INSTRUCTION MANUAL MODEL 465M OZONE MONITOR TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TELEDYNE API) 9480 Carroll Park Drive San Diego, CA 92121-5201 TOLL-FREE: 800-324-5190 TEL: 858-657-9800 FAX: 858-657-9816 EMAIL: api-sales@teledyne.com WEB SITE: http://www.teledyne-api.
THIS PAGE IS INTENTIONALLY LEFT BLANK
SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a safety alert symbol and placed throughout this manual and inside the instrument. The symbols with messages are defined as follows: WARNING: Electrical Shock Hazard HAZARD: Strong oxidizer GENERAL WARNING/CAUTION: Read the accompanying message for specific information.
CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement ces consignes. Chaque consigne de sécurité est représentée par un pictogramme d’alerte de sécurité; ces pictogrammes se retrouvent dans ce manuel et à l’intérieur des instruments.
WARRANTY WARRANTY POLICY (02024F) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instruments, Inc., provides that: Prior to shipment, TAPI equipment is thoroughly inspected and tested. Should equipment failure occur, TAPI assures its customers that prompt service and support will be available.
THIS PAGE IS INTENTIONALLY LEFT BLANK P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page iv
TABLE OF CONTENTS 1. INTRODUCTION..........................................................................................................9 1.1. PREFACE ..............................................................................................................................................9 2. SPECIFICATIONS AND AGENCY APPROVALS ................................................11 2.1. SPECIFICATIONS........................................................................................................
. DIGITAL COMMUNICATIONS ..............................................................................43 8.1. DIGITAL COMMUNICATIONS OVERVIEW ............................................................................................43 8.2. CHANGING COM PORT HARDWARE CONFIGURATION .......................................................................43 8.3. RS-232 CONFIGURATION ...................................................................................................................45 8.4.
FIGURES Figure 4-1: M465M Mounting Hole Locations and Dimensions.................................................................16 Figure 4-2. NEMA AC Power Connection to Monitor ................................................................................18 Figure 4-3: Pneumatic Connections, Rack Mount Configuration ................................................................19 Figure 4-4: Pneumatic Connections, NEMA Configuration ....................................................................
TABLES Table 4-1: M465M Ventilation Clearances, NEMA Version........................................................................15 Table 4-2: M465M Ventilation Clearances, Rack Mount Version................................................................16 Table 4-3: Status Output Definitions............................................................................................................24 Table 4-4: Relay Functions..............................................................................
1. INTRODUCTION 1.1. Preface Teledyne API is pleased that you have purchased the M465M. We offer a full one year warranty (see Section 2.2) and we at Teledyne API will be pleased to provide you with any support required so that you may utilize our equipment to the fullest extent. We hope you will not experience any problems with the Teledyne API Model 465M (M465M) but if you do, the built-in tests and diagnostics should allow you to quickly and easily find the problem.
THIS PAGE IS INTENTIONALLY LEFT BLANK P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 10
2. SPECIFICATIONS AND AGENCY APPROVALS 2.1. Specifications Measuring principle UV absorption (Beer Lambert Law) Ranges User selectable to any full scale range from: 0-1000 PPM to 0-10,000 PPM 0-5 g/Nm3 to 0-50 g/Nm3 0-1 wt% to 0-5 wt% Measurement Units PPM, wt%, g/Nm3 Accuracy Precision/Repeatability Display Resolution Response Time (95%) 1% of Full Scale 0.5% of Full Scale 1 PPM, 0.001 wt%, 0.01 g/Nm3 <45 seconds to 95% Pressure, Temperature (NTP = 273.15K, 760 mmHg) 11.0 – 16.0 psia 0.
2.2. Approvals This section presents Safety and Electromagnetic Compatibility (EMC) compliance approvals for the Model 465M monitor. 2.2.1. Safety IEC 61010-1:2001, Safety requirements for electrical equipment for measurement, control, and laboratory use. CE: 2006/95/EC, Low-Voltage Directive North American: cNEMKO (Canada): CAN/CSA-C22.2 No. 61010-1-04 NEMKO-CCL (US): UL No. 61010-1 (2nd Edition) 2.2.2.
