4220 Submerged Probe Flow Meter Installation and Operation Guide Part #60-3223-130 of Assembly #60-3224-052 Copyright © 2001. All rights reserved, Teledyne Isco, Inc.
Foreword This instruction manual is designed to help you gain a thorough understanding of the operation of the equipment. Teledyne Isco recommends that you read this manual completely before placing the equipment in service. Although Teledyne Isco designs reliability into all equipment, there is always the possibility of a malfunction. This manual may help in diagnosing and repairing the malfunction. If the problem persists, call or e-mail the Teledyne Isco Technical Service Department for assistance.
4220 Flow Meter Safety 4220 Flow Meter Safety General Warnings Before installing, operating, or maintaining this equipment, it is imperative that all hazards and preventive measures are fully understood. While specific hazards may vary according to location and application, take heed in the following general warnings: WARNING Liquids associated with this instrument may be classified as carcinogenic, biohazard, flammable, or radioactive.
4220 Flow Meter Safety DANGER DANGER – limited to the most extreme situations to identify an imminent hazard, which if not avoided, will result in death or serious injury. Hazard Symbols The equipment and this manual use symbols used to warn of hazards. The symbols are explained below. Hazard Symbols Warnings and Cautions The exclamation point within the triangle is a warning sign alerting you of important instructions in the instrument’s technical reference manual.
4220 Flow Meter Safety IMPORTANT – PLEASE READ WARNING The installation and use of this product may require you to work in locations where you could be seriously injured or even killed. Take whatever precautions are necessary to ensure your safety before entering the installation. Never work alone or unsupervised. Install and operate this product in accordance with all applicable safety and health regulations, as well as any appropriate local ordinances. This product is often installed in confined spaces.
4220 Flow Meter Safety
4220 Flow Meter Table of Contents Section 1 Introduction 1.1 Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 1.2 Compatible Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 1.3 Operating Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 1.3.1 Submerged Probe Operation . . . . . . . . . . . . . . . . . . . .
3.3 3.4 3.5 3.6 3.7 3.2.1 Low Power Indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3 3.2.2 Isco Sampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3 3.2.3 Isco Nickel-Cadmium Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4 3.2.4 Attaching the Nickel-Cadmium Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4 3.2.5 Isco Lead-Acid Battery . . . . . .
5.9.1 pH Probe Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.9.2 pH Probe Installation Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.9.3 Storage and Maintenance of pH Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.10 The Dissolved Oxygen (D.O.) Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.10.1 How the D.O. Probe Works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
C.3 C.4 C.5 C.6 C.7 C.8 C.9 Parameter to Adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4 Reset Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5 Sampler Pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5 Sampler Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
List of Tables 1-1 1-2 1-3 1-4 1-5 2-1 4-1 5-1 5-2 5-3 5-4 5-5 6-1 A-1 D-1 Controls, Indicators, and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7 Technical Specifications for the 4220 Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . 1-8 Technical Specifications for the 3222 Submerged Probe . . . . . . . . . . . . . . . . . . . . 1-9 Chart Longevity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
vi
4220 Flow Meter Section 1 Introduction This section provides a general introduction to the 4220 Submerged Probe Flow Meter. It includes a description, an explanation of how the unit operates, and technical specifications. Manual Organization This manual provides the information necessary to operate, maintain, and perform minor service on the 4220.
4220 Flow Meter Section 1 Introduction • 674 Tipping Bucket Rain Gauge • High Low Alarm Relay Box Non-Isco Products • IBM Personal Computer or compatible clone with Isco Flowlink software • Laptop Computer with Isco software • YSI 600 Multi-Parameter Sonde Optional Accessories • 4200T Modem - Speech-capable, with connector and cable • D.O. (dissolved oxygen) and pH parameter probes • Submerged probe Extension Cables (Vented), in lengths of 25 and 50 feet.
4220 Flow Meter Section 1 Introduction 1.3.1 Submerged Probe Operation The submerged probe is mounted beneath the surface of the flow stream and measures liquid level by responding to changes in hydrostatic pressure against the probe's sensing element. The probe contains an IC (integrated circuit) differential pressure transducer. This transducer provides an output signal that changes proportionally to the pressure placed against it by the flow stream.
4220 Flow Meter Section 1 Introduction Figure 1-1 Submerged Probe Level Sensor with Standard Tip 1.3.3 Two Probes Available Teledyne Isco offers two different submerged probe level sensors for use with the 4220. The standard unit is intended for operation in depths from 1 inch to 10 feet maximum. The other unit is capable of operation to depths as great as 30 feet. The difference in the probes is in the pressure transducer used inside. The two are not interchangeable.
4220 Flow Meter Section 1 Introduction Figure 1-2 4220 Controls and Indicators 1-5
4220 Flow Meter Section 1 Introduction Pin Function Power Sampler A = Channel 1 (-) Red B = Channel 3 (+) Green C = Channel 1 (+) White D = Channel 2 (-) Brown E = Channel 3 (-) Black F = Channel 2 (+) Blue Ground 12V A B C D E F 12V Ground Flow Pulse Output Bottle Number Input Event Mark Input Inhibit/Enable Output Rain Gauge A B C D 12V Ground SDO Rain Gauge Input Interrogator A B C D E F 12V Ground Sense Line SDO SDI Analog Output Pulse Modem (or Analog Output) B Tip C Ring A B C D E F
4220 Flow Meter Section 1 Introduction Table 1-1 Controls, Indicators, and Connectors Controls Settings Function ON/OFF On - Off Turns the flow meter on and off. Internal memory is protected with a standby battery. See Section 2. Keypad Momentary Switches 24-key, four column matrix. Programs flow meter through a series of keystrokes prompted by messages on the display. Certain keys perform specific functions (printing reports or entering program choices into memory). Arrow keys move through menus.
4220 Flow Meter Section 1 Introduction 1.6 Technical Specifications The technical specifications for the 4220 Flow Meter are found in Table 1-2. The anticipated longevity for a roll of paper used in the internal plotter is shown for various chart speeds in Table 1-4. Battery life expectancy is found in Table 1-5. Table 1-2 Technical Specifications for the 4220 Flow Meter Physical Specifications Size 16" high × 11 1⁄2" wide × 10 1⁄2" deep (40.6 × 29.2 × 26.7 cm) (without power source attached).
4220 Flow Meter Section 1 Introduction Table 1-2 Technical Specifications for the 4220 Flow Meter (Continued) Flowlink Data Storage and Retrieval System Memory Partitions Maximum of 6 user-defined memory partitions for level or event storage. Data Storage Rate of data storage user-selected in 1, 2, 5, 10, 15, 30, 60, or 120 minute intervals. Baud Rates Serial connection - 300, 1200, 2400, 4800, or 9600 baud. Serial connection with the optional internal modem - 2400 baud.
4220 Flow Meter Section 1 Introduction Table 1-3 Technical Specifications for the 3222 Submerged Probe (Continued) Compensated Temperature Range 32° to 122°F (0° to 50° C). Temperature Error (over compensated temperature range) 0.1 to 4.0 ft. (0.03 to 1.2 m) ±0.005 ft. per degree F 4.0 to 10.0 ft. (1.2 to 3.1 m) ±0.007 ft. per degree F Table 1-4 Chart Longevity Chart Speed, Inches/Hour Time to Empty Roll 4 195 Hours (81/2 Days) 2 161/4 Days 1 321/2 Days 0.
4220 Flow Meter Section 1 Introduction Battery capacity is expressed in ampere-hours. The battery manufacturer provides this information for each battery. This value is the product of a load current times an arbitrary time period, ten hours for nickel-cadmium batteries, and twenty hours for lead-acid types. The terminal voltage of the battery at the end of this time period is the discharged cell voltage, 10 volts for nickel-cadmium and 10.5 volts for lead-acid types.
4220 Flow Meter Section 1 Introduction nickel-cadmium batteries show fairly rapid rates of self-discharge. A battery that is fully charged and then placed in storage will lose some capacity each day. In a week, this could easily be 5% or more. When using lead-acid batteries, you must be careful to avoid complete discharge, as this may cause cell reversal, which will ruin the battery.
4220 Flow Meter Section 1 Introduction Fluke® 87 (or other currentaveraging meter) FLUKE 87 TRUE RMS MULTIMETER 00I5 Flow Meter mA A Connect Cable 60-1394-023, or you can make your own. A mA μA Com VΩ + lead – Battery, 12 Volt + + clip A good quality, adjustable, regulated DC power supply can be substituted for the 12-volt battery. The power supply should have at least 3 Amperes output, preferably more, and capable of overcurrent surges.
4220 Flow Meter Section 1 Introduction 1-14
4220 Flow Meter Section 2 Programming 2.1 Getting Started You must program the 4220 Flow Meter to accurately monitor a flow stream. You must also install the submerged probe level sensor. The 4220 will usually also need a primary measuring device, a structure placed across a stream that regulates flow. This section describes programming the flow meter with the aid of the keypad and display. There are nine program steps that control all aspects of the flow meter's operation.
4220 Flow Meter Section 2 Programming condition means that the flow meter will enable the sampler only when a certain condition or set of conditions, sensed by the flow meter, are met.) Total Flow Flow Rate Current Level Date (or pH/D.O.) Time (temperature) Following is a typical programming display on the flow meter. One of the items in the second line will be flashing. The item flashing is the selection currently held in memory.
4220 Flow Meter Section 2 Programming Print Program - Pressing this key will make the flow meter print out a complete list of the current program choices retained in memory. Print Report - One function of the flow meter is to print reports of all recorded activity at regular intervals. The contents of these reports are defined in step 1. If you set up the report generator, you can have a report printed anytime by pressing this key.
