXIT SETUP X.X 0 SETUP X.X 0 DELTA DAYS: 1 2 ENTR EXIT SEQ 1) DISABLED NEXT MODE EXIT SETUP X.
Model GFC7000E Instruction Manual Calibration Procedures 7.6.1. AutoCal with Auto or Dual Reporting Ranges Modes Selected SETUP C.4
Model GFC7000E Instruction Manual EPA Protocol Calibration 8. EPA PROTOCOL CALIBRATION At the writing of this manual there is no EPA requirements for the monitoring of CO2 or published calibration protocols.
Model GFC7000E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 9. MAINTENANCE SCHEDULE & PROCEDURES Predictive diagnostic functions including failure warnings and alarms built into the analyzer’s firmware allow the user to determine when repairs are necessary without performing painstaking preventative maintenance procedures.
Model GFC7000E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES Table 9-1: Item Action Freq Cal Check Req’d.
Model GFC7000E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES Table 9-2: Operating Mode* Function GFC7000E Test Function Record Date Recorded ZERO CAL STABILITY ZERO CAL CO2 MEAS Zero CAL MR RATIO SPAN CAL SAMPLE PRES PHT DRIVE SAMPLE After SLOPE OFFSET 04584 Rev A1 Warn-up SPAN CAL ZERO CAL 147
Model GFC7000E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 9.2. Predicting Failures Using the Test Functions The Test Functions can be used to predict failures by looking at how their values change over time. Initially it may be useful to compare the state of these Test Functions to the values recorded on the printed record of the final calibration performed on your instrument at the factory, p/n 04307. Table 9-3 can be used as a basis for taking action as these values change with time.
Model GFC7000E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 9.3.1. Replacing the Sample Particulate Filter The particulate filter should be inspected often for signs of plugging or contamination. We recommend that when you change the filter, handle it and the wetted surfaces of the filter housing as little as possible. Do not touch any part of the housing, filter element, PTFE retaining ring, glass cover and the o-ring. To change the filter: 1.
Model GFC7000E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 9.3.2. Rebuilding the Sample Pump The diaphragm in the sample pump periodically wears out and must be replaced. A sample rebuild kit is available – see Appendix B of this manual for the part number of the pump rebuild kit. Instructions and diagrams are included with the kit. Always perform a Flow and Leak Check after rebuilding the Sample Pump. 9.3.3.
Model GFC7000E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 4. Pressurize the instrument with the leak checker, allowing enough time to fully pressurize the instrument through the critical flow orifice. Check each fitting with soap bubble solution, looking for bubbles. Once the fittings have been wetted with soap solution, do not re-apply vacuum, as it will suck soap solution into the instrument and contaminate it. Do not exceed 15 psi pressure. 5.
Model GFC7000E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES User Notes 04584 Rev A1 152
Model GFC7000E Instruction Manual THEORY OF OPERATION 10. THEORY OF OPERATION The Model GFC7000E Gas Filter Correlation Carbon Dioxide Analyzer is a microprocessorcontrolled analyzer that determines the concentration of carbon dioxide (CO2) in a sample gas drawn through the instrument. It requires that sample and calibration gasses be supplied at ambient atmospheric pressure in order to establish a stable gas flow through the sample chamber where the gases ability to absorb infrared radiation is measured.
Model GFC7000E Instruction Manual THEORY OF OPERATION 10.1.2. Measurement Fundamentals In the most basic terms, the Model GFC7000E uses a high energy heated element to generate a beam of broad-band IR light with a known intensity (measured during Instrument calibration. This beam is directed through multi-pass cell filled with sample gas. The sample cell uses mirrors at each end to reflect the IR beam back and forth through the sample gas to generate a 2.5 meter absorption path (see Figure 10–1).
Model GFC7000E Instruction Manual THEORY OF OPERATION Measurement Cell (Pure N2) Reference Cell (N2 with CO2) Figure 10-2: GFC Wheel A GFC wheel is a metallic wheel into which two chambers are carved. The chambers are sealed on both sides with material transparent to 4.3 µm IR radiation creating two airtight cavities. Each cavity is filled with specially composed gases. One cell is filled with pure N2 (the measure cell).
Model GFC7000E Instruction Manual THEORY OF OPERATION does not absorb IR light. This results in a fluctuation in the intensity of the IR light striking the photo-detector (See Figure 10-3) that results in the output of the detector resembling a square wave. The Model GFC7000E determines the amount of CO2 in the sample chamber by computing the ratio between the peak of the measurement pulse (CO2 MEAS) and the peak of the reference pulse (CO2 REF).
Model GFC7000E Instruction Manual THEORY OF OPERATION calibration SLOPE and OFFSET values to produce the CO2 concentration which is then normalized for changes in sample pressure.
