6712SR Refrigerated Sampler Installation and Operation Guide Part #69-9003-591 of Assembly #60-9004-336 Copyright © 2001. All rights reserved, Teledyne Isco, Inc.
Foreword This instruction manual is designed to help you gain a thorough understanding of the operation of the equipment. Teledyne Isco recommends that you read this manual completely before placing the equipment in service. Although Teledyne Isco designs reliability into all equipment, there is always the possibility of a malfunction. This manual may help in diagnosing and repairing the malfunction. If the problem persists, call or e-mail the Teledyne Isco Technical Service Department for assistance.
6712SR Refrigerated Sampler Safety 6712SR Refrigerated Sampler Safety General Warnings Before installing, operating, or maintaining this equipment, it is imperative that all hazards and preventive measures are fully understood. While specific hazards may vary according to location and application, take heed in the following general warnings: WARNING This instrument has not been certified for use in “hazardous locations” as defined by the National Electrical Code.
712SR Refrigerated Sampler Safety Hazard Symbols The equipment and this manual use symbols used to warn of hazards. The symbols are explained below. Hazard Symbols Warnings and Cautions The exclamation point within the triangle is a warning sign alerting you of important instructions in the instrument’s technical reference manual. The lightning flash and arrowhead within the triangle is a warning sign alerting you of “dangerous voltage” inside the product. Pinch point.
6712SR Refrigerated Sampler Table of Contents Section 1 Introduction 1.1 1.2 1.3 1.4 1.5 About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . About 700 Series Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SDI-12 Sondes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Memory to Store Monitoring Data . . . . . . . . . . . . . . . . . .
6712SR Refrigerated Sampler Table of Contents 3.6 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8 3.6.1 Information Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9 3.6.2 Help Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9 3.6.3 Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6712SR Refrigerated Sampler Table of Contents 5.3 Storage for Extended Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2 5.3.1 Selecting a Stored Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2 5.4 Programming for 700 series Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4 5.5 Programming SDI-12 Sondes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4 5.
6712SR Refrigerated Sampler Table of Contents 5.20 5.21 5.22 5.23 Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Pressurized Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Command Driven Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Command Driven Sampler Responses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6712SR Refrigerated Sampler Table of Contents Appendix B Material Safety Data Sheets Appendix C General Safety Procedures C.1 Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1 C.1.1 Adverse Atmosphere . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1 C.1.2 Deteriorated Rungs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1 C.1.3 Traffic . . . . .
6712SR Refrigerated Sampler Table of Contents 5-3 Variable-Volume Scenario . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21 5-4 Start Time Diagram for Sampler Enable Settings . . . . . . . . . . . . . . . . . . . . . . . . 5-23 5-5 Event Mark Timing Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31 5-6 Password Protected Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6712SR Refrigerated Sampler Table of Contents 7-4 8-1 8-2 8-3 C-1 6712 Remote Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8 Setting the Clock and Calendar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4 Resetting the Pump Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5 Checking and Replacing the Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6712SR Refrigerated Sampler Table of Contents xii
6712SR Refrigerated Sampler Section 1 Introduction The 6712SR Refrigerated Sampler is an integral part of Teledyne Isco’s comprehensive sampling system. The system includes the 6712SR and 6712SR Refrigerated Samplers, the 6712 Compact Sampler, and the 6712 Standard Sampler. The 6712SR is ideally suited for permanent installation in a wide variety of indoor and outdoor environments.
6712SR Refrigerated Sampler Section 1 Introduction The appendices are: • Appendix A, Menu Flowcharts • Appendix B, Material Safety Data Sheets • Appendix C, General Safety Procedures • Appendix D, Replacement Parts • Appendix E, Accessories List 1.2 About 700 Series Modules The bay on the controller’s side accepts any of Teledyne Isco’s 700 Series Modules. The 700 Series includes: • 701 pH Parameter Module for monitoring pH and temperature.
6712SR Refrigerated Sampler Section 1 Introduction sampler is equipped with the optional dial-out modem), or from a 581 Rapid Transfer Device (RTD). The RTD is a quick and simple way to transfer the data from the field to your computer. Reports are easily collected with Teledyne Isco’s Samplink software. 1.5 Pump Requirements The pump requires a pump tube made specifically for the 6712 and 6700 Series samplers. These pump tubes are easily recognized by their blue alignment collars.
6712SR Refrigerated Sampler Section 1 Introduction Table 1-1 6712SR Sampler Features (Continued) Compatible Teledyne Isco Products • 581 Rapid Transfer Device • 674 Rain Gauge • 700 Series Modules • SDI-12 Sondes • Refrigerator Temperature Sensor Real-Time Displays As the sampler runs a sampling program, it displays the program’s status. The status display may include such information as the time of the next sample, the number of the next bottle, or whether the sampler is disabled or stopped.
6712SR Refrigerated Sampler Section 1 Introduction Table 1-1 6712SR Sampler Features (Continued) Optional Dialout Modem The 6712 controller can be ordered with a factory-installed 2400 baud dialout modem. With the modem you can: •connect to the sampler and download data using Flowlink software. •program the sampler to call a contact list when an “alarm” condition exists. •use the Remote Commands to control the sampler’s operation from a remote location. See Section 5.
6712SR Refrigerated Sampler Section 1 Introduction Table 1-1 6712SR Sampler Features (Continued) EXTENDED PROGRAMMING FEATURES Pacing •Uniform Time Pacing: Sampling at regular time intervals. •Nonuniform Time Pacing: Sampling at irregular time intervals. •Random Interval Pacing: Sampling at unique random time intervals generated by the controller. •Flow Pacing: Sampling at regular flow-volume intervals. •Event Pacing: Sampling each time a user-definable event occurs.
6712SR Refrigerated Sampler Section 1 Introduction Table 1-2 6712SR Sampler Construction Materials Item Material Refrigerator Top Cover Refrigerator Body Fiberglass reinforced plastic with ultraviolet-resistant gel coat.
6712SR Refrigerated Sampler Section 1 Introduction Table 1-3 Technical Specifications for the 6712SR Refrigerated Sampler (Continued) Weight of Nickel Cadmium Battery: 4.3 lbs (2.0 kg) Weight of Lead Acid Battery: 5.7 lbs (2.6 kg) Weight of Power Pack: 6.0 lbs (2.7 kg) Weight of Battery Backed Power Pack: 6.6 lbs (3.0 kg) Controller Dimensions: Length: 10.3 in (26.0 cm) Width: 12.5 in (31.7 cm) Height: 10 in (25.
6712SR Refrigerated Sampler Section 1 Introduction Table 1-3 Technical Specifications for the 6712SR Refrigerated Sampler (Continued) Sampler Power Requirements: Nominal: 12 volts DC supplied by the refrigerator Operational: 11 to 13 Volts DC Absolute Maximum: 14.4 Volts DC Power Pack Requirements: 120VAC ±12VAC, 60 Hz., 1.0 Amp. Use only Teledyne Isco-made power packs certified by UL. (P/N Note: This is the only version certified by UL. The line cord (mains connect cable) is the “Disconnect Device.
6712SR Refrigerated Sampler Section 1 Introduction Table 1-3 Technical Specifications for the 6712SR Refrigerated Sampler (Continued) 6712 Controller: Software Specifications Suction Line: Programmable 3 to 99 ft (1 to 30 m) lengths of: • 3/8" (0.95 cm) I.D. Vinyl • 3/8" (0.95 cm) I.D. Teflon®-lined with polyethylene jacket Sample Frequency: • From 1 minute to 99 hours and 59 minutes in 1 minute increments between consecutive samples.
6712SR Refrigerated Sampler Section 1 Introduction Table 1-3 Technical Specifications for the 6712SR Refrigerated Sampler (Continued) Module Reading Conversions: • Level and 4–20 mA readings converted to flow rate units: gallons per second gallons per minute million gallons per day cubic feet per second cubic meters per second cubic meters per hour cubic meters per day liters per second • Velocity units: feet per second • Totalized flow units: gallons cubic feet liters meters per second million gallons cu
6712SR Refrigerated Sampler Section 1 Introduction Table 1-3 Technical Specifications for the 6712SR Refrigerated Sampler (Continued) Dry Weight: 120 lbs (54 kg) Refrigerator only, no controller or bottle configuration 135 lbs (61 kg) Includes refrigerator, controller, pump tube, distributor shaft extension, distributor arm and nut, and discharge tube for the 24 plastic bottle configuration.
6712SR Refrigerated Sampler Section 1 Introduction Table 1-4 Connector Specifications Pin Location Diagram POWER SOURCE 2-PIN MALE, SEALED RAIN GAUGE 9-PIN FEMALE, SEALED Pin Function A Ground B +12 volts DC A Connector Icon B E A D F I B H C G FLOW METER 6-PIN MALE, SEALED B 6-PIN FEMALE, SEALED C INTERROGATOR A D C E A D B F 8 2 (IN MODULE BAY) 7 1 3 4 +12 volts DC Ground Programmable I/O1 pin Rain gauge input Not Connected SDI-12 Data Not Connected Programmable I/O2 pin Progr
6712SR Refrigerated Sampler Section 1 Introduction 1-14
6712SR Refrigerated Sampler Section 2 Installation/Preparation 2.1 Preparing the Sampler The 6712SR is shipped to you with the controller mounted on the refrigerator and a distributor shaft extension installed. Use Figure 2-1 for reference when removing the controller. You must remove the controller when: • The internal desiccant needs recharging, the internal battery needs to be replaced or when the controller needs repair. • You install a different bottle kit.
6712SR Refrigerated Sampler Section 2 Installation/Preparation Composite Sampling Multiple-Bottle Sampling Mounting Rod 1/4-20 Set Screw Mounting Rod Distributor Shaft Extension Flat Washer Flat Washer Wing Nut Wing Nut Stop Arm Distributor arm Nut Distributor arm Assembly End of discharge tube below neck of bottle. Figure 2-1 Mounting the Controller 2.1.
6712SR Refrigerated Sampler Section 2 Installation/Preparation Table 2-1 Distributor Arm Positions and Discharge Tube Lengths for Bottle Kits 24-Bottle Kits 1000-milliliter polypropylene bottles 350-milliliter glass bottles Distributor Arm Position: D Discharge Tube: 60-9003-279 Discharge Tube Length: 20 inches (50.8 cm) Distributor Shaft: 60-9003-273 Distributor Shaft Extension: 811/16 inches (22.0 cm) 12-Bottle Kits 2.
6712SR Refrigerated Sampler Section 2 Installation/Preparation 2.2 Installing Bottle Kits Teledyne Isco ships the refrigerator from the factory with the sample bottles in place. When using the sampler for the first time, you will only need to remove the bottle lids. When installing cleaned bottles or a new kit, keep these guidelines in mind: • Do not mix glass and plastic bottles together. • Install all bottles to insure that they remain in place. • Align the racks correctly.
6712SR Refrigerated Sampler Section 2 Installation/Preparation 24-Bottle Kit 1000-milliliter polypropylene bottles. 2-Bottle Kit (Requires a locating base.) Figure 2-2 Installing the Bottle Racks Install the distributor shaft extension. Use the 811/16-inch extension for the 12and 24-bottle kits. Use the 101/2-inch extension for the 8-bottle kit. The instructions on page 2-1 show you how to remove the controller and replace the extensions for each bottle kit.