4220 Flow Meter Section 2 Programming If you want a different choice from the one that is flashing, you can move across the display by using the left and right arrow keys. Each time you press the right arrow key, the flashing selection will move one position to the right. This will continue until the flashing cursor is over the last display. You may notice an arrow pointing to the edge of the display. This indicates additional choices are available beyond what you can see.
4220 Flow Meter Section 2 Programming 2.3 Description of Program Steps The Program Steps for the 4220 are: 1. Operating Mode 2. Flow Conversion 3. Adjust Level/Parameters 4. Reset Totalizer 5. Sampler Pacing 6. Sampler Enable 7. Alarm Dialout 8. Printer 9. Reports/History Note 2.3.
4220 Flow Meter Section 2 Programming 2.3.2 Step 2, Flow Conversion Type Step 2, Flow Conversion Type, determines how the flow meter calculates flow rate and total flow. For the 4220 Flow Meter, flow rate is calculated by knowing the measured level and (usually) the characteristics of a structure called a primary measuring device. A primary measuring device is a structure placed across a flow stream through which the entire stream must flow.
4220 Flow Meter Section 2 Programming An Equation is used when you have a non-standard primary device, or want to use different values from those programmed into the look-up tables of the flow meter. Equation uses the standard flow equation: Q = k1HP1+ k2HP2 Where Q equals flow rate; k1 and k2 are constants; H is level (or head), and P1 and P2 are the powers to which the two H terms are raised. (Your equation may not have the second term, in which case you would enter 0 for the second constant, k2.
4220 Flow Meter Section 2 Programming The port is not dedicated to a particular sensor, except through programming. You can change the sensor. For example, you can change from a pH probe to a D.O. sensor if you change the programming. Selection of one parameter will keep the other from showing up on the menus. Note, however, that it is possible to measure several different stream conditions including pH and D.O. at the same time with the YSI 600 Sonde.
4220 Flow Meter Section 2 Programming control. There are several possible options—DISABLE, CONDITIONAL, VOLUME, and FLOWLINK. DISABLE will keep the sampler from receiving a flow pulse from the flow meter. VOLUME allows the flow meter to signal the sampler whenever a specific flow volume has passed by. Flowlink (this option will only appear if you are using Flowlink software) allows the sampler to be signalled from the flow meter as a result of conditions determined by Flowlink.
4220 Flow Meter Section 2 Programming combination of these conditions. The YSI 600 Sonde provides several measurements at the same time. You must have the appropriate sensors for rainfall, temperature, D. O., pH and the YSI outputs. Note 2.3.7 Step 7 - Alarm Dialout Mode You must have the optional modem to make use of this program step. The menus will not even appear unless the flow meter has a modem installed.
4220 Flow Meter Section 2 Programming Typically included are such items as maximum and minimum flow rates, the time they were reached, sample records, etc. The flow meter lets you create two separate reports, and lets you define what appears on them to a great extent. Note that the contents of the reports are defined in Setup in step 1. Step 9 merely lets you turn them off and on and set the timing. You can define the start time, the interval between reports and other aspects of the report.
4220 Flow Meter Section 2 Programming SETUP OPTIONS: 'EXIT' TO QUIT • OPTIONAL OUTPUTS • • REPORT SETUP • SETUP OPTIONS: 'EXIT' TO QUIT • LCD BACKLIGHT • • (LANGUAGE) • • PROGRAM LOCK • SETUP OPTIONS: 'EXIT' TO QUIT • (LANGUAGE) • • PROGRAM LOCK • • PROGRAM • LANGUAGE may not appear on your machine. This is intended primarily for export models, as they are programmed in different languages. Domestic models contain only English.
4220 Flow Meter Section 2 Programming YSI 600 READING INTERVAL refers to the YSI 600 Sonde. The flow meter can measure several different aspects of the stream at the same time, including pH, D.O., temperature, plus conductivity. If you select LEVEL READING INTERVAL, the following will appear: LEVEL READING INTERVAL • CONTINUOUS • • 15 SEC • • 30 SEC • • 1 MIN • • 2 MIN • • 5 MIN • The LEVEL READING INTERVAL option is a way to conserve power in battery-powered installations.
20 Flow Meter Section 2 Programming Return to the Setup menu if the unit has not already done so. This time select ENABLE/ALARM HYSTERESIS from the menu. Press Enter. The following will appear: LEVEL ENABLE/ALARM HYSTERESIS X.XXX FT (or other units of measure) The HYSTERESIS menu lets you set the range over which the level (or other condition) can vary before the flow meter responds to the change.
4220 Flow Meter Section 2 Programming The following will appear if you are measuring D. O: DO ENABLE/ALARM HYSTERESIS X.XXX PPM (or mg/l depending on units selected in Program.) A set of menus similar to those shown above for pH, D.O., and temperature will then appear for the YSI 600 Sonde, if you have selected it. You can set hysteresis for YSI-pH, YSI-D.O., YSI-Conductivity, and YSI-temperature, if these parameters have been turned on in previous program selections. 2.4.
4220 Flow Meter Section 2 Programming or two current loops supplied. If you use both the external converter and the internal board, you can have a total of four current loops controlled by the same flow meter. The internal current loops are brought out to a 6-pin M/S connector in the slot where the Modem connector is usually mounted. To activate the internal analog output(s), return to the main screen and press 4, 2, 0.
4220 Flow Meter Section 2 Programming occurs, return to the 4 - 2 - 0 option, and at the prompt enter 0 for the number of outputs activated. This will restore the external 4-20 mA converter capability. MANUAL CONTROL will appear if you continue moving to the right. “RANGE” will appear with the ANALOG OUTPUT menu if the optional internal 4-20 mA converter is present in the flow meter.
4220 Flow Meter Section 2 Programming Programming for the conditions and values that determine the operation of the 4-20 mA loop (or loops) is done in Programming Step 2, following the entry of FLOW RATE AT MAXIMUM HEAD. Note The information in the following section is provided for those who can write their own software programs to process the data transmitted from the Serial Output port. Special cables may be required. Contact Teledyne Isco technical support for more information.
4220 Flow Meter Section 2 Programming Table 2-1 ASCII Output Codes (Continued) Code Parameter Units YCO YSI 600 Conductivity Millisiemens per centimeter YSP YSI 600 Specific Conductance Millisiemens per centimeter YSA YSI 600 Salinity YTD YSI 600 Total Dissolved Solids Milligrams per liter Parts per thousand YTE YSI 600Temperature YSP YSI 600 Specific Conductance Millisiemens per centimeter Degrees Celsius YCO YSI 600 Conductance Millisiemens per centimeter YSA Parts per thousand YSI 60
4220 Flow Meter Section 2 Programming Note It is important to use CHECKSUM if you plan to use internal modems or the interrogator. The UART is shared with these devices. If you select SERIAL OUTPUT from the OPTIONAL OUTPUTS menu, the following display will appear: PERIODIC SERIAL OUTPUT • ON • • OFF Selection of OFF from this menu will disable this feature, and there will be no further references to it. The Serial Output data appears on the Interrogator connector of the flow meter.
4220 Flow Meter Section 2 Programming If you select any of these outputs, the flow meter will request that you turn them on or off. If you are running on battery and do not intend to use either of these options, select OFF. Otherwise, select ON. After the OPTIONAL OUTPUTS menus have been set, press Enter. The display will return to the SETUP menu. This time, select REPORT SETUP with the arrow key. Press Enter.
4220 Flow Meter Section 2 Programming Select YES if you want DO/PH to appear in the report. Note that you must have the appropriate sensor connected to the flow meter to sense parameters; the flow meter is capable of sensing temperature, pH and temperature, and D.O. (dissolved oxygen) and temperature. Press Enter. The following will appear: TEMPERATURE IN REPORT • YES • • NO • Select YES if you want TEMPERATURE to appear in the report.
4220 Flow Meter Section 2 Programming Press Exit to leave the program. Then press Enter and reselect SETUP. The Setup menu will reappear: SETUP OPTIONS: 'EXIT' TO QUIT • STATUS • • REPORT SETUP • • LCD BACKLIGHT • SET CLOCK, SITE ID, MEASUREMENT SETUP, PROGRAM LOCK, PROGRAM are off-screen and can be reached with the arrow keys. Select LCD BACKLIGHT with the arrow key. Press Enter.
4220 Flow Meter Section 2 Programming Select the language appropriate for your application. The menus and the printed reports will appear in the selected language. Press Enter. The SETUP menu will reappear. This time select PROGRAM LOCK from the menu. Press Enter: PROGRAM LOCK • ON • • OFF • PROGRAM LOCK keeps the program from being changed. Select OFF while you are programming, and then go back and select ON if you need to lock the program.
4220 Flow Meter Section 2 Programming temperature. Press Enter. The following will appear. You will have to press the right arrow key several times to see all of the options displayed in the following menu: FLOW RATE UNITS OF MEASURE • GPS • • GPM • • GPH • • MGD • • CFS • • CFM • • CF- Pressing the right arrow key several times will move other units onto the display.
4220 Flow Meter Section 2 Programming Their use in streams with intermittent flow (such as storm drainage) is not recommended. Note that the pH probe is a consumable item, and will eventually need replacement in any case. RAIN GAUGE • INCHES • • MM • • NOT MEASURED • You must have an Isco 674 Rain Gauge (or approved equivalent) connected to the flow meter through the Rain Gauge Port to sense rainfall. MM = millimeters. The rain gauge is factory-calibrated.