Model GFC7000E Instruction Manual THEORY OF OPERATION 10.1.4. Ambient CO2 Interference Rejection CO2 absorbs IR light very well. So well that even the narrow volume of ambient air between the IR source and the sample chamber is enough to alter the analyzer’s measured concentration of CO2. Also, ambient air, which averages around 350 ppm to 400 ppm, will vary significantly over the course of the day.
Model GFC7000E Instruction Manual THEORY OF OPERATION 10.2.1.
Model GFC7000E Instruction Manual THEORY OF OPERATION CRITICAL FLOW ORIFICE AREA OF LOW PRESSURE AREA OF HIGH PRESSURE Sonic Shockwave SPRING Figure 10-8: O-RINGS FILTER Flow Control Assembly & Critical Flow Orifice The actual flow rate of gas through the orifice (volume of gas per unit of time), depends on the size and shape of the aperture in the orifice. The larger the hole, the more gas molecules, moving at the speed of sound, pass through the orifice.
Model GFC7000E Instruction Manual THEORY OF OPERATION Particulate Filter The Model GFC7000E Analyzer comes equipped with a 47 mm diameter, Teflon, particulate filter with a 5 micron pore size. The filter is accessible through the front panel, which folds down to allow access, and should be changed according to the suggested maintenance schedule described in Table 9-1. 10.2.1.4.
Model GFC7000E Instruction Manual THEORY OF OPERATION • Through the analyzer’s keyboard and vacuum florescent display over a clocked, digital, serial I/O bus (using a protocol called I2C); • RS 232 & RS485 Serial I/O channels; • Via an optional Ethernet communications card: • Various DCV and DCA analog outputs, and • Several sets of Digital I/O channels.
Model GFC7000E Instruction Manual THEORY OF OPERATION 10.3.2. CPU The Model GFC7000E’s CPU is a, low power (5 VDC, 0.8A max), high performance, 386-based microcomputer running MS-DOS. Its operation and assembly conform to the PC/104 Specification version 2.3 for embedded PC and PC/AT applications. It has 2 MB of DRAM on board and operates at 40MHz over an internal 32-bit data and address bus.
Model GFC7000E Instruction Manual THEORY OF OPERATION 10.3.3.2. IR Source The light used to detect CO2 in the sample chamber is generated by a element heated to approximately 1100oC producing infrared radiation across a broad band. This radiation is optically filtered after it has passed through the GFC Wheel and the sample chamber and just before it reaches the photo-detector to eliminate all black body radiation and other extraneous IR emitted by the various components of those components. 10.3.3.3.
Model GFC7000E Instruction Manual THEORY OF OPERATION synchronous / demodulation circuitry of the analyzer to latch onto the most stable part of each measurement and reference IR pulse.
Model GFC7000E Instruction Manual THEORY OF OPERATION 56V Bias CO2 MEAS Photodetector Sample & Hold Circuits Variable Gain Amp Dark Switch Pre Amp TEC Control PHT DRIVE E-Test Generator Signal Conditioner CO2 Reference (x4) Thermo-Electric Cooler Signal Amplifiers Conditioner Control Circuit E Test A Gate E Test B Gate Dark Test Gate Compact Programmable Logic Device Measure Gate Measure Dark Gate Reference Gate Reference Dark Gate Phase Lock Warning M/R Sensor Segment From GFC Wheel Sen
Model GFC7000E Instruction Manual THEORY OF OPERATION The four sample and hold circuits are: DESIGNATION ACTIVE WHEN: IR BEAM PASSING THROUGH SEGMENT SENSOR PULSE IS: Measure Gate MEASUREMENT cell of GFC Wheel HIGH Measure Dark Gate MEASUREMENT Cell of GFC Wheel LOW Reference Gate REFERENCE cell of GFC Wheel HIGH Reference Dark Gate REFERENCE cell of GFC Wheel LOW Timing for activating the Sample and Hold circuits is provided by a phase lock loop circuit (PLL).
Model GFC7000E Instruction Manual Table 10-1: THEORY OF OPERATION Sync/Demod Status LED Activity LED FUNCTION STATUS OK FAULT STATUS D1 M/R Sensor Status LED flashes approximately 2/second LED is stuck ON or OFF D2 Segment Sensor Status LED flashes approximately 6/second LED is stuck ON or OFF See Section 11.1.4 for more information. 10.3.4.5. Photo-Detector Temperature Control The synch/demod board also contains circuitry that controls the IR photo-detector’s thermoelectric coolers.
Model GFC7000E Instruction Manual THEORY OF OPERATION GFC Wheel Motor Control: The GFC wheel operates from a AC voltage supplied by a multi-input transformer located on the relay board. The step-down ratio of this transformer is controlled by factory-installed jumpers to adjust for 100 VAC, 115 VAC or 230 VAC line power. Other circuitry slightly alters the phase of the AC power supplied to the motor during start up based on whether line power is 50Hz or 60 Hz.