6712SR Refrigerated Sampler Section 2 Installation/Preparation Install the distributor-shaft extension. Use the 51/4-inch extension for the 2-bottle kit. The instructions on page 2-1 show you how to remove the controller and replace the extensions for each bottle kit. Next, replace the controller, and adjust the distributor arm to position A. Install the arm and stop. See Installing the Distributor Arm and Discharge Tube on page 2-2.
6712SR Refrigerated Sampler Section 2 Installation/Preparation 2.3 Suction Line The suction line is the tubing from the sampling point to the pump intake. The 6712SR uses a 3/8 -inch ID suction line of lengths 3 to 99 feet. Teledyne Isco offers vinyl or Teflon® suction lines. The Teflon® tubing has a polyethylene jacket to protect it from kinks and abrasions. The vinyl line contains a very low parts-per-million level of phenols. If phenol content affects your sample analysis, use the Teflon suction line.
6712SR Refrigerated Sampler Section 2 Installation/Preparation Table 2-2 Selecting the Right Strainer Strainer Dimensions Application Maximum Depth 10.5 3/8 Stainless Steel Ends, Polypropylene Center 1 .400 Diameter Routine applications. Vinyl Line: 22 feet (6.7m). Low flow applications Vinyl Line: 22 feet (6.7m). Highly acidic liquids. Vinyl Line: 4 feet (1.2m). 1/4 Diameter Teflon Line: 15 feet (4.5m). 7 1 3/8 Stainless Steel Low Flow .400 diameter .562 diameter 11.5 1 3/8 CPVC .
6712SR Refrigerated Sampler Section 2 Installation/Preparation a. Controller. b. Peristaltic pump. e c. Liquid Detector. f g d. Strainer c e. Suction line b h f. Stainless Steel Coupling g. A pump tube routed from the liquid detector’s intake port t h r o u g h t h e p u m p, o u t t h e detector’s discharge port, and over the side of the controller to a bulkhead fitting. a i h. Bulkhead fitting. i. Discharge tube running from the bulkhead fitting through the spring and distributor arm. j j.
6712SR Refrigerated Sampler Section 2 Installation/Preparation 2.6 Installing the Sampler The following segment briefly describes initial sampler installation. CAUTION Tests indicate that the pump produces sound levels in excess of 85db at one meter. Prolonged exposure to this sound could result in hearing loss and requires the use of protective ear plugs.
6712SR Refrigerated Sampler Section 2 Installation/Preparation When installing the sampler, be sure the vertical distance between the level of the liquid and the pump is as small as possible. CAUTION The refrigerator’s lower compartment is not watertight. Do not install in a location where the refrigerator’s lower compartment could become submerged. 2.7 Connecting External Instruments The 6712SR sampler can be used in conjunction with several types of external instruments. 2.7.
6712SR Refrigerated Sampler Section 2 Installation/Preparation WARNING Most Teledyne Isco flow meters are not certified for use in hazardous locations as defined by the National Electrical Code. Contact your Teledyne Isco representative for your equipment’s status. Never operate equipment in hazardous locations that has not been certified as safe for such use. Refer to Appendix C in the back of this manual for more safety information. Note Select Extended Programming to use the SDI-12 sonde.
6712SR Refrigerated Sampler Section 2 Installation/Preparation 2.8 Locking the Sampler The refrigerator is equipped with a lockable latch for the top cover and a lockable catch for the door handle. To lock the refrigerator, use two padlocks, one for the top cover and one for the door. 2.9 Servicing the Sampler Servicing the sampler includes: • Collecting the filled sample bottles and replacing them with clean bottles; • Collecting the data recorded by the sampler during the program.
6712SR Refrigerated Sampler Section 2 Installation/Preparation 2-14
6712SR Refrigerated Sampler Section 3 Getting Started Before programming the 6712SR, you should become familiar with its keys and how to use the screens. The 6712SR Sampler has two sets of programming modes. The first mode, Standard Programming, lets you set up typical sampling programs easily and efficiently. The second mode, Extended Programming, lets you create sophisticated programs for complex sampling applications.
6712SR Refrigerated Sampler Section 3 Getting Started The sampler will continue to use the program mode you selected, even if the sampler is turned off and on again, until you manually switch to the other programming mode. With the display backlight activated, the screen will glow for several seconds following your last keystroke. If you press a key and the screen text does not change, it is probably because the keystroke turned the backlight on, instead of performing its function. Press the same key again.
6712SR Refrigerated Sampler Section 3 Getting Started 3.1.1 Selecting a Language You have the option of having screens display in English or another language. The default is English. At the main menu, type 6712.8: ENGLISH SPANISH Use the arrow keys to move between the choices (languages other than those shown above could appear on your unit). When the language choice you want is blinking, press ↵(Enter). The ↵ (Enter) key always accepts the blinking option. 3.
6712SR Refrigerated Sampler Section 3 Getting Started 3.3 Quick View Screens Quick View screens are a special type of menu screen. They show the current program settings and let you move quickly through the program. You must change the programming style to QUICK VIEW/CHANGE to see the quick view screens, as shown below: 1 3.3.1 Paging Through Quick View Screens RUN PROGRAM VIEW REPORT OTHER FUNCTIONS Select OTHER FUNCTIONS. Select PROGRAMMING STYLE.
6712SR Refrigerated Sampler Section 3 Getting Started The following Quick View screens show you how to change the sample volume. Begin by paging through the quick view screens until you see the screen containing the sample volume setting: RUN PROGRAM VIEW REPORT OTHER FUNCTIONS ↑ 200 ml SAMPLES •To see the Quick View screens, select PROGRAM. •Press the arrow key until 200 ml SAMPLES blinks. Then press ↵.
6712SR Refrigerated Sampler Section 3 Getting Started Entering Times and Dates ENTER TIME AND DATE: HH:MM DD-MON-YY __:__ __-__-__ •Move from one field to another by pressing the arrow keys. •Change a setting in a field by typing a number. Press ↵ to accept the new setting. When setting the clock and calendar, use a 24-hour clock for times and the day-month-year format for dates. The illustration at the left shows blanks for each position that accepts an entry. The positions are called fields.
6712SR Refrigerated Sampler Section 3 Getting Started To remove days from the list: 1. Press an arrow key until the cursor is on the day you want to remove. Press ↵ to deselect it. 2. When finished, select DONE and press ↵ NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL Using the Clock Start Time Menu •Select CLOCK TIME. FIRST SAMPLE AT: 06:00 •Type the time of day and press ↵. SELECT DAYS: SU MO TU WE TH FR SA DONE •Press an arrow key until the cursor is on the correct day.
6712SR Refrigerated Sampler Section 3 Getting Started Changing the Site Name SITE: “FACTORY01” ABCDEFGHIJKLMNOPQRST UVWXYZ-& “0123456789 BACK-UP DONE SITE: “SACTORY01” ABCDEFGHIJKLMNOPQRST UVWXYZ-& “0123456789 BACK-UP DONE •Press the Down-Right arrow until the S on line 2 blinks. Then, press ↵. SITE: “SICTORY051” ABCDEFGHIJKLMNOPQRST UVWXYZ-& “0123456789 BACK-UP DONE •Press the Up-Left arrow until the I blinks. Press ↵. •Continue pressing arrow keys to select T, E, and space.
6712SR Refrigerated Sampler Section 3 Getting Started 3.6.1 Information Messages These are brief messages about the programming status, such as “ ‘ROM’ PASSED TEST” or the status message that displays when you first start up the sampler. Information messages are displayed for four seconds, unless they are terminated by pressing any key on the keypad. 3.6.2 Help Notes To see a Help note: 1. Press the ? (Help) key. 2.
6712SR Refrigerated Sampler Section 3 Getting Started 3.6.3 Warning Messages Warning messages appear when the sampler determines something is out of the ordinary. For example, the following warning appears when you type a number that is larger than the standard bottle volume: BOTTLE VOLUME IS 3500 ml (300-30000) WARNING! STANDARD BOTTLE VOLUME EXCEEDED! The 6712 does accept nonstandard volumes because you might use a nonstandard container occasionally.
6712SR Refrigerated Sampler Section 3 Getting Started See Running Programs on page 4-10. RUN PROGRAM VIEW REPORT OTHER FUNCTIONS See the screen charts in Appendix A. VIEW DATA PRINT DATA SYSTEM IDs CONFIGURE REPORTS See Reports on page 4-19. SET CLOCK SET PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE See Figure A-8, Manual Functions Screens, on page A-8.
6712SR Refrigerated Sampler Section 3 Getting Started 3-12
6712SR Refrigerated Sampler Section 4 Standard Programming The 6712SR Sampler has two sets of programming screens. The first set, called standard programming screens, lets you set up typical sampling programs easily and efficiently. The second set, extended programming screens, lets you create sophisticated programs for complex sampling applications. All of the sampling features available within the standard programming screens are available in extended programming.
6712SR Refrigerated Sampler Section 4 Standard Programming Use the arrow key to move between the two choices, feet or meters. When the desired choice is blinking, press ↵. The default unit of length for the Spanish option is meters. 4.3 Programming Examples The following table is an example of time-paced sampling. Further details about some of the menu items are provided in later sections.
6712SR Refrigerated Sampler Section 4 Standard Programming Table 4-1 Standard Program: Sample Every 15 Minutes, One Sample Per Bottle (Continued) Type the volume of the sample you want deposited in each bottle. Then, press ↵. 11 SAMPLE VOLUME: 200 ml (10-1000) 12 NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL For this example, select DELAYED START by pressing an arrow until the option blinks. Then, press ↵.
6712SR Refrigerated Sampler Section 4 Standard Programming Table 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample (Continued) Select 24. 4 NUMBER OF BOTTLES: 1 2 4 8 12 24 5 BOTTLE VOLUME IS 1000 ml (300-30000) 6 SUCTION LINE LENGTH IS 10 ft (3-99) 7 TIME PACED FLOW PACED Enter the number of pulses between sample events. 8 FLOW BETWEEN SAMPLE EVENTS: 2 PULSES (1-9999) Select BOTTLES/SAMPLE.
6712SR Refrigerated Sampler Section 4 Standard Programming 4.4 Pacing Standard Time Pacing and Flow Pacing Screens for Samplers without Flow Modules TIME PACED FLOW PACED • When programming, you will be asked to select time or flow paced sampling. Depending upon your choice, you will see one of the following screens: TIME BETWEEN SAMPLE EVENTS: __ HOURS, __ MINUTES FLOW BETWEEN SAMPLE EVENTS: ____ PULSES (1-9999) Sample pacing is the rate at which the sampler takes samples.
6712SR Refrigerated Sampler Section 4 Standard Programming The flow meter or flow logger sends the same pulse for both flow and trigger pacing, and the sampler cannot distinguish between them. Therefore, to program a sampler for trigger pacing, simply program it for flow pacing. 4.5 Distribution Distribution describes how the sampler is to deposit samples. A sample is the volume of liquid deposited in a bottle.