4220 Flow Meter Section 2 Programming These devices require periodic calibration for accurate sensing. See Section 5 for more information about the dissolved oxygen probe. If you are not using the D. O. probe, you would select NOT MEASURED for this step. Selection of PPM or MG/L will keep references to pH from showing up on subsequent menus. TEMPERATURE UNITS • DEG F • • DEG C • • (NOT MEASURED) • This step sets up measurement of the temperature of the flow stream.
4220 Flow Meter Section 2 Programming NO is the default. If you select YES, the following display will appear: WARNING – DO NOT DISCONNECT POWER YSI COMMUNICATION CHECK. PLEASE WAIT... CAUTION Do not disconnect either sonde or flow meter power during a communications check. The memory in the sonde can be damaged by a power failure during an update.
4220 Flow Meter Section 2 Programming This value is provided because conductivity rises (~2%/ °C) with temperature. The default setting is 1.91%. This value is the temperature coefficient for pure KCl (potassium chloride) in water. For other salts this value will be somewhat inaccurate, but it does provide a close approximation for solutions of many common salts, such as NaCl (sodium chloride), NH 4 Cl (ammonium chloride) and sea water. If you use the value of 1.
4220 Flow Meter Section 2 Programming If you select V-NOTCH, the following will appear: SELECT V-NOTCH WEIR ANGLE (IN DEGREES) • 22.5 • • 30 • • 45 • • 60 • • 90 • • 120 • If you select RECTANGULAR, the following will appear: END CONTRACTIONS ON RECTANGULAR WEIR: • YES • • NO • If you select YES, the following will appear: RECTANGULAR WEIR WITH END CONTRACTIONS ENTER CREST LENGTH XX.XXX FEET (or meters) If you select CIPOLLETTI, the following will appear: CIPOLLETTI WEIR ENTER CREST LENGTH XX.
4220 Flow Meter Section 2 Programming If you select LEOPOLD-LAGCO, this will appear: LEOPOLD-LAGCO FLUME SIZE • 4" • • 6" • • 8" • • 10" • • 12" • • 15" • • 18" • • 21" • If you press the right arrow key several times, the sizes shown below will move onto the screen: LEOPOLD-LAGCO FLUME SIZE • 24" • • 30" • If you select HS, the following display will appear: HS FLUME SIZE • 0.4' • • 0.5' • • 0.6' • • 0.8' • • 1.0' • If you select H, the following will appear: H FLUME SIZE • 5' • • .
4220 Flow Meter Section 2 Programming If you select ROUND PIPE for the Manning flow conversion, the following displays will appear: MANNING ROUND PIPE SLOPE = X.XXXXX ROUGH = X.XXXX Slope is entered as a dimensionless quantity, delta Y/ delta X, not as percent slope. Or, as otherwise expressed: ΔX = ΔY For example: Rise Run 1 100 = .01 Roughness coefficients are published in the Isco Open Channel Flow Measurement Handbook. You must know the material the pipe is made of.
4220 Flow Meter Section 2 Programming If you select TRAPEZOID for the Manning flow conversion, the following displays will appear: MANNING TRAPEZOID SLOPE=X.XXXXX ROUGH=X.XXX (Slope and roughness are entered as for ROUND PIPE previously.) Then: MANNING TRAPEZOID TOP WIDTH=X.XXX FEET (or meters) Then: MANNING TRAPEZOID BOTTOM WIDTH=X.
4220 Flow Meter Section 2 Programming The level- to-flow rate data for such devices is usually available from the manufacturer. From this data the flow meter can create a flow conversion based on the relationship between the level and flow rate. After the FLOW RATE UNITS menu has appeared, the next menu is: SET X (1-4): (0) POINTS ENTERED • ADD POINT • • (UNITS) • Then: SET 1 DATA POINT 1 ENTER: 0.00 (level units) 0.
4220 Flow Meter Section 2 Programming The flow meter's internal resolution and its accuracy are based on the value you enter for Maximum Head. The flow meter will display: FLOW RATE AT MAXIMUM HEAD X.XXX CFS (or other units of measure) Programming the 4-20 mA Outputs Note If you do not turn on the 4-20 mA output(s) in step 1, the menus determining its (their) operation will not even appear later in the program.
4220 Flow Meter Section 2 Programming current range is selected). The flow meter will then request that you repeat the process of defining the data type and setting the minimum and maximum values for any of the other analog ports you activated previously in Setup. 2.4.5 Step 3 - Parameter to Adjust This step lets you enter the measured level in the flow stream. It also lets you calibrate the pH (acidity or alkalinity), D.O. (dissolved oxygen) parameter sensors, and the YSI 600 Multiple Parameter Sonde.
4220 Flow Meter Section 2 Programming over 14%! Errors in level measurement have a greater effect on flow calculations at low liquid levels. Dimensional errors tend to be more significant at higher levels. If you select pH for Parameter to Adjust, the following display will appear. (Note that pH will not appear as an option unless you have selected it in step 1.) If you want to measure pH and pH does not appear in this step, you must go back to step 1 and select pH instead of NOT MEASURED.
4220 Flow Meter Section 2 Programming Absolute barometric pressure is barometric pressure not corrected to sea level. The barometric pressure published by the U.S. Weather Bureau is adjusted to sea level. If you use their value, you must convert it to the absolute pressure for your altitude. You should use Weather Bureau barometric pressure only if you are at sea level, or are able to correct the Weather Bureau figure to absolute pressure at your location. Enter the correct value. WRAP D. O.
4220 Flow Meter Section 2 Programming If you select pH for the parameter to calibrate, the following display will appear: YSI 600 pH CALIBRATION • pH 4 & 7 • • pH 7 & 10 • • pH 4, 7, & 10 • You can perform a two- or three-point calibration for pH with the YSI sonde. The menus that follow are similar to those in the preceding section for the Isco pH sensor. Select the calibration that best suits your stream’s profile.
4220 Flow Meter Section 2 Programming Then: PLACE PROBE IN X.XX PPT PRESS ENTER WHEN STABLE: X.XX MS/CM Then: CALIBRATING... PLEASE WAIT... There is no need to calibrate the YSI 600 temperature sensor, as it is self-calibrating.
No Err or/warning Override Ye s - No Good Cal ibrat ion Ye s Error/ Wa rn in g Calibrating... Please Wait... Ri nse probe and place in ph 10.0 solution Press Enter when stable: X.XX pH Ye s Er ror/warning Override Ye s - No G ood Cali brat ion Er ror/ Wa rnin g Calibrating... Please Wait... Rins e Probe And Place In Ph 7.0 Solution Press Enter when stable: X. XX pH Ye s No No No Er ror/warning Override Ye s - No G ood Cali brat ion Error / Wa rn in g Calibrating... Please Wait...
4220 Flow Meter Section 2 Programming 2.4.6 Step 4 - Reset Totalizer This step allows you to reset the flow meter's internal flow totalizer. Note that there is the possibility of more than one totalizer. All models have the capability of maintaining a separate totalizer for the time the sampler is enabled through the sampler enabling feature (step 6). If you select step 4, the following will appear: RESET TOTALIZER • YES • • NO • If you select NO, the flow meter will advance to the next step.
4220 Flow Meter Section 2 Programming trols the flow meter remotely, via phone lines and a modem, or locally with a laptop computer and cable. In any event, if the sampler pacing definition is controlled by Flowlink, it can only be changed through Flowlink. If you select CONDITIONAL for sampler pacing, the following will appear: CONDITION • (LEVEL) • • (FLOW RATE) • • (RAINFALL) • • (D. O.) • (pH), (TEMPERATURE), (YSI pH), (YSI DO), (YSI CONDUCTIVITY) and (YSI TEMP) may also appear.
4220 Flow Meter Section 2 Programming After you have determined what condition will signal the sampler and under what circumstances, the following menu will appear: SELECT OPERATOR • DONE • • OR • • AND • SELECT OPERATOR allows you to trigger the sampler from a single condition or from two conditions. Suppose you wanted to trigger the sampler from only one condition and that condition was LEVEL. You would select level as the condition, and then identify what change in LEVEL would be the trigger.
4220 Flow Meter Section 2 Programming useful where the sampler needs to remain idle for long periods of time, such as storm water runoff applications. When you select or advance to step 6, the following display will appear: SAMPLER ENABLE MODE • DISABLE • • ENABLE • • CONDITIONAL • • (STORM) • (Flowlink) may also appear. The Flowlink menu option will not appear unless Flowlink software is installed. DISABLE means that the sampler will be permanently inhibited by the flow meter.
4220 Flow Meter Section 2 Programming chosen from this menu before the flow meter recognizes the event as a storm. The next menu defines the interval that must pass between storm events. TIME SINCE LAST RAINFALL DAYS: X (allowable entry of 1-7) If you select CONDITION from SAMPLER ENABLE, the following will appear: CONDITION • (LEVEL) • • (FLOW RATE) • • (D. O.) • • (pH) • (TEMPERATURE), (RAINFALL), (YSI pH), (YSI DO), (YSI CONDUCTIVITY) and (YSI TEMP) may also appear.
4220 Flow Meter Section 2 Programming determine whether both conditions are necessary for enabling (AND) or whether either condition will enable the sampler (OR). If you select FLOW RATE: FLOW RATE • GREATER THAN • • LESS THAN • • RATE OF CHANGE • You would select one of these options, and then enter a value, as for LEVEL, previously. The rest of the menus will appear the same as they did for LEVEL and FLOW RATE. D. O.
4220 Flow Meter Section 2 Programming Note You must have the optional internal modem installed and connected to a telephone network to make use of this feature. This menu will not even appear if you do not have a modem. The flow meter will automatically advance to the next step. More information about the modem can be found in Section 5.