3. PRINCIPLE OF OPERATION The detection of ozone molecules is based on absorption of 254 nm UV light due to an internal electronic resonance of the O3 molecule. The M465M uses a mercury lamp constructed so that a large majority of the light emitted is at the 254nm wavelength. Light from the lamp shines through an absorption cell through which the sample gas is passed.
length determines how many molecules are present in the column of gas in the absorption tube. The intensity of light is converted into a voltage by the detector/preamp module. The voltage is converted into a number by a high resolution A/D (analog-to-digital) converter. The digitized signal, along with the other variables, is used by the CPU to compute the concentration using the above formula.
4. GETTING STARTED The M465M is shipped with the following standard equipment: 1. Power cord (Rack mount version only) 2. Instruction manual. Upon receiving the M465M please do the following: 1. Verify that there is no apparent shipping damage. (If damage has occurred please advise shipper first, then Teledyne API.) 4.1. Mechanical Installation (for NEMA 4X Enclosure) Mount the enclosure securely to a vertical surface. Figure 4-1 shows the locations of the four mounting holes.
Figure 4-1: M465M Mounting Hole Locations and Dimensions 4.2. Mechanical Installation (19” Rack Version) The Rack Mount version of the M465M was designed to be operated as a bench-top monitor or to be mounted in a standard 19” RETMA instrumentation rack. For rack installations, the four mounting feet should be removed from the bottom of the monitor. 4.2.1. Ventilation Clearance When installing the M465M be sure to leave sufficient ventilation clearance.
4.3. AC Power Connection (for Monitors in NEMA 4X Enclosures) CAUTION Disconnect power to the AC mains before making or removing any electrical connections to the M465M. CAUTION A proper earth ground connection must be made to the receptacle labeled “Earth Ground” on the 3 pin AC connector. Failure to do so may result in a shock hazard and malfunction of the monitor 1.
3. For the NEMA configuration, the electrical connection must be hard-wired to the 3-pin connector on the mainboard labeled J23. AC power connection to the monitor should be made with 12-14 AWG stranded copper wire, connected to the monitor as follows: a. Earth Ground (green): Connect the earth ground wire to the screw terminal lug on the chassis as shown in the Figure below. Do not connect the earth ground wire directly to the PCB connector on the mainboard PCB. b.
should light with a sequence of messages, including the instrument serial number and software revision. (When the instrument first powers up, the display will show ‘xxxx’, indicating that the instrument is in start-up mode, waiting for the UV detector readings to stabilize). After the start-up period, the instrument will begin displaying the ozone concentration and the optional stream selector will begin operating, if installed. 4.4.
Figure 4-4: Pneumatic Connections, NEMA Configuration 4.4.1. Ozone Inlet Connection Sample gas connections are made to the ¼” stainless steel compression fitting on the rear panel. (See Figure 4-3 and Figure 4-4) 1/4" O.D. FEP (Teflon) or Stainless Steel tubing is recommended to connect the sample source to the monitor. Any fittings used in the sample lines should be constructed of Stainless Steel or Teflon.
4.4.2. Exhaust Connection Connect a ¼” OD vent line to the “Exhaust” fitting on the rear panel. This line must be vented outdoors or to an appropriate discharge system capable of handling ozone. Local regulations regarding the discharge of ozone to the atmosphere must be observed. Note that the monitor may be equipped with an internal ozone scrubber for removal of ozone from the exhaust stream. For safety reasons, the exhaust port must still be vented appropriately as described above.
Figure 4-6: NEMA Electrical I/O Connections P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 22
4.5. Electrical I/O Connections 4.5.1. Location of I/O Connectors in the NEMA Configuration In the NEMA configuration of the M465M, the I/O connectors are located internally in the instrument. See Figure 4-6 for their location. Connection to these terminals is usually made via a conduit connection to the NEMA enclosure. A standard conduit penetration may be added to one of the walls of the enclosure for this purpose, or the existing penetration may be used.
The analog output is automatically scaled to the range of the instrument, i.e. if the monitor range is 1000 PPB, then 5V would correspond to 1000 PPB and 0V would correspond to 0 PPB. See Section 7.2.5 for information on changing the range of the monitor. 4.5.3. Status Outputs There are four instrument status outputs located at the “General I/O” connector. These status outputs mirror the state of the four diagnostic LED’s on the front panel.