6712SR Refrigerated Sampler Section 4 Standard Programming Sequential 1 2 3 4 5 6 DISTRIBUTION: SEQUENTIAL STANDARD PROGRAMMING SEQUENTIAL BOTTLES/SAMPLE SAMPLES/BOT TLE RUN CONTINUOUSLY? YES NO Each bottle receives one sample from one sample event.
6712SR Refrigerated Sampler Section 4 Standard Programming 4.6 Start Times It is important to understand the difference between the time at which you run a program and the program’s start time. Running a program simply means selecting RUN from the sampler’s main menu. The start time is the time at which the sampler begins the program’s first sample interval countdown. The start time is controlled by your selections from the start time screens.
6712SR Refrigerated Sampler Section 4 Standard Programming Clock Start Time Menu Delayed Start Screen 1 NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL 1 • Select CLOCK TIME. • Select DELAYED START. FIRST SAMPLE AFTER A ___MINUTE DELAY (1-999) 2 • Enter the number of minutes to delay the countdown to the first sample. NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL FIRST SAMPLE AT 06:00 2 • Enter the time of day.
6712SR Refrigerated Sampler Section 4 Standard Programming Sampler enabled at start time A sample event at the start time occurs for time-paced programs. Sampler disabled at start time by signal from pin F A sample is not taken at start time if the sampler is disabled. No samples are taken while disabled. Figure 4-2 Start Time Diagram 4.7 Running Programs To run a program, select RUN from the main menu. You may also select YES at the RUN THIS PROGRAM NOW? screen at the end of the programming screens.
6712SR Refrigerated Sampler Section 4 Standard Programming PROGRAMMING COMPLETE RUN THIS PROGRAM NOW? YES NO If RUN had been previously Stopped and no programming changes were made. Multiple Bottles ENTER START BOTTLE: __ (1–MAX) OR For the first run using a 24-bottle base. RUN PROGRAM VIEW REPORT OTHER FUNCTIONS If applicable PLEASE WAIT! ...
6712SR Refrigerated Sampler Section 4 Standard Programming An * (asterisk) appears next to the reading if the module was unable to take a reading. If an asterisk appears, the reading displayed is the last available reading. Flow rates will be reported as zero flow if the error persists for more than five minutes (for a level or velocity reading that flow is calculated from). BOTTLE 2 IN 00:12:00 ***** gpm 5.
12SR Refrigerated Sampler Section 4 Standard Programming 4.8.2 Resume Program Selecting this option will cause the sampler to exit the manual paused state and return to normal program operation. A manual resume event will be logged. 4.8.3 View Data This option allows you to view the data recorded by the sampler. See Reports on page 4-19 for more information. 4.8.4 Grab Sample When selected, you will be asked to enter the volume to be pumped.
6712SR Refrigerated Sampler Section 4 Standard Programming If the original count remaining is less than the new pacing interval, the original count will continue to count down to the next sample event. Subsequent samples are then paced by the new interval. 4.8.9 Adjust Volume 4.8.10 Power Used Select this option to change the sample volume within the limits of the currently programmed sample distribution. If you change a sample volume, a volume changed event is logged.
6712SR Refrigerated Sampler Section 4 Standard Programming 4.11 Grab Samples Taking a Grab Sample 1 GRAB SAMPLE CALIBRATE VOLUME OPERATE PUMP MOVE DISTRIBUTOR •Select GRAB SAMPLE 2 SAMPLE VOLUME: ___ ml (10-9990) •Enter the desired volume Grab samples let you take a single sample on demand, collecting the sample in an external container (Figure 4-4).
6712SR Refrigerated Sampler Section 4 Standard Programming 4.12 Calibrate Volume The sampler delivers accurate sample volumes without calibration. If you find that sample volumes vary significantly from the programmed values, first check the suction line for proper installation. Be sure it slopes continuously downhill to the liquid source and drains completely after each sampling cycle.
6712SR Refrigerated Sampler Section 4 Standard Programming Calibrating Sample Volumes 1 RUN PROGRAM VIEW REPORT OTHER FUNCTIONS 9 •Measure the volume delivered and enter the amount here.
6712SR Refrigerated Sampler Section 4 Standard Programming Operating the Pump GRAB SAMPLE CALIBRATE VOLUME OPERATE PUMP MOVE DISTRIBUTOR CAUTION Tests indicate that the pump produces sound levels in excess of 85db at one meter. PROLONGED EXPOSURE TO THIS SOUND LEVEL COULD RESULT IN HEARING LOSS AND REQUIRES THE USE OF PROTECTIVE EAR PLUGS. SELECT DIRECTION PUMP FORWARD PUMP REVERSE PURGING. . . PRESS THE RED ‘STOP’ KEY WHEN DONE PURGED _____COUNTS 4.
6712SR Refrigerated Sampler Section 4 Standard Programming 4.15 Reports The 6712 records a variety of data while running a sampling program. It uses the data to produce four reports: • The Program Settings report, listing the program settings. • The Sampling Results report, listing the program settings, time of samples and other program events. • The Combined Results report, combining the sample event times with rainfall or module data. Interrogator Icon 4.15.
6712SR Refrigerated Sampler Section 4 Standard Programming • ↵ to start the automatic displays again. Viewing Data RUN PROGRAM VIEW REPORT OTHER FUNCTIONS 1 • Select VIEW REPORT. VIEW DATA SYSTEM IDs CONFIGURE REPORTS 2 If you select Module Data in step 3, the sampler displays the screen shown in step 4. Select the Daily Summary option and the sampler displays the stored data just as the Sampling Report was described above.
6712SR Refrigerated Sampler Section 4 Standard Programming 4.15.4 Sampling Results Report As the sampler runs a program, it records the program’s events. Program events include such items as sample events, program enables and disables, power losses, and so on. Table 4-3 lists each event with a short description. You can view the report by following the steps in Viewing the Data on page 4-19 or print the log as the Sampling Results report.
6712SR Refrigerated Sampler Section 4 Standard Programming Table 4-3 Reports: Program Events, Source Codes, and Error Codes * Indicates events and codes that occur in Standard Programming Program Events (listed in the order that they most likely would occur) *Program Run The sampler ran the sampling program. *Program Started The sampler started the sampling program. *Program Enabled (Disabled) The sampler was enabled or disabled according to program settings or an external instrument.
6712SR Refrigerated Sampler Section 4 Standard Programming Table 4-3 Reports: Program Events, Source Codes, and Error Codes (Continued) (The Source and Error Codes Listed below refer to why a sample was taken and what happened while it was being taken.) Source Codes *C Calibrate sample Sample volume delivered during calibration. *D Disable Sample event at disable time. *E Enable Sample event at enable time. Ev Event Event-paced sample event. *F Flow Flow-paced sample event.
6712SR Refrigerated Sampler Section 4 Standard Programming SAMPLER ID# 3687447734 06:32 15-FEB-01 Hardware: A0 Software: 1.
6712SR Refrigerated Sampler Section 4 Standard Programming SAMPLER ID# 3687447734 06:32 15-FEB-01 Hardware: A0 Software: 1.
6712SR Refrigerated Sampler Section 4 Standard Programming SAMPLER ID# 3687447734 06:32 15-FEB-01 Hardware: A0 Software: 1.02 BUBBLER MODULE: 2215419583 Hardware: A00000 Software: 1.07 *********** COMBINED RESULTS *********** SITE: FACTORY051 Program Started at 15:03 WE 14-FEB-01 Nominal Sample Volume = 200 ml FLOW TOTAL LEVEL RATE FLOW SAMPLE BOTTLE TIME ft cfs Mgal ------- ------ ---- ----- ----- ------------1,1 1 15:03 0.009 2.495 000000.000 1,1 2 15:18 0.012 2.576 000000.017 1,1 3 15:33 0.012 2.
6712SR Refrigerated Sampler Section 4 Standard Programming SAMPLER ID# 3687447734 06:32 19-DEC-02 Hardware: A0 Software: 1.02 *********** COMBINED RESULTS *********** SITE: FACTORY051 Program Started at 15:03 WE 18-DEC-02 Nominal Sample Volume = 200 ml FR-TEMP SAMPLE BOTTLE TIME C ------- ------ ------NO FR-TEMPERATURE ---------------------------------------- SAMPLER ID# 3687447734 06:32 19-DEC-02 Hardware: A0 Software: 0.01.
6712SR Refrigerated Sampler Section 4 Standard Programming SAMPLER ID# 3687447734 16:41 15-FEB-01 Hardware: A0 Software: 1.02 BUBBLER MODULE: 2215419583 Hardware: A00000 Software: 1.07 SITE: FACTORY051 Summary Report for 15-FEB-01 (TH) Day's Flow: 000000.054 Mgal Average Flow Rate: 0.4437 cfs 07:20 Minimum Flow Rate: 0.0237 cfs 13:25 Maximum Flow Rate: 7.
6712SR Refrigerated Sampler Section 4 Standard Programming 4.16 System IDs System IDs RUN PROGRAM VIEW REPORT OTHER FUNCTIONS • Select VIEW REPORT. VIEW DATA SYSTEM IDs CONFIGURE REPORTS • Select SYSTEM IDs. SAMPLER 6712 ID __________ HARDWARE: __ SOFTWARE: __ • Displayed briefly In addition to viewing report data and configuring reports, the View Report option on the main menu lets you view system identification information.
6712SR Refrigerated Sampler Section 4 Standard Programming CAUTION Most Teledyne Isco accessories and interfacing equipment are not certified for us in hazardous locations as defined by the National Electrical Code. Check applicable guidelines, codes, and regulations of federal, state, city, and county agencies. Refer to Appendix C in the back of this manual for more safety information.
6712SR Refrigerated Sampler Section 4 Standard Programming Table 4-4 Programming Example With 730 Module Installed (Continued) 8 9 MODE OF OPERATION: FLOWMETER LEVEL ONLY WEIR FLUME DATA POINTS EQUATION MANNING FLOW-INSERT 10 V-NOTCH RECTANGULAR CIPOLLETTI 11 V-NOTCH WEIR ANGLE: 120 90 60 45 30 22.5 If you choose Flowmeter, proceed to screen #9 below. If you choose LEVEL ONLY, proceed to screen #12. When you select Flowmeter, you will have several options for performing the flow conversion.
6712SR Refrigerated Sampler Section 4 Standard Programming Table 4-4 Programming Example With 730 Module Installed (Continued) b 17 FLOW BETWEEN SAMPLE EVENTS: 2 PULSES (1-9999) If the module is being used for level only, enter the number of pulses between sample events. Select BOTTLES/SAMPLE.
6712SR Refrigerated Sampler Section 4 Standard Programming Table 4-5 Programming Example With 750 Module Installed 1 MODULE INSERTED--DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE With the sampler off, insert the module in the bay on the controller. Turn the sampler on by pressing the On/Off key. Press ↵. Download stored data, if you want to keep it. Pressing DONE erases it and advances to the next screen. 2 6712 SAMPLER STANDARD PROGRAMMING For HELP at any screen press ? key.