4220 Flow Meter Section 2 Programming least one of the options will be available to you. As mentioned for other program steps, some of these menu options may not appear depending on the flow meter you have and choices you made earlier in the program. D. O., pH, TEMPERATURE, and RAINFALL all require activation in step 1 and connection of the appropriate sensor to the flow meter. D. O. and pH will never appear together, as only one can be used at a time.
4220 Flow Meter Section 2 Programming arranged in decreasing priority. NUM 1 carries the highest priority, followed by NUM 2 and so on. After you have entered the numbers, the flow meter will request: DELAY BETWEEN DIALOUTS XX MINUTES This is the time delay between calling the first number and calling the second, etc. This option gives you time to respond to an alarm before the flow meter dials the next number. Select a value between 1 and 99 minutes.
4220 Flow Meter Section 2 Programming (TEMPERATURE), (YSI pH), (YSI D.O.), (YSI CONDUCTIVITY), and (YSI TEMP) can also appear. Remember that menus in parentheses may not appear due to previous program selections. Either pH or D. O. may appear, but not both. The various YSI options will only appear if you enabled the YSI 600 in step 1. You must have the appropriate probes to sense these conditions. Select OFF if you do not need this line. If you select pH, D. O.
4220 Flow Meter Section 2 Programming (TEMPERATURE), (YSI pH), (YSI D.O.), (YSI SALINITY), and (YSI TEMP) may also appear. Again, as for line A, you can select another condition to plot on the chart. The flow meter will request a full-scale value. The full-scale value can be different from that of line A. INPUT FOR PRINTER LINE C • NONE • • (LEVEL) • • (FLOW RATE) • • (pH) • • (D. O.) • (TEMPERATURE), (YSI pH), (YSI D.O.), (YSI SALINITY), and (YSI TEMP) may also appear.
4220 Flow Meter Section 2 Programming The second report generator lets you program the flow meter to print two independent reports at different intervals, for example. This is useful for those who need both a weekly and a monthly summary of activity on the flow meter. If you select OFF for this step, the program will advance to the next step, and there will be no report B generated.
4220 Flow Meter Section 2 Programming Flow Meter History Contents: D. O. ADJUSTED pH ADJUSTED LEVEL ADJUSTED FLOW CONVERSION CHANGED PLOTTER SPEED CHANGED PLOTTER TURNED ON PLOTTER TURNED OFF TIME CHANGE FROM TIME CHANGE TO REPORT A CHANGED REPORT B CHANGED REPORT A TURNED ON REPORT B TURNED ON REPORT A TURNED OFF REPORT B TURNED OFF TOTALIZER RESET INTERROGATED SAMPLER ENABLED SAMPLER DISABLED ALARM ACKED BY #X ALARM NOT ACKNOWLEDGED YSI pH ADJUSTED YSI D.O.
4220 Flow Meter Section 3 Installation 3.1 Introduction 3.1.1 Preparation for Use This section of the manual contains detailed information on the installation of the 4220 Flow Meter. Included are sections on the power sources, mounting methods, interconnection wiring, installation of the submerged probe and setup procedure for the unit. The flow meter is shipped with a roll of paper installed and a standard program in the system memory, also called the default program.
4220 Flow Meter Section 3 Installation cartridge should be colored blue. If it is violet or pinkish, you need to regenerate it. You will need to remove the particles from the cartridge for regeneration; you cannot regenerate the cartridge like the case desiccant canister. The cartridge is plastic and will melt. See Section 6 for details on regenerating the desiccant cartridge. CAUTION Do not operate the flow meter with the door open or the desiccators saturated.
4220 Flow Meter Section 3 Installation 3.1.4 Opening the Case 3.2 Connection to a Power Source To access the flow meter controls and internal plotter, you must open the case. Unlatch the two catches on the right side of the case and pull open the lid. You will need to open the case whenever you change the plotter chart or change the programming with the keypad. You can read the display through the window, so periodic readings may be taken without opening the case.
4220 Flow Meter Section 3 Installation Note Do not attempt to run a sampler from a power source installed on a flow meter. While a flow meter will operate satisfactorily powered from a sampler, the reverse is not true. The sampler's pump draws heavy current, particularly on startup. The flow meter-to-sampler connect cable cannot carry such currents. If you are using a single power source for a flow meter-sampler combination, always install the power supply on the sampler. 3.2.
4220 Flow Meter Section 3 Installation 3.2.5 Isco Lead-Acid Battery Teledyne Isco also offers a 6.5 ampere-hour 12-volt rechargeable lead-acid battery to power the flow meter. This battery is similar in size to the nickel-cadmium battery, except somewhat larger, reflecting its 50% greater capacity. For those familiar with nickel-cadmium batteries, operation and maintenance of these batteries differs somewhat.
4220 Flow Meter Section 3 Installation Figure 3-2 Power Pack Installed on a Flow Meter They are both capable of powering the flow meter. The Battery-Backed Power Pack provides 12 VDC at 5 Amperes, and is backed up by a 1.2 Ampere- hour nickel-cadmium battery. This is built in a package the same size as the standard power supply, and is intended for use where short-term power interruptions are frequent but unacceptable for flow meter operation.
4220 Flow Meter Section 3 Installation Teledyne Isco offers a special optional connect cable to power the flow meter from a separate battery. Mount the battery securely, in an upright position, so it will not inadvertently tip over, or be at risk of having its cable pulled off. CAUTION Be sure of proper polarity before attaching clips to the battery. Never attach the flow meter to a source of unknown polarity or voltage. If in doubt, check with a reliable DC voltmeter.
4220 Flow Meter Section 3 Installation 3.3.2 Location of the Flow Meter Because the flow meter uses a submerged probe for level measurement, you do not have to install it directly above the primary measuring device, or even particularly close to the flow stream. You can install the flow meter in a convenient, protected location and route the submerged probe cable to the place where the level sensor is mounted.
4220 Flow Meter Section 3 Installation submerged probe extension cables.) Do not try to patch extension cords. If you need to operate at greater lengths, use the Quick-Disconnect Box. To use the Quick-Disconnect box, you will need a cable of the correct length with an M/S connector to plug into the flow meter. Teledyne Isco will build the cable with the proper connector on one end and stripped wire ends on the other as a special order.
4220 Flow Meter Section 3 Installation mersion when the stream is high, you should protect the desiccator and the reference port by venting them at a point the water cannot possibly reach. Remember that water or other blockage inside the submerged probe's vent tube will ruin the submerged probe. It is even more important to protect the Quick-Disconnect Box this way than the flow meter, because of the likelihood that the submerged probe will be directly connected to the Quick Disconnect Box.
4220 Flow Meter Section 3 Installation WARNING The 4220 Flow Meter has not been approved for use in “hazardous locations” as such locations are defined by the National Electrical Code. Before any flow meter is installed, the proper safety precautions must be taken. The discussions of safety procedures are only general guidelines. Each situation in which you install a flow meter varies. You must take into account the individual circumstances you are in.
4220 Flow Meter Section 3 Installation 3-12
4220 Flow Meter Section 4 Submerged Probe Installation The following sections describe installation of the submerged probe. The first section presents general mounting considerations common to all submerged probe mounting techniques. The following sections describe probe installation using the two systems available for mounting the submerged probe in pipes or round-bottomed flow streams. For pipes up to 15 inches (38.1 cm) in diameter, spring stainless steel self-expanding mounting rings are available.
4220 Flow Meter Section 4 Submerged Probe Installation The probe has several ports through which liquid enters the body to contact the pressure transducer. The only way for the probe to malfunction is for all the ports to be completely blocked. Most substances likely to cover the probe are somewhat porous, which still allows the hydrostatic pressure of the flow stream over the probe to reach the transducer. 4.1.
4220 Flow Meter Section 4 Submerged Probe Installation Slanted Nose Section Slanted Nose Section – Under certain conditions - low flow rates in debris-laden small sewers, for example - the submerged probe may catch and retain the debris, obstructing the flow stream and causing erroneous level readings. To avoid this problem, use the slanted nose section. This nose section has a slanted leading edge that tends to shed debris more readily than the standard nose section.
4220 Flow Meter Section 4 Submerged Probe Installation Compress ring into gap to install in pipe, then... ...outward force of ring against pipe wall holds ring in place inside pipe. Figure 4-1 Sensor Installed on a Spring Ring Attaching the Sensor to the Ring Attach the probe to the ring either by using two 4-40 x 3/16" countersink screws or by snapping the optional probe carrier to the ring. This second method of attaching the sensor allows for easy removal in case service is needed later.
4220 Flow Meter Section 4 Submerged Probe Installation Figure 4-2 Spring Ring Preparation F lo w Figure 4-3 Mounting Ring in a Round Pipe CAUTION Make sure the sensor cable is securely fastened along the back (downstream) edge of the ring. Otherwise, the sensor may provide inaccurate level readings under conditions of high velocity. Do not overtighten the plastic cable ties; they should be tightened just enough to secure the cable in place, without greatly indenting the cable.
4220 Flow Meter Section 4 Submerged Probe Installation The spring ring may need anchoring. Under conditions of high velocity (greater than 5 feet per second or 1.5 meters per second), the ring may not have sufficient outward spring force to maintain a tight fit inside the pipe. The ring may start to lift off the bottom of the pipe in a waving fashion, or may even be carried downstream.
4220 Flow Meter Section 4 Submerged Probe Installation Assembling the Rings The scissors mechanism will work best if the respective assembly is installed to allow the scissors to expand approximately in the middle of the adjustment. Do not overtighten the mechanism. It is designed to flex somewhat to provide a positive lock, once moderately tightened. When assembling the ring sections, make sure the tongue sections are correctly inserted into the slotted sections.