Model GFC7000E Instruction Manual THEORY OF OPERATION DC VOLTAGE TEST POINTS STATUS LED’s RELAY PCA PN 04135 Figure 10-14: Location of relay board Status LED’s 10.3.5.2. I2C Watch Dog Circuitry Special circuitry on the relay board monitors the activity on the I2C bus and drives LED D1. Should this LED ever stay ON or OFF for 30 seconds, the watchdog circuit will automatically shut of all valves as well as turn off the IR Source and all heaters.
Model GFC7000E Instruction Manual THEORY OF OPERATION For calibration purposes, two reference voltages are supplied to the A/D converter: Reference Ground and +4.096 VDC. During calibration, the device measures these two voltages, outputs their digital equivalent to the CPU. The CPU uses these values to compute the converter’s offset and slope and uses these factors for subsequent conversions. See Section 6.9.4 for instructions on performing this calibration. 10.3.6.2.
Model GFC7000E Instruction Manual THEORY OF OPERATION 10.3.6.4. Analog Outputs The analyzer comes equipped with four analog outputs: A1, A2, A3 and a fourth that is a spare. A2 and A1 Output The first two, A2 and A1 are normally set up to operate in parallel so that the same data can be sent to two different recording devices. While the names imply that one should be used for sending data to a chart recorder and the other for interfacing with a datalogger, either can be used for both applications.
Model GFC7000E Instruction Manual THEORY OF OPERATION 10.3.6.6. External Digital I/O This External Digital I/O performs two functions. Status Outputs Logic-Level voltages are output through an optically isolated 8-pin connector located on the rear panel of the analyzer. These outputs convey good/bad and on/off information about certain analyzer conditions. They can be used to interface with certain types of programmable devices (see Section 6.13.1.1).
Model GFC7000E Instruction Manual THEORY OF OPERATION CAUTION Should the AC power circuit breaker trip, investigate and correct the condition causing this situation before turning the analyzer back on.
Model GFC7000E Instruction Manual COMM A Male RS–232 ONLY RS-232 or RS–485 COMM B Female ETHERNET OPTION THEORY OF OPERATION Control Inputs: 1–6 Status Outputs: 1–8 A1 A2 CPU Mother Board PC/104 BUS Analog Outputs KEYBOARD Optional 4-20 mA I2C BUS A3 I2C BUS A4ST FRONT PANEL DISPLAY RELAY BOARD Figure 10-16: Interface Block Diagram 10.4.1. Front Panel Interface MODE FIELD MESSAGE FIELD LOCKING SCREW CONCENTRATION FIELD STATUS LED’s FASTENER FASTENER SAMPLE A
Model GFC7000E Instruction Manual Table 10-3: THEORY OF OPERATION Front Panel Status LED’s NAME COLOR STATE DEFINITION SAMPLE Green Off On Blinking Unit is not operating in sample mode, iDAS is disabled. Sample Mode active; Front Panel Display being updated, iDAS data being stored.
Model GFC7000E Instruction Manual THEORY OF OPERATION Serial Data Display Power Watchdog Clock I2C Interface Display Controller Display Write Display Data Decoder Parallel Data Keypad Decoder I2C to Relay Board I2C to/from CPU Key Press Detect Keyboard Interrupt Status Bit 10.4.1.4. Keyboard/Display Interface Electronics From 5 VDC Power Supply Sample LED (Green) Cal LED (Yellow) Maint. Switch 2nd Lang. Switch 2 x 40 CHAR.
Model GFC7000E Instruction Manual THEORY OF OPERATION This IC performs several functions: • Using a dedicated digital status bit, it sends an interrupt signal alerting the CPU that new data from the keyboard is ready to send. • Upon acknowledgement by the CPU that it has received the new keyboard data, the I2C interface chip resets the key-depress-detect flip-flop. • In response to commands from the CPU, it turns the front panel status LEDs on and off and activates the beeper.
Model GFC7000E Instruction Manual THEORY OF OPERATION DOS Shell AI FIRMWARE Memory Handling IDAS Records Calibration Data System Status Data Analyzer Operations Calibration Procedures Configuration Procedures Autonomic Systems Diagnostic Routines PC/104 BUS ANALYZER HARDWARE Interface Handling Measurement Algorithm Sensor input Data Display Messages Keypad Analog Output Data RS232 & RS485 External Digital I/O PC/104 BUS Linearization Table Figure 10-19: Basic Software Operation 10.5.1.
Model GFC7000E Instruction Manual THEORY OF OPERATION During instrument calibration (see Chapter 7) the user enters expected values for zero and span via the front panel keypad and commands the instrument to make readings of calibrated sample gases for both levels.