6712SR Refrigerated Sampler Section 4 Standard Programming Table 4-5 Programming Example With 750 Module Installed (Continued) ROUND PIPE 13 DIAMETER = 2.00 ft 14 CURRENT LEVEL IS __.__ ft ADJUST LEVEL TO __.__ ft 15 ZERO LEVEL OFFSET: _.___ ft 16 MAXIMUM DEPTH FOR VELOCITY MEASUREMENT 2” 3” 4” 17 NUMBER OF BOTTLES: 1 2 4 8 12 24 Enter the diameter of the round pipe. Press ↵. The current level will be displayed. Enter an adjustment if needed. If the adjustment is different by more than 0.
6712SR Refrigerated Sampler Section 5 Extended Programming The 6712SR Sampler has two sets of programming screens. The first set, called standard programming screens, lets you set up typical sampling programs easily and efficiently. The second set, extended programming screens, lets you create sophisticated programs for complex sampling applications. All of the sampling features available with standard programming screens, discussed in Section 4, are available with extended programming screens.
6712SR Refrigerated Sampler Section 5 Extended Programming Each part has its own group of bottles. Because settings for each part are independent of each other, the sampler, when running a two-part program, fills each bottle group as if they were being filled by two different programs. Two-part programs finish when both parts are done. Two-part programming is ideal for sampling storm events.
6712SR Refrigerated Sampler Section 5 Extended Programming The current program is the program you see when you select PROGRAM from the main menu and page through the programming screens. It is the program the sampler uses when you select RUN from the main menu. You cannot delete programs, but you can modify them as often as necessary. To run a program with different settings without losing the settings in your current program, select a stored program and modify its settings.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.4 Programming for 700 series Modules The bay on the controller’s side accepts any of Teledyne Isco’s 700 Series Modules. These optional accessories are not required to operate the 6712. However, the modules offer an economical way to combine flow-rate or parameter monitoring with sampling.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.6.1 Storm Water Runoff Sampling Storm water runoff sampling is an ideal application for two-part programming. Typical monitoring permits require that a series of timed samples be taken during the initial portion of the storm event, followed by flow paced samples during the remainder of the event. This example assumes that a Rain Gauge and a Bubbler Flow Module are connected to the sampler.
6712SR Refrigerated Sampler Section 5 Extended Programming once enabled, stay enabled, no sample taken when the sampler is enabled; first sample taken 20 minutes after the sampler is enabled. • Part ‘B’ pause and resumes: None. • Start time: Delayed start of 1 minute. • 730 Bubbler Module: Installed Table 5-2 Storm Water Runoff Sampling 1 6712 SAMPLER EXTENDED PROGRAMMING For HELP at any screen press ? key. 2 RUN “EXTENDED 1” PROGRAM VIEW REPORT OTHER FUNCTIONS Select PROGRAM.
6712SR Refrigerated Sampler Section 5 Extended Programming Table 5-2 Storm Water Runoff Sampling (Continued) 11 PROGRAM MODULE? YES NO 12 MODE OF OPERATION FLOWMETER LEVEL ONLY 13 14 15 16 Select FLOWMETER WEIR FLUME DATA POINTS EQUATION MANNING FLOW-INSERT Select FLOW-INSERT. WEIR/ORIFICE TYPE V-NOTCH ROUND Select ROUND. FLOW INSERT SIZE: 6" 8" 10" 12" Select 12. NEW MODULE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE CURRENT LEVEL IS ___.___ ft. 17 Select YES Select DONE.
6712SR Refrigerated Sampler Section 5 Extended Programming Table 5-2 Storm Water Runoff Sampling (Continued) Enter 0. 23 0 RINSE CYCLES (0-3) 24 RETRY UP TO 0 TIMES WHEN SAMPLING (0-3) 25 ONE-PART PROGRAM TWO-PART PROGRAM 26 24 BOTTLES AVAILABLE ASSIGN BOTTLES 1 THRU 6 TO PART ‘A’ (1-23) Select UNIFORM TIME PACED. 27 UNIFORM TIME PACED FLOW PACED EVENT PACED NONUNIFORM TIME Enter 0 for hours, 5 for minutes.
6712SR Refrigerated Sampler Section 5 Extended Programming Table 5-2 Storm Water Runoff Sampling (Continued) 35 36 37 38 ENABLE: RAIN LEVEL FLOW NONE Select RAIN. ENABLE: RAIN Select AND. AND OR Note: All valid enable options will appear here, so your screen may differ. DONE ENABLE: RAIN AND LEVEL FLOW NONE Select LEVEL. “RAIN” SET POINT: 0.15 INCHES PER __HOURS, __MINUTES Enter 0.15 for inches, then enter 30 minutes.
6712SR Refrigerated Sampler Section 5 Extended Programming Table 5-2 Storm Water Runoff Sampling (Continued) 48 PACED BY: FLOW PULSES FLOW MODULE VOLUME Select FLOW MODULE VOLUME. 49 FLOW BETWEEN SAMPLE EVENTS: 2500 gal (0.001-99999) 50 SAMPLE AT START? YES NO 51 1 BOTTLES PER SAMPLE EVENT (1- 18) 52 SWITCH BOTTLES ON: NUMBER OF SAMPLES TIME 53 SWITCH BOTTLES EVERY 1 SAMPLES (1- 50) 54 RUN CONTINUOUSLY? YES NO 55 SAMPLE VOLUME: 200 ml (10-1000) Select RAIN.
6712SR Refrigerated Sampler Section 5 Extended Programming Table 5-2 Storm Water Runoff Sampling (Continued) Select YES. 61 RESET RAIN HISTORY? YES NO Select SET POINT. 62 “LEVEL” CONDITION: SET POINT RANGE RATE OF CHANGE “LEVEL” SET POINT 0.25 ft Enter 0.25. 63 (0.100-30.000) 64 ONCE ENABLED, STAY ENABLED? YES NO 65 SAMPLE AT ENABLE? YES NO 66 START OF SAMPLING AFTER A 15 MINUTE DELAY (1-999) 67 PAUSE RESUME 1. HH:MM DD HH:MM DD 2.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.7 Suction Head Extended Screen for Suction Head Settings AUTO SUCTION HEAD ENTER HEAD Suction head is the vertical distance between the surface of the liquid source and the sampler’s pump. At each sample event, the sampler determines the suction head automatically. In extended programming, you can manually enter the suction head if you want. Select ENTER HEAD only when the head remains stable and you can measure it accurately.
6712SR Refrigerated Sampler Section 5 Extended Programming conditions becomes true. The default distribution setting for event-paced sampling is sequential (each time an event occurs, the sampler takes one sample, placing it in one bottle). However, any distribution setting may be used.
6712SR Refrigerated Sampler Section 5 Extended Programming Table 5-3 Event Paced Sampling (Continued) 4 “LEVEL” SET POINT: 1.000 ft Enter a set point of 1 foot. ( 0.001 - 30.00) 5 ENABLED WHEN: ABOVE SET POINT BELOW SET POINT Select ABOVE SET POINT. 6 EVENT02: RAIN LEVEL FLOW TEMP5 pH5 SP_CO5 DO5 DONE Choose RATE OF CHANGE. 7 “LEVEL” CONDITION: SET POINT RANGE RATE OF CHANGE 8 CONDITION IS TRUE WHEN “LEVEL” RISES FALLS Choose RISES.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.9.3 Nonuniform Time Pacing The extended time pacing options are uniform time and nonuniform time. “Uniform time” pacing is identical to “time” pacing in standard programming. Nonuniform time pacing uses an irregular interval between sample events; each interval is individually programmable. There are three types of nonuniform time pacing: • Nonuniform clock time pacing. • Nonuniform interval pacing. • Random interval pacing. 5.9.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.10 Distribution Distribution describes how the sampler is to deposit samples. A sample is the volume of liquid deposited in a bottle. A sample event includes the full sampling cycle and may deposit a sample into more than one bottle. You can program the sampler for five distribution methods: • Sequential • Bottles per sample • Samples per bottle • Composite • Multiple Bottle Compositing 5.10.
6712SR Refrigerated Sampler Section 5 Extended Programming Time Switched Distribution 1 2 BOTTLES PER SAMPLE EVENT (1-max) •Enter the number of bottles that should receive samples at each sample event. 2 SWITCH BOTTLES ON: NUMBER OF SAMPLES TIME • Select TIME. 3 If the program settings instruct the sampler to take a sample at the start time, the sampler also draws a sample at the switch time. When this occurs, the pacing will also be reset at the switch time.
6712SR Refrigerated Sampler Section 5 Extended Programming Sequential EXTENDED PROGRAMMING 1 2 3 4 5 6 Each bottle receives one sample from one sample event. DISTRIBUTION: SEQUENTIAL 1 BOTTLES PER SAMPLE EVENT (1-24) S W I T C H B O T T L E S ON: NUMBER OF SAMPLES TIME SWITCH BOTTLES EVERY 1 SAMPLES (1-50) RUN CONTINUOUSLY? YES NO Bottles-per-Sample EXTENDED PROGRAMMING 1 1 2 2 3 3 Multiple bottles receive a sample from one sample event.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.11 Flow Proportional Sample Volumes The 6712 can collect flow proportional sample volumes. For some sampling protocols, these flow-weighted, variable-volume samples can be more representative of the stream. This is because stream characteristics often change as the flow fluctuates. Variable-volume sampling is possible when the sampler is programmed for Uniform Time pacing.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.11.1 Calculating Variable Sample Volume Settings The entry for the SAMPLE VOLUME display can be calculated based on expected flow rates and collection volume requirements. The number of pulses or flow module volume that you enter should ensure that the minimum volume of liquid required for analysis is collected without exceeding the bottle volume capacity of the sampler.
6712SR Refrigerated Sampler Section 5 Extended Programming 666,667 gal 9400 ml 500,000 gal 3000 ml e Gr 0 1 2 s at e 3 x tE c pe st E Lea te low dF dF cte xp e 4 Hours 5 low 6 7 8 Figure 5-3 Variable-Volume Scenario 5.11.2 Calculations when Using an External Flow Meter SAMPLE VOLUME: 10 ml FOR EVERY ___ PULSES The highest possible SAMPLE VOLUME value is determined by the minimum volume required for analysis and the least expected flow rate over the sampling duration.
6712SR Refrigerated Sampler Section 5 Extended Programming These factors would be applied to the following equation. (Greatest Total Flow Vol) / (Flow Pulse Vol) 10 ml for every __ pulses = (Total Bottle Vol) / (10 ml per unit) From the scenario we can apply the factors to the equation. (666,667 gallons / 100 gallons) 10 ml for every 7.0 pulses = (9,400 ml / 10 ml) At high flow rates, a number below 7 would overfill the bottle. At low flow rates, a number greater than 16 would collect too little sample.
6712SR Refrigerated Sampler Section 5 Extended Programming When the sampler detects that it is disabled at start time, it suspends the program until it is enabled. Once enabled, the sampler collects a sample only when programmed to take a sample at the enable time. The diagram in Figure 5-4 shows the sampler’s response when enabled or disabled at the start time. Sampler enabled at start time A sample event at the start time always occurs for time-paced programs (except for random).