4.5.4. Hi-Current Relay Outputs Three form C relay outputs are provided on the rear panel on the nine pin connector (See Figure 4-5). The relays are labeled “Relay 1,” “Relay 2,” and “Relay 3.” The relays are dry contact type form C (SPDT) relays with Normally Open (N.O.), Normally Closed (N.C.) and Common (COM) contacts. The relays are capable of driving loads of up to 250VAC, 5A. For maximum contact life, the relays should only be used to drive resistive loads.
With the instrument configured for RS485, connection can be made to the DB-9 connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O” connector. See Section 8.2 for additional details. 4.5.5.3. Ethernet Connection With the instrument configured for Ethernet, connection should be made to the RJ-45 port labeled “Ethernet.” Connection can be made using a standard Category 5 or better network patch cable.
5. FREQUENTLY ASKED QUESTIONS 1. Q: How long does the UV Lamp last? A: The typical lifetime is about 2-3 years. The lamp output will steadily decrease until lamp adjustment can no longer be performed. The "Check Lamp" warning LED will tell the operator when this adjustment should be attempted. When this adjustment can no longer be performed, the lamp should be replaced. See Chapter 9 for more information.
THIS PAGE IS INTENTIONALLY LEFT BLANK P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 28
6. OPTIONAL HARDWARE AND SOFTWARE 6.1. Rack Mount Slides Rack mount slides can be attached to the sides of the Rack Mount version of the M465M, allowing it to slide completely out of a RETMA rack, facilitating maintenance activities. 6.2. Low Flow Alarm The M465M can be fitted with an internal electronic flow-switch that enables the instrument to detect gas flow and provide a Pneumatic Error if the flow is outside it’s normal limits.
THIS PAGE IS INTENTIONALLY LEFT BLANK P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 30
7. OPERATION 7.1. Front Panel Overview The M465M front panel provides a number of functions in addition to displaying the current ozone concentration being measured Figure 7-1 below shows the display and keypad portion of the front panel. Figure 7-1: Front Panel Display 7.1.1. Ozone Concentration Display The 4-character LED display in the center of the front panel normally shows the current ozone concentration being measured by the M465M.
7.1.3. Concentration Alarm Status/Reset The M465M front panel displays the concentration alarm status. An LED labeled ‘ALARM STATUS’ shows the alarm status as follows: Table 7-1: Alarm LED’s Alarm Status LED Color Alarm State Green No alarm active Amber ‘HI’ alarm active Red ‘HI-HI’ alarm active If the alarms are configured for latching operation, then the RESET button can be used to attempt to reset all alarms.
7.2. Front Panel Menus The M465M has an interactive menu structure that can be operated from the front panel. From within this menu structure, the operator can view real-time parameters such as temperature and pressure, view configuration information, and edit setup parameters such as alarm limits. See Figure 7-2 and Figure 7-3 shows the menu structure as well as the buttons that are used to navigate it. Starting with the display in Concentration mode, press CFG to enter Menu mode.
Figure 7-2: Front Panel Menu Diagram P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 34
Figure 7-3: Front Panel Menu Diagram P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 35
7.2.3. DIAG Menu The DIAG menu (See Figure 7-2 and Figure 7-3) contains functions that are useful for testing and configuring external equipment that may be connected to the rear panel I/O of the instrument. 7.2.3.1. Analog Step-Test When placed in this mode, the instrument will automatically step the analog output through 5 points from 0 to 100 % and display the current value on the front panel.
Table 7-4: Signal I/O List Signal Name Rear Panel Label Function AUX 1 IN CTRL IN - 1 UNDEFINED AUX 2 IN CTRL IN - 1 UNDEFINED RELAY 1 RELAY 1 SENSOR OK RELAY 2 RELAY 2 HI ALARM RELAY 3 RELAY 3 HI-HI ALARM STAT OUT 1 STATUS OUT – 1 SENSOR OK STAT OUT 2 STATUS OUT – 2 INVALID READING STAT OUT 3 STATUS OUT – 3 CHECK LAMP STAT OUT 4 STATUS OUT – 4 PNEUMATIC ERROR STAT AUX 1 AUX OUT – 1 UNDEFINED STAT AUX 2 AUX OUT – 2 UNDEFINED 7.2.3.3.