Model GFC7000E Instruction Manual THEORY OF OPERATION Depending on the sampling frequency and the number of data parameters the iDAS can store several months of data, which are retained even when the instrument is powered off or a new firmware is installed. The iDAS permits users to access the data through the instrument’s front panel or the remote interface. The latter can automatically download stored data for further processing. For information on using the iDAS, refer to Sections 6.12.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11. TROUBLESHOOTING & REPAIR PROCEDURES This section contains a variety of methods for identifying the source of performance problems with the analyzer. Also included in this section are procedures that are used in repairing the instrument. CAUTION The operations outlined in this chapter are to be performed by qualified maintenance personnel only. CAUTION Risk of electrical shock.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.1.1. Interpreting WARNING Messages The most common and/or serious instrument failures will result in a warning message being displayed on the front panel. Table 11-1 lists warning messages, along with their meaning and recommended corrective action.
Model GFC7000E Instruction Manual Table 11-1: WARNING MESSAGE BENCH TEMP WARNING Warning Messages - Indicated Failures FAULT CONDITION The optical bench temp is controlled at 48 ± 5 ° C. BOX TEMP WARNING Box Temp is < 5 °C or > 48 °C. CANNOT DYN SPAN CANNOT DYN ZERO CONFIG INITIALIZED Dynamic Span operation failed Dynamic Zero operation failed Configuration and Calibration data reset to original Factory state. Concentration alarm 1 is enabled and the measured CO2 level is ≥ the set point.
Model GFC7000E Instruction Manual Table 11-1: WARNING MESSAGE SAMPLE PRES WARN SAMPLE TEMP WARN SOURCE WARNING Warning Messages – Indicated Failures (cont.) FAULT CONDITION Sample Pressure is <10 in-Hg or > 35 in-Hg Normally 29.92 in-Hg at sea level decreasing at 1 in-Hg per 1000 ft of altitude (with no flow – pump disconnected). Sample temperature is < 10oC or > 100oC. Occurs when CO2 Ref is <1250 mVDC or >4950 mVDC. Either of these conditions will result in an invalid M/R ratio.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES Functions whose values are still within the acceptable range but have significantly changed from the measurement recorded on the factory data sheet may also indicate a failure. A worksheet has been provided in Appendix C to assist in recording the value of these test functions. The following table contains some of the more common causes for these values to be out of range.
Model GFC7000E Instruction Manual Table 11-2: TEST FUNCTIONS TROUBLESHOOTING & REPAIR PROCEDURES Test Functions - Indicated Failures (cont.
Model GFC7000E Instruction Manual SAMPLE* TROUBLESHOOTING & REPAIR PROCEDURES RANGE = 500.000 PPM CO2 =X.XXX < TST TST > CAL SAMPLE SETUP ENTER SETUP PASS : 818 8 1 8 ENTR EXIT PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE EXIT SECONDARY SETUP MENU SETUP X.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES because it may be possible to recover operation of the analyzer. If after 30 – 60 seconds neither the DS5 is flashing or no characters have been written to the front panel display then the CPU is bad and must be replaced. Mother Board P/N 04069 CPU Status LED Figure 11-3: CPU Status Indicator 11.1.4.2.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES D1 – M/R Sensor Status JP4 Connector to Opto-Pickup Board D2 – Segment Sensor Status Figure 11-4: Sync/Demod Board Status LED Locations 11.1.4.3. Relay Board Status LED’s There are eight LED’s located on the Relay Board. The most important of which is D1, which indicates the health of the I2C buss.
Model GFC7000E Instruction Manual Table 11-5: LED FUNCTION TROUBLESHOOTING & REPAIR PROCEDURES Relay Board Status LED Failure Indications SIGNAL I/O PARAMETER ACTIVATED BY DIAGNOSTIC TECHNIQUE VIEW RESULT D2 Yellow WHEEL HEATER WHEEL_HEATER WHEEL_TEMP D3 Yellow BENCH HEATER BENCH_HEATER BENCH_TEMP D4 Yellow SPARE N/A N/A D5 Green SAMPLE/CAL GAS VALVE OPTION CAL_VALVE N/A D6 Green ZERO/SPAN GAS VALVE OPTION SPAN_VALVE N/A D7 Green SHUTOFF VALVE OPTION SHUTOFF_VALVE N/A D8 Green
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES independent flow meter to perform a flow check as described in Section 9.3.4. If this test shows the flow to be correct, check the pressure sensors as described in Section 11.4.6.5. 11.1.6. Typical Sample Gas Flow Problems 11.1.6.1. Flow is Zero The unit displays a SAMPLE FLOW warning message on the front panel display or the SAMPLE FLOW test function reports a zero or very low flow rate.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 5. If an IZS option is installed in the instrument, press CALZ and CALS. then suspect a bad sample/cal valve. If the flow increases 11.1.6.3. High Flow The most common cause of high flow is a leak in the sample flow control assembly or between there and the pump.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES ambient air, these natural, diurnal fluctuations might be detected by the instrument and cause its in its calculation of the CO2 concentration of the sample gas to drift. Another possible symptom of poor or stopped purge gas flow would be the inability to measure zero concentrations accurately at the end of a work day on a system that was calibrated at the beginning of a workday.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 4. Confirm the sample pressure, wheel temperature, bench temperature, and sample flow readings are correct and have steady readings. 5. Disconnect the exhaust line from the optical bench near the rear of the instrument and plug this line into the SAMPLE inlet creating a pneumatic loop. The CO2 concentration (either zero or span) now must be constant.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.3.1. Temperature Problems Individual control loops are used to maintain the set point of the absorption bench, filter wheel, and IR photo-detector temperatures. If any of these temperatures are out of range or are poorly controlled, the MGFC7000E will perform poorly. 