6712SR Refrigerated Sampler Section 5 Extended Programming 5.12.1 Sampler Enable This example uses a sampler with an attached 730 Bubbler Flow Module, Rain Gauge and an SDI-12 sonde at address 5 with four parameters. Table 5-4 Sampler Enable ENABLE: In this example, you will be enabling the sampler if it rains .5” in 2 hours and the level goes above 1 foot. 1 NONE PROGRAMMED ENABLE: RAIN LEVEL FLOW TEMP5 pH5 SP_CO5 DO5 NONE Select RAIN. 2 ENABLE: RAIN Choose AND.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.13 Sampler Enable Responses In addition to programming enable conditions, extended programming lets you control the sampler’s response to its enable state. You can: • Stay enabled after the first enable. • Set up a repeatable enable. • Take a sample at enable or disable. • Delay the start of sampling after the enable. • Reset the sampling interval countdown each time the sampler is enabled.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.13.4 Delay To Start Of Sampling If you have selected ONCE ENABLED STAY ENABLED and have chosen to not sample at the enable, you have the opportunity to suspend the start of sampling. This choice is useful for storm programs that require a flow-paced sample taken after the initial grab sample. START SAMPLING AFTER A ___ MINUTE DELAY (1-999) 5.13.5 “Dry Period” Option For samplers used with a 0.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.14 Pauses/Resumes Pause and Resume Screens Flowcharts containing the pause and resume screen appear in Appendix A. PAUSE RESUME 1. 17:00 MO 08:00 TU 2. 17:00 TU 08:00 WE CLEAR DONE SAMPLE AT RESUME? YES NO Pause and resume settings, available in extended programming, create an intermittent sampling schedule. A program with pauses and resumes begins sampling at its programmed start time, continuing until the first pause time and day of the week.
6712SR Refrigerated Sampler Section 5 Extended Programming Adjust Suction Head – If you have chosen to manually enter the suction head, and are in the manual paused state, the scrolling menu will have an option to adjust the suction head. If you make changes to the suction head value, the sampler logs a changed suction head event. 5.16 Sampling Reports The 6712 records a variety of data while running a sampling program. Sample reports can be found in Section 4.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.18.2 Programming Style The sampler has two different styles of programming screens: normal and quick view. These are both explained in Section 3. When you are in extended programming, the default style is Quick View, which allows you to move more quickly through the screens. If you prefer a step-by-step display for the screens, switch to the Normal setting. 5.18.
6712SR Refrigerated Sampler Section 5 Extended Programming Bottle Full Detect BOTTLE FULL DETECT? YES NO Bottle Full Detection – The sampler can detect a full bottle condition when collecting samples in a single bottle (composite sampling). When the sampler detects a full bottle, it stops the sampling routine. A BOTTLE FULL condition will also stop a sampling routine that is set to RUN CONTINUOUSLY. The Bottle Full Detect screen allows you to enable or disable this feature.
6712SR Refrigerated Sampler Section 5 Extended Programming • 3-second pulse at the beginning of forward pumping only, marking the time the sampler deposited the sample. • Variable duration pulse generated during the pump cycle, from the beginning of the pre-sample purge to the end of the postsample purge. • Variable duration pulse generated only during forward pumping. Figure 5-5 shows event mark timing diagrams. Bottle Number – Concurrent with the event mark, the sampler also sends a bottle number signal.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.18.6 Serial Output Programming the serial output PERIODIC SERIAL OUTPUT? YES NO Serial ASCII data can be output through the controller’s interrogator port. The output frequency is user-selectable: every 15 seconds, every minute, every 5 minutes, every 15 minutes, or by command. To receive serial data as a periodic output: 1. Select YES at screen asking if you want periodic serial output.
6712SR Refrigerated Sampler Section 5 Extended Programming Table 5-5 Serial Data Codes Identifier Parameter Units B?1 Bottle number and time days since 1-Jan 1900 CR Current day’s rain tips (rolls over every 256 tips) CS2 Checksum unsigned long DE Description 6712 Sampler FL Flow cubic meters per second ID Unit specific identifier 10 digit unsigned long LE Level meters MO Model 6712 PE Percent percent of full scale PH pH none PR Previous day’s rain tips (rolls over every 2
6712SR Refrigerated Sampler Section 5 Extended Programming Table 5-5 Serial Data Codes (Continued) Identifier Parameter Units CLa Chloride milligrams per liter CPa Chlorophyll micrograms per liter FLa Flow rate cubic meters per second VOa Flow volume cubic meters xxDa4 Other data not listed above • The three most recent samples will have entries as B?. The “?” will be the first bottle number in the set. If the sample is a grab sample, it will be shown as B0.
6712SR Refrigerated Sampler Section 5 Extended Programming From Standby RUN PROGRAM VIEW REPORT OTHER FUNCTIONS ENTER PASSWORD: ____ Programming Screens VIEW DATA SYSTEM IDs CONFIGURE REPORTS ENTER PASSWORD: ____ ENTER PASSWORD: ____ MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE ENTER PASSWORD: ____ STOP PROGRAM RESUME PROGRAM VIEW DATA GRAB SAMPLE SAMPLINK REPORT FLOWLINK REPORT CUSTOM REPORT ALL REPORTS While Running a Program Stop Key SAMPLE 3 IN 00:03:26 ENTER PASSWORD: ____ On/Off Key F
6712SR Refrigerated Sampler Section 5 Extended Programming DO YOU HAVE A RAIN GAuGE CONNECTED? YES NO • Select YES. RAIN GAUGE: 0.01 inch TIP 0.1 mm TIP When asked if you have a rain gauge connected, select YES to record the rain gauge readings and add RAIN as a sampler enable option. Select the appropriate tip for the rain gauge. The sampler will then create a rain memory partition and update the available sampler enable conditions. Select NO to disable the readings and the sampler enable option.
6712SR Refrigerated Sampler Section 5 Extended Programming To configure an attached temperature sensor, select “FR-TEMP” a s t h e i n p u t f o r I / O 3 . T h e 6 7 1 2 w i l l d i s p l ay t h e N E W HARDWARE SETUP screen. Select DONE and the 6712 reconfigures its memory partitions to record the temperature readings. Temperature readings are updated approximately every 2.5 seconds. The readings are stored as FR-TEMP readings at the data storage interval.
6712SR Refrigerated Sampler Section 5 Extended Programming Example: Dialout Alarms If the controller is equipped with a dialout modem, it can be programmed to notify a telephone or pager contact list when an alarm condition is true. Before proceeding the I/O conditions must be defined. Refer to section 5.19.6. These steps program a dialout alarm when the condition defined for I/O1 is true.
6712SR Refrigerated Sampler Section 5 Extended Programming Extended characters, the Comma (,), Pound (#), and Star (*), also may be included in the 25 digits. To enter an extended character, press the Decimal Point key on the Avalanche keypad. The Avalanche displays a screen used to select and enter the extended characters. Each comma entered will cause the modem to pause for two seconds before proceeding with the remaining digits. Phone numbers that are no longer needed can be removed.
6712SR Refrigerated Sampler Section 5 Extended Programming When you enter each of the (up to three) phone numbers for the sampler to call, if text messaging is active, a request for the TAP (Telocator Alphanumeric Protocol) server number and communications parameters will be made. As with voice messaging, the sampler will dial each telephone number up to five times per alarm condition, until the message is acknowledged. 5.
6712SR Refrigerated Sampler Section 5 Extended Programming • Add or remove a SDI-12 sonde parameter with the Hardware Setup • Reinitialize the sampler The stored data is lost as the sampler reformats its memory. As a precaution, before clearing the memory, the sampler display will show the following message: INTERVAL CHANGED -DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE If you want to keep the stored data, download it before proceeding.
6712SR Refrigerated Sampler Section 5 Extended Programming For more consistent sample volumes from lines under constant pressure, calibrate the samples. Teledyne Isco does not recommend sampling from pressurized lines. 5.22 Command Driven Operation The 6712 controller can be used as a slave to an external controller. The external controller is responsible for determining when to take a sample, how much volume to pump, and where to put the sample.
6712SR Refrigerated Sampler Section 5 Extended Programming 5.23 Command Driven Sampler Responses The 6712 Sampler’s response string is the same for each of the three commands. A sample response string is shown below: MO,6712,ID,2424741493,TI,35523.50000,STS,1,STI,35523.4187 5,BTL,2,SVO,100,SOR,0,CS,xxxx Note that the sampler’s response is a series of alphabetical and numerical pairs. The alphabetical characters are headings that describe the following numerical value or code.
6712SR Refrigerated Sampler Section 5 Extended Programming 8 = PUMP LATCH OPEN. 9 = SAMPLER SHUT OFF (while sampling). 11 = NO DISTRIBUTOR. 12 = SAMPLE IN PROGRESS. CS,xxxx is the byte-by-byte sum of the entire string “MO,...,CS,” For commands that are not understood (INVALID COMMAND or CHECKSUM MISMATCH) or when no samples have been taken, the most recent bottle information is left off the response string.
6712SR Refrigerated Sampler Section 6 SDI-12 Sondes In order to provide you with the flexibility to use the water parameter sensors that best meet your needs, Teledyne Isco has implemented support for SDI-12 protocol. Support of this standard allows you to use SDI-12 compatible devices in conjunction with your 6712SR sampler in Extended Programming mode. 6.
6712SR Refrigerated Sampler Section 6 SDI-12 Sondes SDI-12 Sonde Data Units: Parameter Temperature Conductivity Specific Conductance Total Dissolved Solids Salinity Dissolved Oxygen pH ORP Level Ammonium-Nitrogen Ammonia-Nitrogen Nitrate-Nitrogen Turbidity Chloride Chlorophyll Other Range -40 - 100 °C 0 - 100 mS/cm 0 - 100 mS/cm 0 - 90 g/l 0 - 70 ppt 0 - 20 mg/l 0 - 14 -999 - 999 mv 0 - 6.5279 m 0 - 200 mgN/l 0 - 200 mgN/l 0 - 200mgN/l 0 - 5000 NTU 0 - 1000 mg/l 0-400 ug/l -99999.0-99999.0 Resolution 0.
6712SR Refrigerated Sampler Section 6 SDI-12 Sondes Press the arrow keys until “SDI-12 DATA” is blinking, then press ↵. You will see the screen below: NO SDI-12 SONDE ↑ SETUP KNOWN SONDES SCAN SDI-12 SONDES ↓ Select “SCAN SDI-12 SONDES” and press ↵. The system will scan to see what sondes are connected. If it detects a different sonde, it will indicate so, and advance to the setup screens. 6.
6712SR Refrigerated Sampler Section 6 SDI-12 Sondes Connecting Other SDI-12 Sondes SELECT ADDRESS OF SONDE TO SETUP: __ __ DONE 3 •Select the address of the sonde you want to set up. 3 Company Model Optional Information SONDE AT ADDRESS __ HAS __ DATA VALUES 4 •You can wait a few seconds, or press ↵ to advance to the next screen: DATA FROM SONDE __ 1 = OTHER 2 = OTHER 3 = OTHER ---> 5 •Press an arrow key until the cursor is on the data you want to identify. Then, press ↵.
6712SR Refrigerated Sampler Section 6 SDI-12 Sondes Select this screen to display the following: NO SDI-12 SONDE SETUP KNOWN SONDES SCAN SDI-12 SONDES Select “SETUP KNOWN SONDES” and then indicate the address of the sonde you want to setup. A screen containing the sonde model and number of data values will appear briefly. Then you will be asked to verify, parameter by parameter, what data you want to store.