4220 Flow Meter Section 4 Submerged Probe Installation 9" 1 21.5" Extension #1 68-3000-038 (pair) 60-3004-172 (each) 31.5" 41.5" 43.7 " 2 Extension #2 68-3000-039 (pair) 60-3004-173 (each) 3.4" min 15.9 max 3 Extension #3 68-3000-040 (pair) 60-3004-174 (each) 4 Scissors Mechanism 60-3004-170 Extension #4 68-3000-041 (pair) 60-3004-175 (each) Flow Standard Base Section 60-3004-169 The table below indicates recommended part configurations for various pipe diameters.
4220 Flow Meter Section 4 Submerged Probe Installation To prevent debris from catching on the cable, it is important to attach the cable to the mounting ring so it offers as little resistance to the flow as possible. Attach the sensor cable to the downstream edge of the ring, using the self-locking plastic ties supplied with the ring. Place the ties through the holes in the edge of the mounting ring and then lock them around the cable.
4220 Flow Meter Section 4 Submerged Probe Installation A variety of threaded fasteners should be available in various overall and thread lengths; a minimum thread length of 1/4 in. (0.63 cm) is recommended. Be sure to follow the manufacturer's instructions carefully, and observe all safety precautions. 4.3.5 Completing the Probe Installation The probe installation is finished by coiling any excess sensor cable and securing it using cable clamps or other means.
4220 Flow Meter Section 4 Submerged Probe Installation the mounting ring are not suitable, you must build your own mounting hardware. Figure 4-6 shows the dimensions of the probe for your reference when building your own hardware. Figure 4-7 shows the probe installed in several primary devices. Certain flume manufacturers produce flumes with built-in cavities for Isco’s submerged probe. Contact the flume manufacturer for details on the special flumes.
4220 Flow Meter Section 4 Submerged Probe Installation that certain flume manufacturers produce flumes with built-in cavities designed specifically for use with Isco submerged probe level sensors. Typically, these are Parshall or Palmer-Bowlus flumes. Contact the manufacturer for details on these flumes. Bore 0.437" X 0.330" Deep (1.11 X 0.838 cm) 0.812" (2.06 cm) 0.141" Dia. Thru. (0.358 cm) 0.670" (1.70 cm) 0.875" (2.22 cm) 3.625" (9.21 cm) 7.249" (18.42 cm) 9.38" (23.
4220 Flow Meter Section 4 Submerged Probe Installation Weir Round Pipe Sensor Mounting Plate Spring Ring Round Pipe: Probe installed from above, outside of manhole, by using Street Level Installation System. Street Level Installation Components 1. Handle 2. Ring Palmer-Bowlus Flume Base Bottom Assembly Parshall Flume Round Pipe Probe Mounted in Recess Contact the factory or your Teledyne Isco Representative for additional information about the Street Level Installation system.
4220 Flow Meter Section 4 Submerged Probe Installation 4.4.4 Securing Probe with a Weighted Plate In situations with a minimal flow velocity (for example, in a stilling well), you can simply attach the probe to a weighted plate and submerge it in the flow. 4.4.5 Completing the Probe Installation The submerged probe installation is finished by coiling any excess sensor cable and securing it using cable clamps or other means.
4220 Flow Meter Section 5 Options and Accessories This section describes accessories available for use with 4200 Series Flow Meters. There are two groups of accessories. One group of options you can use with any of the 4200 Series. The other group are accessories for a specific type of level sensing and can only be used with one flow meter of the series. This section covers the accessories that are usable with any 4200 Series Flow Meter.
4220 Flow Meter Section 5 Options and Accessories mode. It operates at 300/1200/2400 baud. This modem has speech and tone capabilities and comes with a connect cable to attach to the telephone line. Note The modem is disabled when an interrogator cable is connected to the flow meter’s interrogator port. It cannot receive incoming calls, and the alarm dialout will be rendered inoperable, while this cable is connected. Disconnect the interrogator cable in order to use the 4200T Modem. 5.1.
4220 Flow Meter Section 5 Options and Accessories telephone company will provide advance notice in order for you to make the necessary modifications in order to maintain uninterrupted service. “If trouble is experienced with this equipment, please contact the Teledyne Isco Customer Service Department, (800) 228-4373 or, outside the U.S.A., call (402) 464-0231, for repair and (or) warranty information.
4220 Flow Meter Section 5 Options and Accessories You may find the following booklet, prepared by the FCC, helpful: How to Identify and Resolve Radio-TV Interference Problems. This book is available from the U.S. Government P r i n t i n g O f f i c e , Wa s h i n g t o n , D. C. 2 0 4 0 2 , S t o c k N o . 004-00-00345-4. 5.2 Connection to External Serial Device Teledyne Isco offers two serial outputs from 4200 series flow meters.
4220 Flow Meter Section 5 Options and Accessories mation from the flow meter to a variable analog output current ranging from 4 to 20 milliamperes. When a condition measured by the flow meter is converted into an analog output, 4 mA becomes the 0% or baseline for the condition, while 20 mA becomes the 100% or full-scale of the condition. Teledyne Isco offers two different arrangements for providing the 4-20 mA outputs.
4220 Flow Meter Section 5 Options and Accessories Table 5-1 4-20 mA Output Interface Specifications 5.3.2 Internal Multiple Analog Output Board Adjustments / Calibration Factory-calibrated; when used within range, no adjustment needed, Maximum Distance 1,500 ft. (457.3 m) using 18 AWG wire. For those needing more than one analog output, Teledyne Isco offers the Multiple Analog Output Board, which is installed inside the flow meter. This board provides from one to three isolated analog outputs.
4220 Flow Meter Section 5 Options and Accessories Table 5-2 Multiple Analog Output Board Specifications Operating Temperature 0 to 140° F (–18 to 60° C) Output Accuracy ±0.5% of full-scale Resolution 0.1% of full-scale (0-20 mA) Electrical Isolation Isolated from each other and from the flow meter. Calibration Factory-calibrated; no further adjustments necessary. Maximum output distance 1500 feet (457.3 m) using 18 AWG wire.
4220 Flow Meter Section 5 Options and Accessories • It must provide a contact closure for every 0.01" (or 0.1 mm) of rainfall A cable assembly is available to connect a user-supplied rain gauge to the flow meter. More information about the rain gauge is available from the factory or the Rain Gauge Manual. Figure 5-1 674 Tipping Bucket Rain Gauge 5.5 Isco Flowlink Software Teledyne Isco offers a proprietary software system for data acquisition, storage, retrieval, and analysis.
4220 Flow Meter Section 5 Options and Accessories and the other for low. The availability of form C contacts (both NO [normally open] and NC [normally closed] contacts) means you can switch loads either on or off. Relay contacts are rated 3 Amperes at 24 volts AC or DC. The alarm box operates on 12 VDC supplied from the flow meter. Current consumption in standby condition is approximately 10 mA. In alarm, current increases to 190 mA.
4220 Flow Meter Section 5 Options and Accessories other end of the cable has 3 stripped wires. Attach them to the alarm box according to the instructions in the alarm box instruction manual. BLACK +12 VDC GREEN –12 VDC WHITE DATA 5.7 Parameter Sensing with Isco 4200 Series Flow Meters The Isco 4220 Flow Meter has the capability of displaying, recording and (if Flowlink software is used) storing data provided from parameter sensors.
4220 Flow Meter Section 5 Options and Accessories The Parameter Modules themselves plug into the Parameter Probe connector on the flow meter. CAUTION The pH and D.O. probes require continuous submersion after installation, or they will lose sensitivity. Prolonged dehydration of the sensor bulb may damage or even ruin the pH probe. 5.8 The Temperature Probe The temperature probe is the simplest of the three, consisting of a thermistor inside a metal housing.
4220 Flow Meter Section 5 Options and Accessories The dissociation constant is a number indicating the degree of ionic dissociation for a substance after it is dissolved in water. Dissociation constants vary widely for substances depending on the nature of the substance’s chemical bonds. Ionic salts tend to have higher constants. Each number on the pH scale between 7 and 0 equals a tenfold increase in H + ion. Each number between 7 and 14 equals a tenfold decrease of H+ ion.
4220 Flow Meter Section 5 Options and Accessories pH measurement is affected significantly by temperature, like any other chemical reaction. Consequently, temperature compensation is provided. in the flow meter. The Isco pH probe has a built-in temperature sensor that is exposed for faster response. The pH probe connects to a Parameter Module that plugs into the Parameter Port on the flow meter. The probe has a 25 ft. (7.6 m) cable. For greater distances, contact the factory.
4220 Flow Meter Section 5 Options and Accessories 3. Now go to step 3. PARAMETER TO ADJUST - NONE, LEVEL, pH. (Other selections may appear.) Select pH. If the pH menu does not appear in step 3, go back to step 1 to make sure you have turned it on. Then the pH selection of calibrations will appear. 4. Place the pH probe in the appropriate buffer solution until the stainless steel body is one inch beneath the surface. 5. If you select 4 & 7, the following will appear.
4220 Flow Meter Section 5 Options and Accessories Greasy substances, being nonconductors of electricity, weaken the electrical potentials formed between the glass mono-electrode and the solution, slowing or halting the response altogether. The pH Sensor operates satisfactorily mounted either horizontally or vertically in the stream. However, horizontal mounting is more secure, and presents less of a debris trap.