11.3.1.1. Box or Sample Temperature Box Temperature The box temperature sensor is mounted to the motherboard and cannot be disconnected to check its resistance.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.3.1.3. GFC Wheel Temperature Like the bench heater above there are three possible causes for the GFC wheel temperature to have failed. 1. The wheel heater has failed. Check the resistance between pins 1 and 4 on the white five-pin connector just below the sample temperature sensor on the bench (pin 1 is the pointed end). It should be approximately 275 ohms. 2.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES MGFC7000E Final Test And Validation Data Sheet-p/n 04271 shipped with the unit from Teledyne Instruments). 1. The most common cause of excessive noise is leaks. Leak check and flow check the instrument described in Section 9.3. 2. Detector failure – caused by failure of the hermetic seal or over-temperature due to poor heat sinking of the detector can to the optical bench.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.4.2. DC Power Supply If you have determined that the analyzer’s AC mains power is working, but the unit is still not operating properly, there may be a problem with one of the instrument’s switching power supplies. The supplies can have two faults, namely no DC output, and noisy output.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES • D1 on the relay board is flashing, or • D1 is not flashing but pressing a key on the front panel results in a change to the display. 11.4.4. Keyboard/Display Interface The front panel keyboard, display and Keyboard Display Interface PCA (03975 or 04258) can be verified by observing the operation of the display when power is applied to the instrument and when a key is pressed on the front panel.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES IR Source Drive U5 No The IR source drive output can be verified by measuring the voltage at J16 with the IR source disconnected. It should be 11.5± 0.5 VDC. 11.4.6. Sensor Assembly 11.4.6.1. Sync/Demodulator Assembly To verify that the Sync/Demodulator Assembly is working follow the procedure below: 1. Verify that D1 and D2 are flashing. • If not check the opto pickup assembly, Section 11.4.6.2 and the GFC wheel drive, Section 11.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.4.6.3. GFC Wheel Drive If the D1 and D2 on the sync demodulator board are not flashing then: 1. Check for power to the motor by measuring between pins 1 and 3 on the connector feeding the motor. For instruments configured for 120 or 220-240VAC there should be approximately 88 VAC for instruments configured for 100VAC, it should be the voltage of the AC mains, approximately 100VAC. 2.
Model GFC7000E Instruction Manual • TROUBLESHOOTING & REPAIR PROCEDURES Measure the voltage across TP3 and TP1. With proper flow (800 sccm at the sample inlet) this should be approximately 4.5V (this voltage will vary with altitude). With flow stopped (sample inlet blocked) the voltage should be approximately 1V. If the voltage is incorrect, the flow sensor is bad, the board is bad or there is a leak upstream of the sensor. 11.4.7. Motherboard 11.4.7.1.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES described in Section 6.9.4.2 and then perform an Analog Output Step Test as described in Section 6.9.3. For each step the output should be within 1% of the nominal value listed in the table below. Table 11-11: Analog Output Test Function - Nominal Values Current Outputs OUTPUT RANGE 2 -20 4 -20 NOMINAL OUTPUT VALUES STEP % CURRENT V(250 OHMS) CURRENT V(250 OHMS) 1 0 2 mA 0.5V 4 1 2 20 5.6 1.4 7.2 1.8 3 40 9.2 2.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES Connect a jumper from the +5 pin on the Status connector to the x5V on the Control In connector. Connect a second jumper from the ‘-‘ pin on the Status connector to the A pin on the Control In connector. The instrument should switch from SAMPLE mode to ZERO CAL R mode. Connect a second jumper from the ‘-‘ pin on the Status connector to the B pin on the Control In connector.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES Problems with RS-232 connections usually center around 4 general areas: 1. Incorrect cabling and connectors. See Table 6-15 for connector and pin-out information. 2. The BAUD rate and protocol are incorrectly configured. See Section 6.10.7. 3. If a modem is being used, additional configuration and wiring rules must be observed. See Section 6.13.2.6 4. Incorrect setting of the DTE – DCE Switch is set correctly See Section 6.10.5 5.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES please use the following procedure (see the Spare Parts list in Appendix B for part numbers and kits): 1. Turn off power to the analyzer. 2. Locate the assembly attached to the sample pump, see Figure 3–11. 3. Disconnect the pneumatic connection from the flow assembly and the assembly from the pump. 4. Remove the fitting and the components as shown in the exploded view in Figure 11.6. 5.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.5.2. Removing/Replacing the GFC Wheel When Removing or replacing the GFC Wheel it is important to perform the disassembly in the following order to avoid damaging the components: 1. Turn off the analyzer. 2. Remove the top cover as described in “Getting Started” Section 3.1. 3. Open the Instrument’s hinged front panel. 4. Locate the GFC wheel/motor assembly (see Figure 3-11). 5.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11 12 Figure 11-8: Removing the GFC Wheel 11. Remove the ONE (1) screw fastening the GFC wheel/mask assembly to the gfc motor hub. 12. Remove the GFC wheel/mask assembly. 13. Follow the previous steps in reverse order to put the GFC wheel/motor assembly back together. 11.5.3. Disk-On-Chip Replacement Procedure Replacing the Disk-on-Chip, may be necessary in certain rare circumstances or to load new instrument software.