6712SR Refrigerated Sampler Section 6 SDI-12 Sondes 6.6.1 The Calibration Screens To access the calibration screens from the main menu, select Other Functions, Hardware. Advance through the screens until you see one that displays “SDI-12 DATA.” Select this screen to display the following: NO SDI-12 SONDE SETUP KNOWN SONDES SCAN SDI-12 SONDES Select “SETUP KNOWN SONDES” and then select “DONE.” You will see a screen similar to the following that will list the parameters you can calibrate.
6712SR Refrigerated Sampler Section 6 SDI-12 Sondes Use the arrow keys to select the desired par am eter to validate. Any parameter can be validated. You will be prompted to place the probe in your validation standard. When stable, press ↵. The reading will be displayed at the bottom of the screen. When you are done validating, select “DONE” to return to the Hardware options screens. 6.
6712SR Refrigerated Sampler Section 6 SDI-12 Sondes 6-8
6712SR Refrigereated Sampler Section 7 Remote Operation The 6712 controller can be operated remotely from external devices. External controlling devices may be a computer via RS-232 serial communications or a telephone via touch-tone comm a n d s. R e m o t e t e l e p h o n e o p e r a t i o n r e q u i r e s t h e factory-installed 2400 baud talking modem. 7.1 Computer Operation The 6712 controller can be operated from a remote location using a computer.
6712SR Refrigereated Sampler Section 7 Remote Operation 7.1.1 Menu Control Menu commands can be entered from a remote location using a computer. The menu commands are listed in Table 7-1. To access the menu commands, type MENU at the “>” prompt and press ENTER. The screen will return a list of the available options. Note that the current state of the controller will determine the available menu options. Then, type the desired command(s), confirming with ENTER.
6712SR Refrigereated Sampler Section 7 Remote Operation Note You can retrieve the sampling reports at any time using the same serial connection. Type “Q” to exit the menu. Then at the > prompt, type “REPORT” and press Enter. The sampler controller will send reports as configured. 7.1.2 External Program Control Under external program control, the pump and distributor operate according to received commands.
6712SR Refrigereated Sampler Section 7 Remote Operation Take a Sample – This directs the sampler to take a sample. The command specifies how much volume to take, and which bottle to place the sample in. The sampler responds by taking a sample and returning a status string that reports the new status. A command to take a sample would appear as: BTL,2,SVO,100,CS,1039 This example instructs the sampler to place the distributor arm over bottle 2 (BTL,2) and take a 100 ml sample (SVO,100).
6712SR Refrigereated Sampler Section 7 Remote Operation 23 = VOLUME OUT OF RANGE. (the sample volume requested is outside its range (10-9990 ml) STI,35523.41875, is the most recent sample time. BTL,2, is the bottle receiving the most recent sample. SVO,100, is the most recent sample’s volume. SOR,0, is the results of attempting the most recent sample. The possible results are: · 0 = SAMPLE OK. · 1 = NO LIQUID FOUND. · 2 = LIQUID LOST (not enough liquid to satisfy requirements.
6712SR Refrigereated Sampler Section 7 Remote Operation Table 7-2 Serial Data Codes (Continued) Identifier Parameter Units PE Percent percent of full scale PH pH none PR Previous day’s rain tips2 RA Rain tips2 RTE Refrigerator Temperature degrees celsius SS Sampler enable state logical TE Temperature degrees celsius TI Current time days since 1-Jan-1900 VE Velocity meters per second VO Volume cubic meters VSI Velocity signal strength percent VSP Velocity spectrum strengt
6712SR Refrigereated Sampler Section 7 Remote Operation Table 7-2 Serial Data Codes (Continued) Identifier Parameter xxD_ 7.1.3 Remote Control of Sampler Keypad 4, 5 Units Other data not listed above 1 The three most recent samples will have entries as B?. The “?” will be the first bottle number in the set. If the sample is a grab sample, it will be shown as B0. 2 Rainfall tips roll over every 256 counts. 3 The checksum does not include the checksum, carriage return, and linefeed.
6712SR Refrigereated Sampler Section 7 Remote Operation 3. If the command is valid for its current state of operation, the sampler responds with “[command number], please acknowledge.” If the command is not valid the sampler responds with the command number followed by three quick beeps. Enter a valid command. 4. Acknowledge the valid command by pressing the number. For the example, press 1 . 5. If the number is incorrect, the sampler responds with your entry followed by three quick beeps.
6712SR Refrigerated Sampler Section 8 Maintenance This section describes how to maintain and troubleshoot your 6712SR sampler. If you think your unit requires repair, or if you have questions concerning its operation or maintenance, contact Teledyne Isco’s Technical Service Department: Phone: (800) 228-4373 (402) 464-0231 FAX: (402) 465-3022 E-mail: IscoService@teledyne.com 8.1 Maintenance Checklist • Clean the condenser coil annually. • Clean the refrigerator’s air filter every 3 months.
6712SR Refrigerated Sampler Section 8 Maintenance Make sure the two pump drain holes (located on the bottom right-hand side of the pump, beneath the pump band) are open and free of debris or buildup. 8.2.3 Sample Bottles The sample bottles have a wide mouth to facilitate cleaning. Wash them with a brush and soapy water, or use a dishwasher. You can autoclave the glass bottles. 8.2.
6712SR Refrigerated Sampler Section 8 Maintenance protocols are based on U.S. Environmental Protection Agency Publications EPA-600/4-77-039 (Sampling of Water and Wastewater by Dr. Phillip E. Shelley). 8.3.1 Teledyne Isco Glass Sample Bottles 1. One spectro-grade acetone rinse. 2. Dishwasher cycle (wash and tap water rinse, no detergent). 3. Acid wash with at least 20 percent hydrochloric acid. 4. Dishwasher cycle (wash and tap water rinse, no detergent). 5. Replace in covered Teledyne Isco tubs. 8.3.
6712SR Refrigerated Sampler Section 8 Maintenance 8.4 Maintenance Screens When you select OTHER FUNCTIONS from the main menu, and then select Maintenance, the following screen will be displayed: SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS Using these options, you can set the sampler’s time and date, set the number of pump counts for the pump tube warning, reset the battery warning for the internal battery, and run a set of diagnostic tests for troubleshooting purposes. 8.4.
6712SR Refrigerated Sampler Section 8 Maintenance Table 8-1 Setting the Clock and Calendar (Continued) 5 8.4.2 Pump Tube Alarm SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS Press STOP to return to the main menu. The sampler tracks the pump counts in both the forward and reverse cycles with a resettable counter. When the counter reaches the default count of 1,000,000 or the count you have e ntere d, the sa mp le r d is p lay s a m ess ag e, “WA RN IN G ! REPLACE PUMP TUBE.
6712SR Refrigerated Sampler Section 8 Maintenance 8.4.3 Internal Battery Warning The internal lithium battery preserves the sampler’s memory when power is removed. For an explanation of how to replace the lithium battery, go to Section 8.8. Table 8-3 Checking and Replacing the Internal Battery 1 RUN PROGRAM VIEW REPORT OTHER FUNCTIONS 2 MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE HARDWARE Select INTERNAL BATTERY.
6712SR Refrigerated Sampler Section 8 Maintenance PLEASE WAIT! . . . ‘RAM’ PASSED TEST PURGING . . . PLEASE WAIT! . . . ‘ROM’ PASSED TEST •If RAM or ROM fails the test, the sampler will report that ‘RAM’ (or ’ROM’) FAILED TEST. Contact Teledyne Isco’s Customer Service for assistance. ABCDEFGHIJKLMNOPQRST ABCDEFGHIJKLMNOPQRST ABCDEFGHIJKLMNOPQRST ABCDEFGHIJKLMNOPQRST PURGING . . . ON/OFF RATIO = _.__ TEST DISTRIBUTOR YES NO FINDING ARM FLEXURE TEST PUMP? YES NO NOW AT BOTTLE __ PUMPING . . .
6712SR Refrigerated Sampler Section 8 Maintenance CAUTION Tests indicate that the pump produces sound levels in excess of 85db at one meter. Prolonged exposure to this sound level could result in hearing loss and requires the use of protective ear plugs. Distributor Test – The distributor test is provided for factory personnel to verify the distributor’s position as it rotates through the 24 positions. (screens 9 through 11).
6712SR Refrigerated Sampler Section 8 Maintenance 6. The blue collars on the tube help align the tube in the detector and pump. Align the tube by placing the collars in the grooves inside the liquid detector. 7. Close the detector’s cover and tighten the black knob. Close the pump strap and latch it. 8. Reset the pump-tube counter. (See example in table 8-2.) 9. Take a “dry” grab sample to test the tube installation.
6712SR Refrigerated Sampler Section 8 Maintenance Figure 8-1 Removing and Replacing the Pump Tube Figure 8-2 Placing the Pump Tube in the Liquid Detector 8-10
6712SR Refrigerated Sampler Section 8 Maintenance 8.6 Opening the Controller Case You will need to open the controller case to gain access to: • The desiccant bag that dries the controller’s interior • The battery that provides backup power to the controller CAUTION The circuit boards can be damaged from a discharge of static electricity. Always use an antistatic mat when opening the controller to protect the circuit boards. To open the case: 1.
6712SR Refrigerated Sampler Section 8 Maintenance control box exceeds 20%. As more moisture accumulates, the areas numbered “30” and “40” will turn light pink or white, indicating relative humidities of 30% and 40% respectively. If the 30% area of the humidity indicator turns light pink or white, open the control unit, inspect it for leaks, and replace the desiccant bag. CAUTION Do not attempt to reactivate the used desiccant.
6712SR Refrigerated Sampler Section 8 Maintenance Keypad Connector Module Connector Battery Liquid Detector Connector Connector Board Connector Pump Count Connector Figure 8-4 Main Circuit Board Connections 8.
6712SR Refrigerated Sampler Section 8 Maintenance Note Teledyne Isco recommends you replace the battery every five years, as soon as the battery warning appears. To ensure that the replacement meets Teledyne Isco’s specifications, use only Teledyne Isco replacement batteries. (See Appendix D, Replacement Parts.) To replace the internal battery: 1. Open the controller case. Separate the case bottom and top by removing the two distributor connectors. 2. Lift the desiccant box from the case. 3.
6712SR Refrigerated Sampler Section 8 Maintenance 8.9 Error Messages If the sampler detects an error that prevents it from taking a sample or continuing the program, it displays an error message.