4220 Flow Meter Section 5 Options and Accessories Figure 5-4 pH Parameter Module Note When installing the pH probe and its sensor carrier, make sure the mounting slots on the carrier are completely pressed into the mating tabs on the ring. The probe relies on a full engagement between tabs and slots for secure mounting. If the slots are loose against the tabs, the probe may be swept away by the force of the stream.
4220 Flow Meter Section 5 Options and Accessories The Isco amplifier box extends the allowable distance between the probe and the flow meter. The probe has a 25-foot cable, so you must mount the amplifier within this distance. The maximum distance between the amplifier box and the flow meter is 1,000 feet. Table 5-3 pH Probe Specifications Description Submersible, horizontal-mounting probe with combination-type electrodes. Porous Teflon® liquid junction to resist fouling and coating.
4220 Flow Meter Section 5 Options and Accessories tap water could revive the electrode in more severe situations. If none of these solutions work, it may be necessary to replace the probe. 5.10 The Dissolved Oxygen (D.O.) Probe Note The 270 D.O. module has been discontinued. Probes, service kits, and accessories are still available to maintain existing field units. This probe measures the amount of oxygen dissolved in a stream or waterway.
4220 Flow Meter Section 5 Options and Accessories Tests conducted by Isco with probes installed in various waste streams have indicated that greases and solids quickly coat the probe's membrane, making it impossible for oxygen to enter the reaction chamber. This will result in an abnormally low reading, or no reading at all. • Cleaning the membrane is generally not effective, as it tends to drive the solids further into the membrane pores.
4220 Flow Meter Section 5 Options and Accessories 4. Place the O-ring on top of the membrane, generally conforming to the circumference edge of the probe. 5. Place the thumb and index finger from both hands opposite each other on the O-ring at equal distances. 6. Roll the O-ring down over the end of the probe, being careful not to touch the membrane where it covers the probe. 7. Trim off excess membrane with scissors or a sharp knife.
4220 Flow Meter Section 5 Options and Accessories use. Contents of the flow stream are also important, as some substances will foul the membrane very quickly. Erratic readings will result from loose, wrinkled or fouled membranes, or from large bubbles in the electrolyte reservoir. If erratic readings or evidence of membrane damage occur, you should replace the membrane and the KCl electrolyte.
4220 Flow Meter Section 5 Options and Accessories probe. Use only 2-mil membranes with D.O. probes connected to Isco flow meters. Table 5-4 D.O. Probe Specifications Cathode Gold Anode Silver Membrane FEP Teflon; 2 mil standard Electrolyte Half-saturated KCl (Potassium Chloride) Temp. Range 0° to +45° C Polarizing Voltage 0.8 Volts (Nominal) Probe Current in Air at 30° C 19 microamps (nominal) Probe Current in Nitrogen at 30° C 0.15 microamps or less Connection 12 ft. (3.
4220 Flow Meter Section 5 Options and Accessories Note You must use the Isco Temperature Probe with the D.O. Probe to provide temperature compensation. Wrap both the D.O. Probe and Temperature Probe in a damp cloth. Wait ten minutes for it to stabilize, then proceed. Go to step 1 on the flow meter. Select PROGRAM, then step through the units of measure with Enter until you reach the menu for pH measurement. Select NOT MEASURED. Press Enter. Then D.O. UNITS will appear. Select either MG/L or PPM.
4220 Flow Meter Section 5 Options and Accessories Scissors Rings. You can also install them in other ways with custom hardware as is appropriate for your situation. For details on the Spring Rings and the Scissors Ring, refer to Section 4. For installations where it is desirable to mount the parameter probe further upstream from an outfall, the four-foot probe extension is recommended. Mounting Ring Direction of Flow Probe Probe Extension Figure 5-7 Mounting of Four Foot Probe Extension 5.
4220 Flow Meter Section 5 Options and Accessories the flow meter. You can use the YSI 600 as deep as 200 feet below the surface, or in as little as a few inches of water. The fast sensor response of the YSI 600 makes it ideal for vertical profiling. Its small size means it can fit inside two-inch diameter monitoring wells. A Rapid Pulse Dissolved Oxygen Sensor eliminates the need for stirring, providing accurate results without an expensive and bulky stirrer.
4220 Flow Meter Section 5 Options and Accessories Table 5-5 YSI 600 Probe Specifications (Continued) Accuracy ± 0.4° C Resolution 0.1° C Dissolved Oxygen % Saturation Sensor Type Rapid Pulse - Polarographic Range 0 to 200 % Air Saturation Accuracy ± 2% Air Saturation Resolution 0.1% Air Saturation Dissolved Oxygen mg/L Sensor Type Calculated from % air saturation, temperature, and salinity Range 0 to 20 mg./L Accuracy ± 0.2 mg./L Resolution 0.01 mg.
4220 Flow Meter Section 6 Maintenance and Service 6.1 Routine Maintenance and Minor Service The following sections provide routine maintenance and servicing instructions. Included are sections on cleaning the flow meter, reactivating the desiccators, maintaining the bubble line, servicing the internal printer, elementary troubleshooting and servicing CMOS circuitry. Teledyne Isco recommends that you become familiar with the maintenance procedures presented here.
4220 Flow Meter Section 6 Maintenance and Service 6.2 Reactivation of the Desiccators The 4220 has a reusable desiccant canister held by a steel clamp on the inside of the case lid. There is also a tubular desiccant cartridge on the top of the case next to the connectors. The canister contains silica gel that adsorbs moisture trapped inside the flow meter's case when it is closed. This keeps the inside of the case completely dry during shipment, storage and use.
4220 Flow Meter Section 6 Maintenance and Service 6.2.2 Regenerating the External Desiccant Cartridge Teledyne Isco uses two types of silica gel: • One looks like small beads or pellets that are blue-black when dry, pale pink to transparent when saturated. • The other looks like coarse sand, yellow when dry, dark green when saturated. Regenerate desiccant by heating at 212° - 350° (100° - 175°C). MSDS (Material Safety Data Sheets) for silica gel chemicals are provided at the back of this manual.
4220 Flow Meter Section 6 Maintenance and Service Note Regeneration of the desiccators is extremely important. Saturated desiccators let the flow meter draw moisture inside, exposing both mechanical and electronic components to water and/or chemical contamination. The air in many installations contains fumes that will form acids in the presence of moisture. These acids may corrode electrical components, particularly connectors and circuit boards.
4220 Flow Meter Section 6 Maintenance and Service 6.3.2 Cleaning the Submerged Probe Without Disassembly Occasionally, organic materials may become jammed inside the submerged probe's housing. If this material swells as it becomes saturated with water, it will exert pressure on the stainless diaphragm placed over the transducer.
4220 Flow Meter Section 6 Maintenance and Service Again, any deformation of the stainless steel diaphragm will permanently disable the submerged probe. If you must remove the nose of the probe, do it very carefully. A small warning disk is located in front of the transducer. Do not remove the warning disk. After cleaning, reinstall the nose section and tighten the screws. The mounting hole must be aligned with the grounding point. Figure 6-3 Alignment of Grounding Point 6.3.
4220 Flow Meter Section 6 Maintenance and Service 6.4 Maintenance of the Printer The internal printer needs little maintenance beyond changing the chart roll and the ink ribbon. Refer to the pictures provided for each section. Also refer to the label inside the cabinet. END CAP REROLL SPOOL ASSEMBLY SUPPLY SPOOL ASSEMBLY END CAP 1. REMOVE REROLL AND SUPPLY ASSEMBLIES, 4. INSERT PAPER END INTO PLOTTER ENTRANCE AND REMOVE WHITE CAPS. (MARKED BY ARROW ON PLOTTER) AND PRESS CHART ADVANCE. 2.
4220 Flow Meter Section 6 Maintenance and Service 6. Remove the feed spool by pulling on the handle extending from the right side of the printer. 7. Snap off the other white end cap as described previously. Save the white end caps; you will reuse them. 8. Remove the empty roll from the spool by holding the handle in one hand and pulling the roll from the spool with the other. To install the new chart paper: After you remove the empty roll, 1.
4220 Flow Meter Section 6 Maintenance and Service Figure 6-5 Changing the Ink Ribbon 6.4.2 Ink Ribbon Replacement Ribbon life will vary greatly from one installation to another depending on how often the printer has to print. When the characters on the chart become difficult to read, you should replace the ribbon. If possible, try to replace the ribbon at the same time you change the paper roll, as it is easier to replace the ribbon when the roll of paper is out of the way. To replace the ribbon: 1.
4220 Flow Meter Section 6 Maintenance and Service 7. Locate the three small pins on each spool of the ink ribbon and turn the spools so the pins face the gears on the two ribbon shafts. 8. Replace the two spools on their respective shafts, pushing the detector levers out of the way so the spools will easily re-engage their gears. 9. Gently rotate each spool to tighten the ink ribbon. Reinstall the paper take-up roll if necessary. 6.4.
4220 Flow Meter Section 6 Maintenance and Service Note If you disassemble the flow meter for servicing, you will also remove the aluminum chassis covers to access the circuitry. Always replace these covers when repairs have been completed. The covers protect the circuit boards and also reduce signal emissions that could interfere with the operation of nearby electronic equipment. For the same reason, do not remove any of the ferrite beads or alter the wiring harnesses inside the cabinet in any way.
4220 Flow Meter Section 6 Maintenance and Service Figure 6-7 Location of the Three Fuses 6.5.3 Display Warnings The LCD will show various warnings and error messages to warn of problems in the program, or difficulties inside the flow meter. Some messages request routine maintenance; others point out programming errors; still others indicate serious internal difficulties. Following are typical warning messages displayed by the LCD: CHECK PRINTER FUSE PRINTER JAMMED PAPER OUT 6.5.