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES User Notes 04584 Rev A1 211
Model GFC7000E Instruction Manual A Primer on Electro-Static Discharge 12. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne Instruments considers the prevention of damage caused by the discharge of static electricity to be extremely important part of making sure that your analyzer continues to provide reliable service for a long time. This section describes how static electricity occurs, why it is so dangerous to electronic components and assemblies as well as how to prevent that damage from occurring. 12.
Model GFC7000E Instruction Manual Table 12-1: A Primer on Electro-Static Discharge Static Generation Voltages for Typical Activities 65-90% RH 10-25% RH 1,500V 35,000V Walking across vinyl tile 250V 12,000V Worker at bench 100V 6,000V Poly bag picked up from bench 1,200V 20,000V Moving around in a chair padded with urethane foam 1,500V 18,000V MEANS OF GENERATION Walking across nylon carpet 12.2.
Model GFC7000E Instruction Manual A Primer on Electro-Static Discharge Potentially damaging electro-static discharges can occur: • Any time a charged surface (including the human body) discharges to a device. Even simple contact of a finger to the leads of an sensitive device or assembly can allow enough discharge to cause damage. A similar discharge can occur from a charged conductive object, such as a metallic tool or fixture.
Model GFC7000E Instruction Manual • A Primer on Electro-Static Discharge As long as my analyzer is properly installed it is safe from damage caused by static discharges: It is true that when properly installed the chassis ground of your analyzer is tied to earth ground and its electronic components are prevented from building static electric charges themselves.
Model GFC7000E Instruction Manual A Primer on Electro-Static Discharge alligator clips are available for use in work areas where there is no available grounded plug. Also, anti-ESD wrist straps include a current limiting resistor (usually around one megohm) that protects you should you accidentally short yourself to the instrument’s power supply. • Simply touching a grounded piece of metal is insufficient.
Model GFC7000E Instruction Manual A Primer on Electro-Static Discharge 12.4.2. Basic anti-ESD Procedures for Analyzer Repair and Maintenance 12.4.2.1. Working at the Instrument Rack When working on the analyzer while it is in the instrument rack and plugged into a properly grounded power supply 1. Attach you anti-ESD wrist strap to ground before doing anything else. • Use a wrist strap terminated with an alligator clip and attach it to a bare metal portion of the instrument chassis.
Model GFC7000E Instruction Manual A Primer on Electro-Static Discharge 6. Disconnecting your wrist strap is always the last action taken before leaving the work bench. 12.4.2.3. Transferring Components from Rack To Bench and Back When transferring a sensitive device from an installed Teledyne Instruments analyzer to a AntiESD workbench or back: 1. Follow the instructions listed above for working at the instrument rack and work station. 2.
Model GFC7000E Instruction Manual A Primer on Electro-Static Discharge Static charges will build up on the outer surface of the anti-ESD container during shipping as the packing materials vibrate and rub against each other.
Model GFC7000E Instruction ManualAPPENDIX A - Version Specific Software Documentation APPENDIX A - Version Specific Software Documentation APPENDIX A-1: Model GFC7000E Software Menu Trees APPENDIX A-2: Model GFC7000E Setup Variables Available Via Serial I/O APPENDIX A-3: Model GFC7000E Warnings and Test Measurements Via Serial I/O APPENDIX A-4: Model GFC7000E Signal I/O Definitions APPENDIX A-5: Model GFC7000E iDAS Functions 045150110 Rev A1 221
Model GFC7000E Instruction Manual APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0 APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0 SAMPLE TEST1 Only appear if reporting range is set for AUTO range mode. CLR1,3 ENTER SETUP PASS: 818 LOW HIGH (Primary Setup Menu) CFG RANGE STABIL PRES SAMP FL PMT NORM PMT UV LAMP LAMP RATIO STR.
Model GFC7000E Instruction Manual APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0 SAMPLE TEST1 CALZ CAL Only appear if reporting range is set for AUTO range mode.