NOTE While measuring zero air, a certain amount of noise or “dithering” of the concentration about the zero point will occur and is normal. This noise is typically 1-3 PPM in magnitude. 7.2.3.5. Span Calibration The Span Calibration also allows the instrument to calibrate its internal ozone offset factor. The Span Calibration however, is done with some controlled concentration gas connected to the “Sample In” port of the M465M.
Figure 7-4: Span Cal Menu 7.2.4. VIEW Menu The VIEW menu allows the operator to view various measurement parameters in realtime. This can be useful for diagnosing various instrument or system problems. NOTE These values are updated in real-time as they are repeatedly scrolled on the display. Table 7-5: VIEW Menu Parameters Parameter Meaning Units Normal Range* MEAS UV detector reading, measure cycle mV 250 – 1230 UV detector reading, reference cycle mV 250 – 1230 Sample pressure psia 9.
7.2.5. SETUP-VARS Menu The VARS menu allows viewing and editing of various global setup variables that effect how the M465M operates. These variables are stored in the instrument’s non-volatile memory. Table 7-6 lists these variables and their function. Table 7-6: VARS List VAR ALARM MODE ALARM DELAY BC FILT SIZE FLOW SWITCH ENABLE COMM ADDRESS UNIT SCROLL DELAY AZERO ENABLE AZERO INTERVAL GAS CARRIER WEIGHT ANALOG RANGE Meaning Controls the latching behavior of the concentration alarms.
7.2.6. SETUP-ALARM Menu The ALARM menu allows the operator to view and edit properties of the concentration alarms. The table below summarizes the alarm configuration settings.
THIS PAGE IS INTENTIONALLY LEFT BLANK P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 42
8. DIGITAL COMMUNICATIONS 8.1. Digital Communications Overview The M465M comes equipped with a digital communications (Com) port that can be connected to a computer or digital data acquisition system. This COM port uses the standardized MODBUS protocol and can be configured for RS232, RS485, or Ethernet (10 Mbit,) allowing for connections to a wide variety of devices and networks. See Section 8.2 for more information on configuring the COM port.
6. For RS-232 operation, set SW2 to either DTE or DCE (See Section 8.3) For RS485 operation, set SW2 to DCE. 7. Re-install the instrument cover. 8. Re-connect power to the instrument. 9. The software portion of the COM port configuration will now automatically be completed after the next instrument boot-up.
8.3. RS-232 Configuration The RS-232 configuration is usually used when making a one-to-one connection between the instrument and a single computer or PLC. The communications protocol used for the RS-232 configuration is MODBUS RTU. For details on the MODBUS RTU specification, please see http://www.modbus.org/. With the instrument configured for RS-232, connection can be made to the DB-9 connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O” connector.
8.4. RS-485 Configuration RS-485 is commonly used for factory automation busses with a computer or PLC acting as a master. The communications protocol used for the RS-485 configuration is MODBUS RTU. For details on the MODBUS RTU specification, please see http://www.modbus.org/. With the instrument configured for RS-485, connection can be made to the DB-9 connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O” connector.
8.5. Ethernet Configuration Ethernet is commonly used for factory automation busses with a computer or PLC acting as a master. The communications protocol used for the Ethernet configuration is MODBUS TCP/IP. For details on the MODBUS TCP/IP specification, please see http://www.modbus.org/. A Windows software application, DeviceInstaller, is available for configuring the Ethernet module used in the M465M. This application is available for download here: http://www.lantronix.
8.5.3. Determining a Dynamically Assigned IP Address (DHCP Mode) The DeviceInstaller software application (See Section 8.5) can be used to search a network for instruments and determine the IP Address assigned by a DHCP server. 1. Click Start->Programs->Lantronix DeviceInstaller->DeviceInstaller. If your PC has more than one network adapter, a message displays. Select an adapter and click OK. 2. Click on the Search icon. After a moment a list of instruments on the network will be displayed.
5. Enter the IP address. The Subnet mask displays automatically based on the IP address; if desired, you may change it. On a local network, you can leave the Default gateway blank (all zeros). Click Next. 6. Click the Assign button and wait several seconds until a confirmation message displays. Click Finish. 7. Select the device from the main window list and select Ping from the Tools menu. The Ping Device dialog box shows the IP address of the selected unit. 8. From the Tools menu, click the Ping button.
8.6. MODBUS Register Maps 8.6.1. Notes on MODBUS registers Concentration values in the MODBUS registers are always expressed in units of PPB (parts-per-billion,) regardless of the concentration units setting of the monitor. 8.6.2.