4220 Flow Meter Section 6 Maintenance and Service To reset the 4220 software: 1. Press Print Program for a record of your program setup. 2. Turn the flow meter off. 3. Hold down the 4 and Exit Program keys simultaneously. 4. While still pressing the 4 and Exit Program keys, turn the flow meter on again. Wait until the display reappears before releasing the keys. Pressing 4 and Exit Program may leave some values programmed into the flow meter's computer.
4220 Flow Meter Section 6 Maintenance and Service 6.6.2 Inspection Protocol When attempting to isolate problems within the unit, you should assume that the CPU and memory are working properly until attempts to find problems in the peripheral circuitry have been exhausted. This is for two reasons: • The likelihood of failure is far greater on transistor drive circuits (heavier currents are handled here), than on the CPU or memory.
4220 Flow Meter Section 6 Maintenance and Service 9. Check the input signals to the unit and see that they are correct. This may require the use of an oscilloscope. You may need to consult Customer Service for the proper appearance of wave forms. 10. Check to see that the crystal oscillator is operating and at the proper frequency. 11. Check the reset circuitry to see that it is working properly. 6.
4220 Flow Meter Section 6 Maintenance and Service circuitry. Simply being connected to the rest of the circuitry provides some protection. Most of the circuitry is well-protected from damage caused by static discharge when the unit is powered up. However, an IC should never be replaced when the unit is turned on. • Individual CMOS semiconductors and built-up printed circuit boards should always be transported in conductive packaging. Foil is satisfactory; metallized plastic bags work well.
4220 Flow Meter Section 6 Maintenance and Service mouse and keyboard commands. If you are unfamiliar with DOS or Windows, please read your DOS or Windows user manuals. 6.8.2 Before Running FLASH UPDATE The windows in FLASH UPDATE contain all the instructions you need to update the instrument. However, there are a few things to consider before running the program. • Updating your instrument erases the data stored in its memory. This includes all readings and most of the program settings.
4220 Flow Meter Section 6 Maintenance and Service 6.8.3 Running FLASH UPDATE 1. Connect the Computer Connect Cable to your computer’s serial port and the instrument’s interrogator connector (marked with the Interrogator icon). 2. Insert the update disk in the floppy disk drive. Interrogator Port Icon 3. Change the DOS prompt to the floppy disk letter prefix. 4. At the DOS command line, type: FLASHLD. The first window in FLASH UPDATE will be the Introduction window. Read it carefully before continuing. 6.
4220 Flow Meter Section 6 Maintenance and Service Options Menu Figure 6-9 Options Menu Select the COM port that corresponds to the serial port used for the Computer Connect Cable. Select Newest Version to see only the most recent update files in a directory. Select All Update Files to see all update files. Select the color scheme that best matches your monitor.
4220 Flow Meter Section 6 Maintenance and Service Table 6-1 Minimum DOS and Computer Hardware Requirements DOS DOS 3.3 or later versions DOS 5.0 or later versions recommended. Microsoft Windows not required. CPU 80286, 80386, 80486 IBM PC or compatible. 80386 or 80486 recommended. (Must operate at 19,200 baud when communicating through the serial port.) 640 kilobytes RAM (Random Access Memory), minimum Serial port For connecting the computer to Isco flow meters, flow loggers, or samplers.
4220 Flow Meter Appendix A Replacement Parts The following illustrations show replacement parts for the 4220 Flow Meter. Find their corresponding line numbers in Table A-1 for part numbers and descriptions. Replacement parts can be purchased by contacting Teledyne Isco’s Customer Service Department. Teledyne Isco, Inc. Customer Service Department P.O. Box 82531 Lincoln, NE 68501 USA Phone: (800) 228-4373 (402) 464-0231 FAX:(402) 465-3022 E-mail:IscoInfo@teledyne.
4220 Flow Meter Appendix A Replacement Parts A-2
4220 Flow Meter Appendix A Replacement Parts A-3
4220 Flow Meter Appendix A Replacement Parts Table A-1 Replacement Parts List No Part Number Complete Part Description 1 60-3214-139 Motor Assy Chart Drive 4200 4 60-3214-093 LCD Module Assy B/L 5 60-3224-070 Case Bottom Sub Assembly 8 60-3214-098 PCB Assembly Keyboard 9 60-3224 -071 PCB Assy 4220 CPU w/Software 10 60-3224-056 PCB Assembly 4220 Amplifier 11 60-3214-120 Wiring Assy - 12 VDC/Sampler 12 60-3214-122 Wiring Assy - Interrogator 13 60-3214-140 Wiring Assy - Rain Gauge
4220 Flow Meter Appendix A Replacement Parts Table A-1 Replacement Parts List (Continued) No Part Number Complete Part Description 67 202-4001-20 O-Ring Silicone #120 .99 ID × .
4220 Flow Meter Appendix A Replacement Parts A-6
4220 Flow Meter Appendix B Accessories List 4220 Flow Meter Appendix B Accessories List 4220 Flow Meter Basic Unit, 68-4220-001 Submerged Probe, 10’ range (with 25-foot cable) ................................................................... 60-3224-002 4220 Flow Meter....................................................................................................................... 60-3224-049 Accessory Package (Includes Desiccant Can and Cannister, Chart Paper Roll, and Printer Ribbon) .............
4220 Flow Meter Appendix B Accessories List Optional Equipment High-Low Alarm Relay Box ..................................................................................................... 60-3404-028 4-20 mA Output Interface........................................................................................................ 60-1784-039 674 Rain Gauge (0.01”) ............................................................................................................ 60-3284-001 674 Rain Gauge (0.1mm)) .
4220 Flow Meter Appendix B Accessories List Miscellaneous – for use with the YSI Sonde Calibration/Transport Bottle Kit............................................................................................. 60-0603-216 Carrying Case for YSI 600, cables, accessories, tools ............................................................. 60-0603-217 YSI 600 Instruction Manual ....................................................................................................
4220 Flow Meter Appendix B Accessories List B-4
4220 Flow Meter Appendix C Programming Worksheets 4220 Flow Meter Appendix C Programming Worksheets Use this form to make a hard copy of the program you use in your 4220. Most program steps can be completed in the shop without the flow meter being installed or at the job site. However, please note the following: • Do not attempt to fill out this sheet without first studying the manual, especially Section 2.
4220 Flow Meter Appendix C Programming Worksheets 7. YSI 600 Reading Interval: Continuous, 15 Sec, 30 Sec, 1Min, 2 Min, 5 Min 8. Level Enable/Alarm: Hysteresis ________Feet (Or Meters) 9. Flow Rate Enable/Alarm Hysteresis _______Units______ 10. Temperature Enable/Alarm Hysteresis ______Deg. F (C) 11. pH Enable/Alarm Hysteresis ________pH units. 12. D.O. Enable/Alarm Hysteresis ________ppm. 13. YSI-pH Enable/Alarm Hysteresis ________pH units. 14. YSI-D.O. Enable/Alarm Hysteresis ________ppm/mg./L 15.
4220 Flow Meter Appendix C Programming Worksheets 43. Rain Gauge: Inches, MM, Not Measured 44. pH Units Of Measure: pH, Not Measured 45. D. O. Units: MG/L, PPM, Not Measured 46. Temperature Units: Deg F, Deg. C, (Not Measured) 47. YSI-pH Units Of Measure: pH, Not Measured 48. YSI-D.O. Units: MG/L, PPM, Not Measured 49. YSI-Conductivity Parameter: Specific Conductivity, Salinity, Conductivity, T.D. S. ________Units________Coeff._______ 50. YSI-Temperature: Deg F, Deg. C, (Not Measured) C.
4220 Flow Meter Appendix C Programming Worksheets 22. Mann. Rectangular: Slope = __.____ Rough = __.____ 23. Mann. Rectangular: Width = __.____ Feet (or meters) 24. Mann. Trapezoid: Slope = __.____ Rough = __.____ 25. Mann. Trapezoid: Top Width = __.____ Feet (or meters) 26. Mann. Trapez.: Bottom Width = __.____ Feet (or meters) 27. Select Data Set: One, Two, Three, Four, (None) 28. Level Units For Data Point Entry: Ft., In, M, Mm 29.
4220 Flow Meter Appendix C Programming Worksheets 6. (YSI 600 D.O. Calibration only): D.O. Standard, Absolute Barometric Pressure, Altitude 7. Altitude Units Of Measure: Ft., M (D.O. probe only)___ (job site only) 8. (D.O. only) Enter Altitude: Altitude = _______ Feet (or meters, at job site only) 9. Wrap D.O. Probe In Moist Cloth: Press Enter When Stable: __.____ MG/L (at job site only) 10. Conductivity Calibration Units: Ms/cm, Ppt__________ 11. Place Probe In ______Ms/cm.
4220 Flow Meter Appendix C Programming Worksheets 13. Printer On/off With Enable: Yes, No C.7 Alarm Dialout Mode You must have a modem installed in the flow meter for any of these menus to appear. 1. Alarm Dialout: Disable, Conditional, Storm, Flowlink 2. Level: Greater Than __.____ Feet. 3. Rainfall Amount: __.____ Inches (other units) 4. Rainfall Time Period: 15 Min, 30 Min, 1 Hr, 2 Hr, 4 Hr, 6 Hr, 8 Hr, 12 Hr, 24 Hr, 48 Hr, 72 Hr 5. Time Since Last Rainfall: Days: ______ (allowable 1-7) 6.