Model GFC7000E Instruction Manual APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0 SETUP ENTER SETUP PASS: 818 CFG PREV DAS ACAL1 NEXT NEXT MODE SET2 PREV ENTR 2 3 MODE SET DUAL AUTO DATE UNIT NEXT SNGL DISABLED ZERO ZERO/SPAN SPAN TIMER ENABLE STARTING DATE STARTING TIME DELTA DAYS DELTA TIME DURATION CALIBRATE PPB PPM UGM MGM % ENTR EDIT LOW3 Go To SECONDARY SETUP MENU (Fig. A-4) HIGH3 RANGE TO CAL3 Figure A-3: 045840110 Rev B.
Model GFC7000E Instruction Manual APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0 SETUP ENTER SETUP PASS: 818 CFG DAS ACAL1 RNGE VIEW PREV CLK MORE EDIT NEXT CONC PNUMTC CALDAT ZTBZRO STBSPN TEMP Cycles through lists of parameters chosen for this iDAS channel PREV PREV NEXT NX10 Selects data point to view.
Model GFC7000E Instruction Manual APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.
Model GFC7000E Instruction Manual APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0 SETUP ENTER SETUP PASS: 818 CFG DAS ACAL1 RNGE PASS COMM CLK MORE VARS ID DIAG COM1 PREV NEXT JUMP EDIT PRINT INET2 COMM - MENU TREE EDIT (Fig A-5) DAS_HOLD_OFF CONC PRECISION DYN_ZERO OFF DYN_SPAN OFF CLOCK_ADJ DHCP INSTRUMENT IP GATEWAY IP SUBNET MASK TCP PORT3 HOSTNAME4 Go To DIAG MENU TREE ON OFF 1 2 3 4 5 (Fig A-6) Only appears if a valve option is installed.
Model GFC7000E Instruction Manual APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0 SETUP . ENTER SETUP PASS: 818 CFG DAS ACAL1 RNGE PASS CLK COMM HESSEN2 ID RESPONSE MODE BCC MORE EDIT GAS LIST EDIT CMD MODE BAUD RATE TEST PORT CO2, 310, REPORTED PREV PREV NEXT INS 1 Only appears if a valve is installed.
Model GFC7000E Instruction Manual APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.
Model GFC7000E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0 APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0 Table A-1: MGFC7000E Setup Variables, Revision E.0 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE DAS_HOLD_OFF Minutes 15 CONC_PRECISION — 3 DYN_ZERO — DYN_SPAN VALUE RANGE DESCRIPTION STANDARD SETUP VARIABLES 0.5–20 Duration of DAS hold off period.
Model GFC7000E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE USER_UNITS — DIL_FACTOR — 1 DARK_CAL_DURATION Seconds 180, 60 4 DARK_MEAS_MV mV DARK_REF_MV DESCRIPTION PPB, PPM, 3, 8, UGM, MGM 3, 8, Concentration units for user interface. 0.1–1000 Dilution factor. Used only if is dilution enabled with FACTORY_OPT variable. 10–600 Duration of dark cal.
Model GFC7000E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0 SETUP VARIABLE NUMERIC UNITS ET_TARGET_CONC PPM 40, 400 ET_CONC_RANGE Conc.
Model GFC7000E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE RS232_MODE BitFlag 0 VALUE RANGE 0–65535 DESCRIPTION RS-232 COM1 mode flags. Add values to combine flags.
Model GFC7000E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0APPENDIX A-3: Warnings and Test Functions, Revision E.
Model GFC7000E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE BENCH_DERIV — 2 WHEEL_CYCLE Seconds WHEEL_PROP 1/ºC WHEEL_INTEG — WHEEL_DERIV — 2 TPC_ENABLE — CONC_LIN_ENABLE — SERIAL_NUMBER — “00000000 ” DISP_INTENSITY — HIGH I2C_RESET_ENABLE — ON FACTORY_OPT BitFlag 045840110 Rev B.
Model GFC7000E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0 0 1 2 3 4 5 6 7 8 9 10 11 12 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESCRIPTION CLOCK_FORMAT — “TIME=%H:%M:%S” Any character in the allowed character set. Up to 100 characters long. Time-of-day clock format flags. Enclose value in double quotes (") when setting from the RS-232 interface. “%a” = Abbreviated weekday name.
Model GFC7000E Instruction Manual APPENDIX A-3: Warnings and Test Functions, Revision E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0 APPENDIX A-3: Warnings and Test Functions, Revision E.0 Table A-2: NAME GFC7000E Warning Messages, Revision E.0 MESSAGE TEXT WSYSRES SYSTEM RESET WDATAINIT DATA INITIALIZED WCONFIGINIT DESCRIPTION Instrument was power-cycled or the CPU was reset. Data storage was erased.