9. MAINTENANCE CAUTION RISK OF ELECTRICAL SHOCK. THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY! 9.1. Maintenance Schedule Table 9-1 below outlines the suggested maintenance procedures and intervals for ensuring the M465M continues to operate accurately and reliably. Table 9-1: Maintenance Schedule Maintenance Item Recommended Interval Section Replace particulate filter element 6 months*1 9.
9.2. Instrument Layout Figure 9-1 and Figure 9-2 shows the internal layout of the M465M. These figures will be referenced in the procedures that follow. Note the caution areas where high voltage (line voltage) may be present when power is connected to the instrument.
Figure 9-2: Instrument Layout, NEMA Configuration P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 53
9.3. Replacing Internal Particulate Filter 1. Disconnect power from the M465M. 2. Loosen the six screws from the top cover (Rack Mount Configuration, see Figure 9-1) or open front panel (NEMA Configuration, see Figure 9-2.) 3. Remove the six screws and top cover from the instrument. 4. Locate the particulate housing filter on the rear panel (Rack Mount Configuration, see Figure 9-1) or the bottom panel (NEMA Configuration, see Figure 9-2.) 5. Loosen the four screws on the sample filter body (see Figure 9-3.
Figure 9-3: Internal Particulate Filter Replacement P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 55
9.4. UV Lamp Adjustment 1. Instrument should be running and warmed up for at least 20 minutes. 2. With instrument running, remove the six screws and the top cover (Rack Mount Configuration) or open front panel (NEMA Configuration.) 3. Locate the Reference UV Detector adjustment pot, PR3, on the Mainboard PCA (see Figure 9-1 or Figure 9-2.) 4. Navigate the front panel menu to VIEW menu and scroll to REF display and press ENT. At this point there should be a scrolling display similar to “REF = XXXX MV.
9. Reconnect power to the instrument and turn on power switch. Let instrument warm up for at least 20 minutes. 10. Perform UV lamp adjustment procedure per Section 9.4. 9.6. Sample Pump Replacement 1. Disconnect power from the M465M. 2. Remove the six screws and the top cover (Rack Mount Configuration) or open front panel (NEMA Configuration.) 3. Locate sample pump assembly (See Figure 9-1 or Figure 9-2.) 4. Cut off the clamps holding the Tygon tubing to the pump nipples and remove tubing.
these should be removed. Use the new valve as a reference to determine which two screws need to be removed. 8. Note the orientation of the valve body on the manifold so that the new valve can be installed in the same orientation. 9. Remove valve body from sensor manifold. 10. Clean any residue or dirt off the surface of the manifold using a lint-free cloth and distilled or DI water. 11. Install the new valve by reversing steps 1-7. 9.8.
SW1 C1 VCC J2 R67 C4 PIC18F8720 C5 3.3V TP1 C2 RS-485/ Analog Out Expansion J3 R2 C3 FLOW R11 C14 C11 U7 C20 + C27 C30 SRAM128KX8-SOP R24 TP8 U13 C41 U16 + C34 D4 M41T0 C36 R35 VR3 VR5 C43 R37 R38 R40 U24 TP14 C54 U27 C59 C63 C60 REF R51 TP18 R53 TP21 R43 C64 C65 R34 I1 TP10 R48 -9V TP16 +12VL J12 D10 J14 R50 SCL C52 U25 TP15 J15 COM R65 TP17 TP20 J13 4-20mA OUTPUT N.O. K3 N.C. J9 J10 JP4 N.C.
9.10. Boot Loader Operation The main function of a boot loader is to load or update the firmware to the microcontroller on the Printed Circuit Board (PCB) of an instrument. There are two types of boot loader setup: hardware and software, described in the following sections. NOTE When an external MODBUS host device is attempting to communicate with the instrument over the serial or Ethernet port while the instrument is starting up, the instrument may halt.
4. Select the COM port that the instrument is connected to. 5. Select the HEX file for the appropriate firmware. 6. Refer to Section 9.10.2.1 to upload the firmware or to Section 9.10.2.2 to verify the firmware. 9.10.2.1. Upload Firmware 1. Press “Write”. The screen will display, “Wait for boot loader”. 2. When the screen displays, “Power cycle instrument now”, power up the instrument. NOTE Search for boot loader will time out in 20 seconds.