4220 Flow Meter Appendix C Programming Worksheets 8. Print Report B at Yr:_____Month:___Day:___ Hr:___ Min:___ 9. Print Flow Meter History: Yes, No 10. Print Flow Meter History: Print Since Last, Print All NOTES Additional table for Data Point Entry Data Point Set #2.
4220 Flow Meter Appendix C Programming Worksheets C-8
4220 Flow Meter Appendix D General Safety Procedures In field installations of 4220 Flow Meters and associated equipment, the safety of the personnel involved should be the foremost consideration. The following sections provide safety procedures for working in and around manholes and sewers. the first section offers general safety advice. The second section deals with the special problem of hazardous gases found in sewers.
4220 Flow Meter Appendix D General Safety Procedures “Lifting Injuries. Unless proper tools are used to remove manhole covers, back injuries or injuries to hands or feet may result. “2. Planning. Advance planning should include arrangements for test equipment, tools, ventilating equipment, protective clothing, traffic warning devices, ladders, safety harness, and adequate number of personnel. Hasty actions may result in serious injuries. Time spent in the manhole should be kept to a minimum. “3.
4220 Flow Meter Appendix D General Safety Procedures the manhole opening. To avoid a serious injury, a person should not be lifted out of a manhole by his arm unless it is a dire emergency. “When more than one person must enter a manhole, the first person should reach the bottom and step off the ladder before the next one starts down. When two men climb at the same time, the upper one can cause the lower one to fall by slipping or stepping on his fingers. “5. Traffic Protection.
4220 Flow Meter Appendix D General Safety Procedures “10. Field Equipment. The following equipment will be available for use: Blowers Gloves Traffic cones Breathing apparatus Hard Hats Coveralls Harnesses First aid kits Manhole irons Emergency flashers Pick axes Flashlights Rain slickers Mirrors Ropes Gas detectors Safety vests Gas masks Waders” D.2 Lethal Atmospheres in Sewers The following is an article written by Dr. Richard D.
4220 Flow Meter Appendix D General Safety Procedures “It seems unlikely that anyone has ever died in a sewer from suffocation, that is, a lack of oxygen. Deaths have often been attributed to ‘asphyxiation.’ This is a word which, according to the dictionary, is used to mean death from an atmosphere that does not support life. The word has sometimes been misinterpreted as meaning suffocation, which is only one kind of asphyxiation. “In nearly all cases of death in sewers, the real killer is hydrogen sulfide.
4220 Flow Meter Appendix D General Safety Procedures number of harmful vapors. They, too, are sensed by smell and explosimeter tests if they get into the public sewer. Such occurrences are rare. “The attempt to instill a sense of urgency about real hazards is diluted if a man is told to give attention to a long list of things that in fact are irrelevant. “Be very careful to avoid high H2S concentrations, flammable atmospheres, and hazards of physical injuries.
4220 Flow Meter Appendix D General Safety Procedures Table D-1 Hazardous Gases (Continued) Gas Chemical Formula Common Properties Specific Gravity or Vapor Density Air =1 Physiological Effect Max Safe 60 Min. Exposure ppm Max. Safe 8 Hour Exposure ppm Explosive Range (% by vol. in air) Limits lower/upper Likely Location of Highest Concentration — At bottom; when heated may stratify at points above bottom. Products of combustion, sewer gas, sludge. Also issues from carbonaceous strata.
4220 Flow Meter Appendix D General Safety Procedures Table D-1 Hazardous Gases (Continued) Gas Gas Hydrogen Sulfide Methane Chemical Formula Chemical Formula H2S CH4 Common Properties Common Properties Specific Gravity or Vapor Density Air =1 Specific Gravity or Vapor Density Air = 1 Irritant and poisonous volatile compound. Rotten egg odor in small concentrations. Exposure for 2 to 15 min. at 0.01% impairs sense of smell. Odor not evident at high concentrations. Colorless. Flammable. 1.
4220 Flow Meter Appendix D General Safety Procedures Table D-1 Hazardous Gases (Continued) Gas Chemical Formula Common Properties Specific Gravity or Vapor Density Air =1 Physiological Effect Max Safe 60 Min. Exposure ppm Max. Safe 8 Hour Exposure ppm No data. Would vary widely with composition. Explosive Range (% by vol. in air) Limits lower/upper Likely Location of Highest Concentration 5.3 Near top of structure. From digestion of sludge. See components.
4220 Flow Meter Appendix D General Safety Procedures D-10
4220 Flow Meter Appendix E Material Safety Data Sheets E.1 Overview This appendix provides Material Safety Data Sheets for the desiccant used by the 4220 Flow Meter. Isco cannot guarantee the accuracy of the data. Specific questions regarding the use and handling of the products should be directed to the manufacturer listed on the MSDS.
4220 Flow Meter Appendix E Material Safety Data Sheets Indicating Silica Gel Material Safety Data Sheet Identity (Trade Name as Used on Label) Manufacturer : MSDS Number* : MULTISORB TECHNOLOGIES, INC. M75 (formerly Multiform Desiccants, Inc.) Address: CAS Number* : 325 Harlem Road Buffalo, NY 14224 Phone Number (For Information): 716/824-8900 Emergency Phone 716/824-8900 Date Prepared: July 6, 2000 G.E.
4220 Flow Meter Appendix E Material Safety Data Sheets Page 2 Section 5 - Health Hazard Data PRIMARY ROUTES OF ENTRY HEALTH HAZARDS Inhalation Ingestion NTP CARCINOGEN LISTED IN Skin Absorption Not Hazardous IARC Monograph Acute May cause eye, skin and mucous membrane irritation. Chronic Prolonged inhalation may cause lung damage. OSHA Not Listed Signs and Symptoms Drying and irritation. of Exposure Medical Conditions Asthma.
4220 Flow Meter Appendix E Material Safety Data Sheets MATERIAL SAFETY DATA SHEET March 8, 2005 M163 Effective Date MSDS Number Section 1 – Product and Company Information Product Name: Silica gel, indicating, yellow Product Use: Grades: Synonyms: Desiccant, absorbent Silica gel, indicating Amorphous silica gel, SiO2, silicon dioxide (amorphous) Company; Street Address: City, State, Zip, Country: Multisorb Technologies, Inc.
4220 Flow Meter Appendix E Material Safety Data Sheets Section 4 – First Aid Measures Eyes: Rinse the eyes well with water while lifting the eye lids. If irritation persists, consult a physician. Skin: Wash affected area with soap and water. Ingestion: Ingestion is unlikely, this material will pass through the body normally. Inhalation: Remove the affected person to fresh air and get medical attention if necessary.
4220 Flow Meter Appendix E Material Safety Data Sheets Section 8 – Exposure Controls/Personal Protection Use exhaust ventilation to keep the airborne concentrations below the exposure limits. Engineering Controls: Respiratory Protection: Use NIOSH approved respirator when the air quality levels exceed the TLV's. Skin Protection: Light gloves will protect against abrasion and drying of the skin. Eye Protection: Safety glasses.
4220 Flow Meter Appendix E Material Safety Data Sheets Section 11 – Toxicological Information This product and its components are not listed on the NTP or OSHA Carcinogen lists. Animal Toxicology Tests for DOT Hazard classification ( Tests Conducted on finely ground silica gel) 1 - hour LC50 (rat) > 2 mg / l 48 - hour oral LD50 (rat) est. > 31,600 mg / kg 48 - hour dermal LD50 (rabbit) est.
4220 Flow Meter Appendix E Material Safety Data Sheets Section 16 – Other Information HMIS – Hazardous Materials Identification System HMIS Rating Health 0 Flammability 0 Reactivity 0 0 - minimal hazard, 1 - slight hazard, 2 - moderate hazard, 3 - serious hazard, 4 - severe hazard This MSDS was prepared by: George E. Mckedy Senior Applications Development Specialist Multisorb Technologies, Inc.
4220 Flow Meter Index Numerics I 4-20 mA Output, 2-15, 5-4 Programming, 2-35 4200T Modem, 5-1 Indicators, 1-4 Installation Flow Meter, 3-7 Parameter Probes, 5-10 Submerged Probe, 4-1 U-Channel or Invert Mounting, 4-9 A Accessories, 1-2, B-1 Alarm Box, 2-20, 5-8 Analog Output, 2-15, 5-4 Programming, 2-35 Associated Equipment, 1-1 B Battery Calculations, 1-10 Battery Life, 1-10 C Chart Paper Replacement, 6-7 Compatible Equipment, 1-1 Connectors, 1-4 Controls, 1-4 Current Draw, 1-12 D Desiccant Caniste
4220 Flow Meter Index P Parameter Probes, 5-10 D.O.
Compliance Statements DECLARATION OF CONFORMITY Manufacturer's Address: Equipment Type/Environment: Trade Name/Model No: Year of Issue: Standard IEC 801.2 Description Electrostatic Discharge IEC 801.3 Radiated RF Immunity IEC 801.
DECLARATION OF CONFORMITY Application of Council Directive: Manufacturer's Name: Manufacturer's Address: Standard of C on fo rm ity Equipment Type/Environment: Trade Name/Model No: Year of Issue: Standards to which Conformity is Declared: 89/336/EEC – The EMC Directive 73/23/EEC – The Low Voltage Directive Teledyne Isco, Inc. 4700 Superior, Lincoln, Nebraska 68504 USA Mailing Address: P.O.
Warranty
Teledyne Isco One Year Limited Factory Service Warranty * Teledyne Isco warrants covered products against failure due to faulty parts or workmanship for a period of one year (365 days) from their shipping date, or from the date of installation by an authorized Teledyne Isco Service Engineer, as may be appropriate. During the warranty period, repairs, replacements, and labor shall be provided at no charge.