MODEL 329R OXYGEN DETECTOR INSTRUCTION MANUAL P/N M57726.
Copyright © 1992 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-1580.
Table of Contents 1.0 Introduction Main Features ............................................................................................... 1 Micro-Fuel Cell ............................................................................................ 2 E-2 Cell Warranty Conditions ...................................................................... 2 2.0 Installation & Operation Precautions....................................................................................................
TELEDYNE ANALYTICAL INSTRUMENTS
329R Introduction 1.0 Introduction The Teledyne Analytical Instruments Model 329R Oxygen Detector measures the percent of oxygen in a wide range of gases and gas mixtures. Modular and compact design provide outstanding performance and reliable on-line monitoring and make the 329R virtually maintenance-free. A special range on the 329R enables you to calibrate in ambient air.
1.0 Introduction 329R 5. Analog Voltage Output. The Model 329R provides customer connections to external devices, such as computers, recorders, alarms, etc., at 0–10 VDC, negative ground, plus contact closures for range identification. Custom voltage outputs, either grounded or isolated, are available as options. 6. Analog Current Output. The Model 329R has a 4–20 mA current output suitable for connection to external devices that use a 4–20 mA input.
329R Installation & Operation 2.0 Installation & Operation Precautions 1. Read this chapter in its entirety before installing the unit. 2. Upon receipt of the instrument, inspect the entire unit for damage. Notify the shipper if any damage is found. Check the unit, probe, and included accessories for broken or loose parts. 3. The sample must be free of entrained solids and water. A high humidity sample is ideal since it will prevent water loss from the cell electrolyte. 4.
2.0 Installation & Operation The analysis unit should be installed in the process at a maximum distance of 6 feet away from the main unit, depending upon the level of electromagnetic noise in the operating environment. Analysis Unit Connection The receptacle for the analysis unit cable is located in the lower lefthand corner of the rear panel. The 6-pin Minifit fitting is keyed to fit only one way in the receptacle; do not force it in.
329R Installation & Operation 2.0 Alarm Terminal Connection SPDT relay contacts are provided for each of the 3 alarms. There are 9 terminals available on the back panel of the system for alarm circuit connection. The explanation for the relay terminals abbreviations is as follows: NC = Normally closed C = Common NO= Normally open The alarm relay circuits are designed for failsafe operation, meaning that the relays are activated (energized) during normal operation.
2.0 Installation & Operation Setting Alarms The alarms can be adjusted over the full scale range, and they are set as a percent of full scale. Thus, an alarm set to indicate 9.6 on the display while on the 10% O2 range will activate at 9.6% O2 on the 10% O2 range and at 2.88% O2 on the 3% O2 range (96% of full scale). To set the alarms: 1. Place the range switch located below the front panel display in the 10% range position (or highest range available). 2. Place the power switch in the ON position. 3.
329R Installation & Operation 2.
2.
329R Maintenance 3.0 Maintenance Replacing the Fuse If the unit will not turn on when plugged in and switched ON (via the power switch), or you suspect a blown fuse, replace the fuse using the following procedure: 1. Disconnect the AC power and place the power switch located on the rear panel in the OFF position. 2. The fuse is located next to the power cord receptacle in the upper left-hand corner of the rear panel. Remove the power cord from the receptacle. 3.
3.0 Maintenance 329R 5. Replace the fuse holder into its receptacle, pushing in firmly until it clicks. In units with DC power, the fuse is located on the rear panel above the ON/OFF switch. Open the fuse holder by unscrewing and removing the knob marked FUSE. The fuse will be located inside the knob.
329R Maintenance 3.0 is punctured and air permitted to enter, the cell will immediately start to react to the presence of oxygen. Warranty Conditions The Class E-2 cell used in the 329R is warranted for twenty-four (24) months of service from the date of shipment. Customers having warranty claims must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty material or workmanship, the cell will be replaced free of charge.
3.
329R Appendix Appendix Specifications Ranges: Air Calibration Range: Analog Output: Display: Alarms: Accuracy: Response: Operating Temp. Range: Power: Enclosure: Standard: 0-3% and 0-10% O2 with 0.01% O2 resolution over full scale. Optional: 0-1% O2 0-25% O2 0-100% O2 0-25% O2 with 0.1% O2 resolution over full scale. Standard: 0–10 VDC, negative ground 4-20 mA, negative ground plus contact closures for range identification.
Appendix 329R Cell Type: E-2 Cell Warranty: 24 months Cell Life (Expected): 48 months Max. Distance Between Control Module and Probe: 6 feet (Maximum distance depends on the level of electromagnetic noise in the operating environment.) Dimensions: 2.75 "H × 6.0 "W × 2.75 "D (63.5 mm × 152.4 mm × 63.5 mm) Faceplate: 3.75 "H × 7.0 "W (95.25 mm × 177.8 mm) Probe unit: 4.0 "H × 6.0 "W × 2.5 "D (101.6 mm × 152.4 mm × 63.5 mm) Spare Parts List QTY.
329R Appendix Drawing List C-57490 A-58154 C-57304 D-57302 C-57305 C-57307 B-58123 Note: Outline diagram Probe interconnection diagram Electronic PCB assembly diagram Electronic PCB schematic diagram Power supply PCB schematic diagram Power supply PCB assembly diagram Analysis unit (probe) outline 329R analyzers with 25% and 100% range modifications incorporate the following changes: 1. 2. 3. 4. 100% range modified circuit changes: JP2 - connects pins 1 and 2 R52 is 12.5K R30 is 7.5K R35 is 1.5K 1.
Appendix 329R Material Safety Data Sheet Section I – Product Identification Product Name: Manufacturer: Address: Phone: Date Prepared or Last Revised: Emergency Phone Number: Micro-Fuel Cells Mini-Micro-Fuel Cells, all classes Super Cells, all classes except T–5F Electrochemical Oxygen Sensors, all classes.
329R Appendix Section III – Physical Hazards Potential for fire and explosion: The electrolyte in the Micro-Fuel Cells is not flammable. There are no fire or explosion hazards associated with Micro-Fuel Cells. Potential for reactivity: The sensors are stable under normal conditions of use. Avoid contact between the sensor electrolyte and strong acids.
Appendix 329R Section V – Emergency and First Aid Procedures Eye Contact: Skin Contact: Ingestion: Inhalation: Flush eyes with water for at least 15 minutes and get immediate medical attention. Wash affected area with plenty of water and remove contaminated clothing. If burning persists, seek medical attention. Give plenty of cold water. Do not induce vomiting. Seek medical attention. Do not administer liquids to an unconscious person. Liquid inhalation is unlikely.