NOTE Search for boot loader will time out in 20 seconds. If the instrument is not powered up in 20 seconds, the screen will display, “No boot loader detected.” Go back to step 1 to verify the firmware. The firmware loader software should display a message “Boot loader found, verifying…”. If this message is not displayed, power down the instrument and receck the COM port configuration and cable (see Section 9.10.1) and retry. 3.
10. TROUBLESHOOTING CAUTION RISK OF ELECTRICAL SHOCK. THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY! 10.1. Reference Drawings The drawings contained in this section are for general reference and may be useful when performing certain troubleshooting activities. 10.1.1. Pneumatic Diagram Figure 10-1 and Figure 10-2 below is a pneumatic diagram that can be referenced when performing troubleshooting on the monitor.
O3 Gas In Filter Water Drop-Out Filter Dryer Flow-Meter Azero Valve O3 Measurement Cell O3 Destruct Flow Switch (Optional) Exhaust Out Pump Flow Control Orifice Figure 10-2: Pneumatic Diagram (NEMA Version with Sample Conditioner) 10.1.2. Interconnect Diagram Figure 10-3 provides details on the electrical connections between the various electronic modules in the M465M.
RACK-MOUNT VERSION ONLY CN0000073 IEC TYPE POWER ENTRY MODULE SW0000051 POWER SWITCH/CIRCUIT BREAKER 05072-0100 MAINBOARD PCA J19 FAN (REAR PANEL) J23 AC IN AC/DC Power Supply J15 FLOW SWITCH (OPTIONAL) J24 “ANALOG IN EXPANSION” 03484-XXXX SENSOR ASSY SW0000062 PRESSURE SENSOR J7 J17 04511-0100 HEATER/ TEMP SENSOR PCA J20 02571-0100 MEAS DETECTOR PCA J16 02571-0100 REF DETECTOR PCA J8 03742 UV LAMP ASSY J18 05313-0100 DISPLAY PCA J4 J2 +15V I2C N/C J14 “MODULE +15” J4 J2 04702 PUMP
10.2. Troubleshooting Using Front Panel Status LED’s or Status Outputs The M465M has 4 front-panel status LED’s that reflect the current operating status of the monitor, and indicate fault conditions. There are also four relay Status Outputs on the rear panel (or internally in the NEMA configuration) that also reflect the state of these status LED’s. See Section 4.5.3 for more information on the Status Outputs.
10.2.1. Sensor OK The Sensor OK LED indicates the status of the O3 sensor module in the monitor. The normal state of this LED (or Status Output) is On. If this LED remains off after the normal warm-up period, then a failure has occurred and the monitor should be removed from service and repaired. The most common cause of this warning is a failure of the UV Lamp. A UV Lamp Adjustment (See Section 9.4) should be attempted.
10.2.4.1. Pressure Too High The monitor inlet is being pressurized. The monitor inlet should be allowed to sample gas at ambient pressure. Disconnect all sample lines from the monitor and read the pressure again. If the value then drops to within the acceptable range, then one of the sample lines is at elevated pressure. If the inlet is being pressurized, then the design of the sample system should be reviewed and corrected. 10.2.4.2.
11. CALIBRATION CAUTION RISK OF ELECTRICAL SHOCK. THE OPERATIONS OUTLINED IN THE FOLLOWING SECTIONS OF THIS CHAPTER ARE TO BE PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY! 11.1. Factory Calibration The M465M is calibrated to internal T-API calibration standards to shipment. A calibration certificate for your instrument can be purchased from Teledyne API if required. Teledyne API also recommends that the M465M be re-calibrated once a year.
THIS PAGE IS INTENTIONALLY LEFT BLANK P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 70
12. TECHNICAL ASSISTANCE If this manual does not solve your problems, technical assistance may be obtained from: Teledyne API, Customer Service, 9480 Carroll Park Drive San Diego, California 92121-5201USA Toll-free Phone: 800-324-5190 Phone: 858-657-9800 Fax: 858-657-9816 Email: api-sales@teledyne.com Website: http://www.Teledyne-API.
THIS PAGE IS INTENTIONALLY LEFT BLANK P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 72
13. M465M SPARE PARTS 13.1.
13.2.