6712SR Refrigerated Sampler Section 8 Maintenance 8.11 Servicing the Refrigerator Before servicing the refrigerator, disconnect it from power. This precaution will protect you from the possibility of electric shock when working with the refrigerator. WARNING Removing the front or back panel exposes electrical and mechanical hazards. Disconnect the power before performing any service activities. Troubleshooting and repair activities should be performed by a qualified refrigeration technician.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts See Running Programs on page -10. RUN PROGRAM VIEW REPORT OTHER FUNCTIONS Stop Key POWER USED: CURRENT: _._ amp-hr PREVIOUS: _._ amp-hr Note A See the charts on pages A-2 through A-19. Note A See View Report on page A-6. See Maintenance Screens on page A-7. MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE See Manual Functions Screens on page A-8.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts RUN PROGRAM VIEW REPORT OTHER FUNCTIONS SITE DESCRIPTION: " FAC TO RY " CHANGE? YES NO YES SITE: __________ ABCDEFGHIJKLMNOPQRST UVWXYZ -&"0123456789 BACK-UP DONE If STOP key pressed. SAVE CHANGES? YES NO NO DONE Programming screens shown are for a 730 Bubbler Flow Module. Screens for other modules may vary.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts Continued from Figure A-2. See Notes B & C. N O D E L AY T O S TA R T D E L AY E D S TA R T CLOCK TIME WAIT FOR PHONE CALL FIRST SAMPLE AFTER A _ _ _ M I N U T E D E L AY (1-999) See Note D.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts SITE: __________ ABCDEFGHIJKLMNOPQRST UVWXYZ -&"0123456789 BACK-UP DONE SITE DESCRIPTION: " FAC TO RY " If applicable UNITS SELECTED F L O W R AT E : c f s FLOW VOLUME: Mgal If STOP key pressed. SAVE CHANGES? YES NO Units Selected screens are for a 730 Bubbler Flow Module. Screens for other modules may vary.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts Continued from Figure A-4.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts RUN PROGRAM VIEW REPORT OTHER FUNCTIONS No valid options N O D ATA TO D I S P L AY SELECT DATA TO VIEW: SAMPLING REPORT MODULE FR-TEMP RAIN SDI-12 VIEW: D A I LY S U M M A R Y CURRENT READINGS pH MODULE VIEW: p H D ATA T E M P E R AT U R E D ATA Valid Options Will Be Shown; Screens Will Vary Depending on Devices Attached V I E W D ATA SYSTEM IDs CONFIGURE REPORT SAMPLER 6712 ID:1234567890 HARDWARE: B0 SOFTWARE: 1.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts RUN PROGRAM VIEW REPORT OTHER FUNCTIONS E N T E R T I M E A N D D AT E : HH:MM DD-MON-YY __:__ __-__-__ MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE ________ PUMP COUNTS, W A R N I N G AT _ _ _ _ _ _ _ . See Note A.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts RUN PROGRAM VIEW REPORT OTHER FUNCTIONS MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE GRAB SAMPLE SAMPLE VOLUME: ____ml (10-9990) P R E S S ↵ W H E N R E A DY ! if not known GRAB SAMPLE C A L I B R AT E V O L U M E O P E R AT E P U M P MOVE DISTRIBUTOR COMPACT PORTABLE STANDARD PORTABLE REFRIGERATOR SAMPLE VOLUME: ____ml (100-9990) C A L I B R AT E V O L U M E P R E S S ↵ W H E N R E A DY ! GO TO BOTTLE __ (1-max) VOLUME DELIVERED: _____ ml
6712SR Refrigerated Sampler Appendix A Menu Flowcharts "EXTENDED "EXTENDED "EXTENDED "EXTENDED RUN "EXTENDED 1" PROGRAM VIEW REPORT OTHER FUNCTIONS 1" 2" 3" 4" If STOP key pressed.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts Continued from Figure A-9. NUMBER OF BOTTLES: 1 2 4 8 12 24 BOTTLE VOLUME IS _____ ml (300-30000) SUCTION LINE LENGTH IS __ ft (3-99) AUTO SUCTION HEAD ENTER HEAD SUCTION HEAD OF __ ft (min-max) _ RINSE CYCLES (0-3) RETRY UP TO _ TIMES WHEN SAMPLING (0-3) O N E - PA R T P R O G R A M T W O - PA R T P R O G R A M See Figure A-11.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts Continued from Figure A-10. TIME BETWEEN SAMPLE EVENTS: __ HOURS, __ MINUTES U N I F O R M T I M E PA C E D F L O W PA C E D E V E N T PA C E D NONUNIFORM TIME PA C E D B Y : FLOW PULSES FLOW MODULE VOLUME FLOW BETWEEN SAMPLE EVENTS: ____ PULSES (1-9999) NONUNIFORM TIME: CLOCK TIMES I N T E RVA L S I N M I N U T E S R A N D O M I N T E RVA L S See Figure A-12: Event Conditions TA K E S A M P L E S AT : 1 . S TA R T T I M E 2. HH:MM 3.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts Continued from Figure A-12. All valid event options will be shown. EVENT: RAIN LEVEL FLOW pHx TEMPx CONDx TDSx DOx NONE I/O RAIN "RAIN" SET POINT: __.__ INCHES PER __HOURS, __MINUTES ENABLED WHEN: ABOVE SET POINT BELOW SET POINT Repeat until conditions are set or until DONE NO EVENT PARAMETERS CURRENTLY AVAILABLE OTHER OPTIONS ENABLED WHEN: ABOVE SET POINT BELOW SET POINT "LEVEL" RANGE: LOWER:_.__ft UPPER:_.__ft (0.001-30.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts Continued from Figure A-21. Continued from Figure A-11. 'A' Done and I/O are not options for I/O Conditions See Note A. See Note A. NONE ENABLE: I/O RAIN LEVEL FLOW TEMPx CONDx TDSx DOx NONE AND or OR ENABLE: RAIN All except NONE AND OR DONE ENABLE: RAIN AND I/O LEVEL FLOW TEMPx CONDx TDSx DOx DONE DONE 'A' DONE I/O RAIN "RAIN" SET POINT: __.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts RUN "EXTENDED 1" PROGRAM VIEW REPORT OTHER FUNCTIONS "EXTENDED "EXTENDED "EXTENDED "EXTENDED SELECT NEW PROGRAM CHANGE PROGRAM NAME NAME: __________ ABCDEFGHIJKLMNOPQRST UVWXYZ -&"0123456789 BACK-UP DONE NAME: __________ ABCDEFGHIJKLMNOPQRST UVWXYZ -&"0123456789 BACK-UP DONE SAVE CHANGES? YES NO PROGRAM NAME: "EXTENDED 1" SITE DESCRIPTIONS: " FAC TO RY " UNITS SELECTED: LENGTH: ft 1" 2" 3" 4" SAVE CHANGES? YES NO SELECT UNITS FOR FLOW RATE: c
6712SR Refrigerated Sampler Appendix A Menu Flowcharts Continued from Figure A-14.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts TIME BETWEEN SAMPLE EVENTS: __ HOURS, __ MINUTES Continued from Figure A-15. PA C I N G : TIME, EVERY __ HOURS, __ MINUTES U N I F O R M T I M E PA C E D F L O W PA C E D E V E N T PA C E D NONUNIFORM TIME PA C E D B Y : FLOW PULSES FLOW MODULE VOLUME FLOW BETWEEN SAMPLE EVENTS: ____ PULSES (1-9999) NONUNIFORM TIME: CLOCK TIMES I N T E RVA L S I N M I N U T E S R A N D O M I N T E RVA L S TA K E S A M P L E S AT : 1 . S TA R T T I M E 2. HH:MM 3.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts Continued from Figure A-16. Continued from Figure A-21. 'A' Done and I/O are not options for I/O Conditions See Note A. 'A' ENABLE: RAIN>0.11"/1:00 AND 0.500
6712SR Refrigerated Sampler Appendix A Menu Flowcharts RUN "ISCO 6712 " PROGRAM VIEW REPORT OTHER FUNCTIONS M E A S U R E M E N T I N T E RVA L 5 15 30 SECONDS 1 2 5 MINUTES S TO R AG E I N T E RVA L MAINTENANCE MANUAL FUNCTIONS S O F T WA R E O P T I O N S HARDWARE DUAL SAMPLER MODE? YES NO USE LIQUID DETECTOR? YES NO BOTTLE FULL DETECT? YES NO PROGRAMMING STYLE: NORMAL QUICK VIEW/CHANGE B A C K L I G H T A LWAY S O F F TIMED BACKLIGHT CONTINUOUS BACKLIGHT AT T H E B E G I N N I N G O F : INITIAL
6712SR Refrigerated Sampler Appendix A Menu Flowcharts RUN "ISCO 6712 " PROGRAM VIEW REPORT OTHER FUNCTIONS MAINTENANCE MANUAL FUNCTIONS S O F T WA R E O P T I O N S HARDWARE USE LIQUID DETECTOR? YES NO LIQUID DETECT ON QUICK VIEW/CHANGE PROGRAMMING STYLE: NORMAL QUICK VIEW/CHANGE TA K E M E A S U R E M E N T S EVERY __ MINUTES M E A S U R E M E N T I N T E RVA L 5 15 30 SECONDS 1 2 5 MINUTES S TO R AG E I N T E RVA L DUAL SAMPLER MODE? YES NO DUAL SAMPLER ON BTL FULL DECTECT OFF TIMED BACKLIGHT BOTTL
6712SR Refrigerated Sampler Appendix A Menu Flowcharts MAINTENANCE MANUAL FUNCTIONS SOFTWARE OPTIONS HARDWARE RUN "EXTENDED 1" PROGRAM VIEW REPORT OTHER FUNCTIONS D O YO U H AV E A RAIN GAUGE ATTACHED? YES NO NO YES RAIN GAUGE 0.01 inch TIP 0.1mm TIP If applicable NEW HARDWARE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE SCAN SDI-12 SONDES AT TURN-ON TIME? YES NO PLEASE WAIT! DONE YES PA R T I T I O N M E M O R Y W I L L L A S T _ _ DAY S If scan finds sonde, see Figure A-24. See Note A.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts RUN "ISCO 6712 " PROGRAM VIEW REPORT OTHER FUNCTIONS MAINTENANCE MANUAL FUNCTIONS SOFTWARE OPTIONS HARDWARE 0.01 inch TIP RAIN GAUGE D O YO U H AV E A R A I N G A U G E AT TA C H E D ? YES NO YES RAIN GAUGE 0.01 inch TIP 0.1mm TIP NO If applicable NEW HARDWARE SETUP-D O W N L O A D D ATA N O W O R L O S E A L L D ATA ! DONE See Figure A-24.
6712SR Refrigerated Sampler Appendix A Menu Flowcharts Notes See Figure A-20. A. Factory installed option. Requires B1 hardware or newer. B. Requires factory-installed Dialout Modem and B2 hardware or newer. If applicable See Note A ANALOG OUTPUT HARDWARE HAS NOT B E E N I N S TA L L E D ! ANALOG OUTPUT _: LEVEL VELOCITY FLOW NONE SETUP MANUAL CONTROL SET ANALOG OUTPUT _ TO __.__ mA ANALOG OUTPUT _: 0-TO-20 mA 4-TO-20 mA DONE ANALOG OUTPUT _: 0 m A AT _ . _ _ f p s 2 0 m A AT _ .