Model GFC7000E Instruction Manual Table A-3: GFC7000E Test Functions, Revision E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0 Table A-3: GFC7000E Test Functions, Revision E.0 TEST FUNCTION NAME1 MESSAGE TEXT RANGE RANGE=50.0 PPM 3 CO RANGE=50.0 PPM 3, 7 RANGE1 RANGE1=50.0 PPM 3 CO RANGE1=50.0 PPM 3, 7 D/A #1 range in dual range mode. RANGE2 RANGE2=50.0 PPM 3 CO RANGE2=50.0 PPM 3, 7 D/A #2 range in dual range mode. RANGE3 CO2 RANGE=50.0 PPM 3, 7 D/A #3 range. 3 STABIL=0.
Model 360E Instruction Manual APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0 APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0 Table A-4: GFC7000E Signal I/O Definitions, Revision E.0 SIGNAL NAME BIT OR CHANNEL NUMBER SYNC_OK DESCRIPTION 0 1–7 1 = sync. OK 0 = sync.
Model 360E Instruction Manual SIGNAL NAME APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0 BIT OR CHANNEL NUMBER DESCRIPTION ST_SPAN_CAL 4 0 = in span calibration 1 = not in span ST_DIAG_MODE 5 0 = in diagnostic mode 1 = not in diagnostic mode ST_CONC_ALARM_1 6 0 = conc. limit 1 exceeded 1 = conc. OK ST_CONC_ALARM_2 7 0 = conc. limit 2 exceeded 1 = conc.
Model 360E Instruction Manual SIGNAL NAME APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0 BIT OR CHANNEL NUMBER DESCRIPTION CO_MEASURE 2 Detector measure reading CO_REFERENCE 3 Detector reference reading 4 Temperature MUX SAMPLE_FLOW 5 Sample flow PHOTO_TEMP 6 Photometer detector temperature TEST_INPUT_7 7 Diagnostic test input TEST_INPUT_8 8 Diagnostic test input REF_4096_MV 9 4.
Model 360E Instruction Manual APPENDIX A-5: GFC7000E iDAS Functions, Revision E.0 APPENDIX A-5: GFC7000E iDAS Functions, Revision E.0 Table A-5: MGFC7000E DAS Trigger Events, Revision E.
Model 360E Instruction Manual Table A-6: APPENDIX A-5: GFC7000E iDAS Functions, Revision E.0 MGFC7000E iDAS Functions, Revision E.0 NAME DESCRIPTION Detector measure reading mV DETREF Detector reference reading mV RATIO M/R ratio.
Model 360E Instruction Manual APPENDIX A-6: Terminal Command Designators, Revision E.0 APPENDIX A-6: Terminal Command Designators, Revision E.0 Table A-7: COMMAND Terminal Command Designators, Revision E.
Model 360E Instruction Manual Table A-8: APPENDIX A-6: Terminal Command Designators, Revision E.0 Terminal Key Assignments, Revision E.0 TERMINAL KEY ASSIGNMENTS ESC Abort line CR (ENTER) Execute command Ctrl-C Switch to computer mode COMPUTER MODE KEY ASSIGNMENTS 045840110 Rev B.
Model 360E Instruction Manual 045840110 Rev B.3 APPENDIX A-6: Terminal Command Designators, Revision E.
Model GFC7000E Instruction Manual APPENDIX B - GFC7000E Spare Parts List APPENDIX B - GFC7000E Spare Parts List NOTE The spare parts list for the Model GFC7000E is currently being compiled. For information is spare parts please contact Teledyne Analytical Instruments customer service department at 16830 Chestnut St., City of Industry, Ca. 91748 Phone: +1 626 961-9221 or 1-626-934-1500. Fax: +1 626-961-2538 Email: ai-customerservice@teledyne.com. 045840120 Rev B.
Model GFC7000E Instruction Manual Appendix C - Repair Questionnaire - GFC7000E Appendix C - Repair Questionnaire - GFC7000E Company: _________________________ Phone Number: ___________ Contact Name: _____________________________ Fax Number: _____________ Email: ____________________ Site Address: __________________________________________________________________ Can we connect to the instrument? If so, provide IP address or modem #:___________________ Model GFC7000E Serial Number:_________________ Fir
Notes and further information: _____________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ ________________________________
Model GFC7000E Instruction Manual APPENDIX D - ELECTRONIC SCHEMATICS APPENDIX D - ELECTRONIC SCHEMATICS Table D-1: DOCUMENT # 04217 04216 04136 03297 04070 03632 03976 04259 04003 04089 04468 045840140 Rev A1 List of Included Electronic Schematics DOCUMENT TITLE Interconnect List - M300E SNs >=100 Interconnect Drawing - M300E SNs >=100 PCA, 04135 Rev A, M300E Relay PCA, 03296, IR Photodetector Preamp and Sync Demodulator PCA, 04069, Motherboard, E-series PCA, 03631, 0-20mA driver PCA, 03975, Keyboard
Model GFC7000E Instruction Manual User Notes 04584 Rev A1 254