6712SR Refrigerated Sampler Appendix A Menu Flowcharts NEW HARDWARE SETUP D O W N L O A D D ATA N O W O R L O S E A L L D ATA ! DONE NO SDI-12 SONDE SETUP KNOWN SONDES SCAN SDI-12 SONDES PLEASE WAIT! LOOKING FOR SDI-12 SONDES NO SDI-12 SONDE AUTO SDI-12 SCAN ON _ _ P R E V I O U S LY CONFIGURED SONDES FOUND __ SONDE(S) FOUND If Sondes Found If No Sondes Found NEW HARDWARE SETUP D O W N L O A D D ATA N O W O R L O S E A L L D ATA ! DONE SCAN SDI-12 SONDES AT T U R N O N T I M E ? YES NO PA R T I T
6712SR Refrigerated Sampler Appendix A Menu Flowcharts A-24
6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets This appendix provides Material Safety Data sheets for the internal desiccant used in the 6712 sampler. Specific questions regarding the use and handling of these products should be directed to the manufacturer listed in the MSDS.
6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets 101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002 MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT® Packaged Desiccant SECTION I -- PRODUCT IDENTIFICATION Trade Name and Synonyms: Chemical Family: Formula: Silica Gel, Synthetic Amorphous Silica, Silicon, Dioxide Synthetic Amorphous Silica SiO2.
6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets 101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002 MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT® Packaged Desiccant SECTION IV -- FIRE EXPLOSION DATA Fire and Explosion Hazard - Negligible fire and explosion hazard when exposed to heat or flame by reaction with incompatible substances. Flash Point - Nonflammable. Firefighting Media - Dry chemical, water spray, or foam.
6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets 101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002 MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT® Packaged Desiccant NOTE TO PHYSICIAN: This product is a desiccant and generates heat as it adsorbs water. The used product can contain material of hazardous nature. Identify that material and treat accordingly.
6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets 101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002 MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT® Packaged Desiccant SECTION IX -- SPECIAL PRECAUTIONS Avoid breathing dust and prolonged contact with skin. Silica gel dust causes eye irritation and breathing dust may be harmful.
6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets Indicating Silica Gel Material Safety Data Sheet Identity (Trade Name as Used on Label) Manufacturer : MSDS Number* : MULTISORB TECHNOLOGIES, INC. M75 (formerly Multiform Desiccants, Inc.) Address: CAS Number* : 325 Harlem Road Buffalo, NY 14224 Phone Number (For Information): 716/824-8900 Emergency Phone 716/824-8900 Date Prepared: July 6, 2000 G.E.
6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets Page 2 Section 5 - Health Hazard Data PRIMARY ROUTES OF ENTRY HEALTH HAZARDS Inhalation Ingestion NTP CARCINOGEN LISTED IN Skin Absorption Not Hazardous IARC Monograph Acute May cause eye, skin and mucous membrane irritation. Chronic Prolonged inhalation may cause lung damage. OSHA Not Listed Signs and Symptoms Drying and irritation. of Exposure Medical Conditions Asthma.
6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets B-8
6712SR Refrigerated Sampler Appendix C General Safety Procedures The safety of the personnel who use the 671SR is a critical consideration. The following procedures, applicable to working in and around manholes and sewers, are those used by Black & Veatch, a respected consulting firm, and are published here by their permission. Field personnel must keep safety uppermost in their minds at all times. When working above ground, rules of common sense and safety prevail.
6712SR Refrigerated Sampler Appendix C General Safety Procedures C.1.4 Falling Objects Items placed near the manhole opening may fall and injure a worker in the manhole. All loose items should be kept away from the manhole opening. This applies to hand tools as well as stones, gravel and other objects. C.1.5 Sharp Edges Sharp edges of items in or near a manhole may cause cuts or bruises. C.1.
6712SR Refrigerated Sampler Appendix C General Safety Procedures A person should never enter a manhole unless he is wearing personal safety equipment, including a safety harness and a hard hat. Two persons should be stationed at the surface continuously while anyone is working inside a manhole, to lift him out if he is overcome or injured. One man cannot lift an unconscious man out of a manhole.
6712SR Refrigerated Sampler Appendix C General Safety Procedures C-4 C.4.5 Field Equipment The following equipment should be available for use: Blowers Gloves Breathing apparatus Hard Hats Coverall Harnesses First aid kits Manhole irons Emergency flashers Pick axes Flashlight Rain slickers Mirror Ropes Gas detectors Safety vests Gas masks Traffic cones Waders” C.4.6 Lethal Atmospheres in Sewers The following is an article written by Dr. Richard D.
6712SR Refrigerated Sampler Appendix C General Safety Procedures does not support life. The word has sometimes been misinterpreted as meaning suffocation, which is only one kind of asphyxiation. “In nearly all cases of death in sewers, the real killer is hydrogen sulfide. It is important that this fact be recognized. Many cities diligently test for explosive gases, which is very important, and they may measure the oxygen concentration which usually is unimportant, but they rarely measure H2S.
6712SR Refrigerated Sampler Appendix C General Safety Procedures “Be very careful to avoid high H2S concentrations, flammable atmospheres, and hazards of physical injuries. Remember that much H2S may be released by the stirring up of sludge in the bottom of a structure. Obey your senses in respect to irritating gases, such as chlorine (unconsciousness comes suddenly from breathing too much). Be cautious about strange odors. Do not determine percent oxygen in the air.
6712SR Refrigerated Sampler Appendix C General Safety Procedures Table C-1 Hazardous Gases (Continued) Formaldehyde CH2O Colorless, pungent suffocating odor. 1.07 Irritating to the nose. — 10 7.0 73.0 Near bottom. Incomplete combustion of organics. Common air pollutant, fungicide. Detectable odor. Gasoline C5H12 to C9H20 Volatile solvent. Colorless. Odor noticeable at 0.03%. Flammable. 3.0 to 4.0 Anesthetic effects when inhaled. Rapidly fatal at 2.4%. Dangerous for short exposure at 1.
6712SR Refrigerated Sampler Appendix C General Safety Procedures Table C-1 Hazardous Gases (Continued) Ozone O3 Irritant and poisonous. Strong electrical odor. Strong oxidizer. Colorless. At 1 ppm, strong sulfur-like odor. Sludge Gas —*** Mostly a simple asphyxiant. May be practically odorless, tasteless. Sulfur Dioxide SO2 Toluene 1.66 Max. naturally occurring level is 0.04 ppm. 0.05 ppm causes irritation of eyes and nose. 1 to 10 ppm causes headache, nausea; can cause coma.
6712SR Refrigerated Sampler Appendix D Replacement Parts The following appendix contains illustrated replacement parts listings, including part descriptions and order numbers.
6712SR Refrigerated Sampler Appendix D Replacement Parts D-2
6712SR Refrigerated Sampler Appendix D Replacement Parts D-3
6712SR Refrigerated Sampler Appendix D Replacement Parts Locknut, 10-32 Screw, 10-32 x 3/4 D-4
6712SR Refrigerated Sampler Appendix D Replacement Parts D-5
6712SR Refrigerated Sampler Appendix D Replacement Parts D-6
6712SR Refrigerated Sampler Appendix D Replacement Parts D-7
6712SR Refrigerated Sampler Appendix D Replacement Parts D-8
6712SR Refrigerated Sampler Appendix E Accessories List E.1 Order Information Prices available on request. Additional items appear in Appendix D, Replacement Parts. Many other items are available. To order any item, contact your sales representative or the factory. Note that the part name listed on your order acknowledgment and invoice may be different than the item name listed here. When examining these documents, use the part number for reference. Teledyne Isco, Inc. P.O.
6712SR Refrigerated Sampler Appendix E Accessories List (Includes bottle; configuration kit; one unlined lid; 2 lids and two discharge tubes). 1-Bottle Kit (9.4-liter glass bottle) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-039 (Includes bottle; configuration kit; 2 lids and two discharge tubes). 1-Bottle Kit (15.0-liter polyethylene bottle) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6712SR Refrigerated Sampler Appendix E Accessories List 9.4-liter (21/2-gallon) Glass Bottle with Two Teflon Lined Lids . . . . . . . . . . . . . . . . . . . . . . . . 68-2930-006 9.4-liter (21/2-gallon) Polyethylene Bottle with Two Unlined Lid . . . . . . . . . . . . . . . . . . . . . . . 68-2740-009 15.0-liter (4-gallon) Polyethylene Bottle with Two Unlined Lids . . . . . . . . . . . . . . . . . . . . . . . 68-2930-004 E.6 Pump Tubes, Suction Line, Strainers Pump Tube (Qty 5) . . . . . . . . . . . . . .
6712SR Refrigerated Sampler Appendix E Accessories List 4200T Modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . factory-installed Contact your sales representative or Teledyne Isco Customer Service. Note The modem is disabled when an interrogator cable is connected to the sampler’s interrogator port. It cannot receive incoming calls, and the alarm dialout will be rendered inoperative, while this cable is connected.
6712SR Refrigerated Sampler Appendix E Accessories List (This cable allows connection of a Hydrolab Quanta SDI-12 sonde and a teledyne Isco Rain Gauge.) SDI-12 Adapter Cable (6 foot length, no connector) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60-9004-341 (This is for use with a customer supplied connector to an SDI-12 device.) SDI-12 Y-Connect Cable (6 foot length, no connector). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6712SR Refrigerated Sampler Appendix E Accessories List E-6
6712SR Refrigerated Sampler Index A Accessories bottle racks E-2 bulk bottles E-2 bulk discharge tubing E-3 cables E-3 data collection accessories E-3 power sources and battery chargers E-4 pump tube E-3 refrigerator temperature sensor E-4 retaining rings E-2 strainers E-3 suction line E-3 Acknowledge dialout alarm 5-38 Alkaline battery battery warning 8-13 opening controller case 8-11 replacing 8-13 Analog output 5-37 B Backlighting 5-30 Bottle kits installing 2-4 Bottle Number Signal 5-31 Bottles-per-sa
6712SR Refrigerated Sampler Index L Line rinses 5-12 Liquid detector enable/disable 5-28 Locating base 2-bottle kit 2-6 M Maintenance 8-4 alkaline battery 8-13 checklist 8-8 cleaning guidelines 8-1 protocols for bottles and tubing 8-2 maintenance programming screens 8-4 diagnostics 8-6 pump counts for pump tube warning 8-8 pump tube alarm 8-8 reinitializing 8-8 opening controller case 8-11 pump tube 8-8 Manual functions 4-14 calibrate volume 4-14 grab samples 4-14 moving distributor 4-18 running pump manu
6712SR Refrigerated Sampler Index measuring 2-9 variable volume sampling 5-19 Sampler enable 4-9, 5-22 "dry period" option 5-26 master/slave sampling 5-29 repeatable enable 5-25 sample at enable or disable 5-25 start delay 5-26 start times 4-8 stay enabled 5-25 Samples per bottle 4-6, 5-16 Sampling cycle 2-8 Sampling programs running programs 4-10, 5-27 Sampling reports Collecting reports 4-19, 5-28 configuring 4-20 sampling results report 4-21 Viewing 4-19 Sampling results report see Sampling reports Sampl
6712SR Refrigerated Sampler Index Index-4
Warranty Teledyne Isco One Year Limited Factory Service Warranty * Teledyne Isco warrants covered products against failure due to faulty parts or workmanship for a period of one year (365 days) from their shipping date, or from the date of installation by an authorized Teledyne Isco Service Engineer, as may be appropriate. During the warranty period, repairs, replacements, and labor shall be provided at no charge.