User guide

Paramagnetic O2 Purity Analyzer Appendix
Teledyne Analytical Instruments 87
Example 1:
With pressure compensation factor set to 0.00 and the vent needle valve
fully open:
Oxygen: 100.01 Pressure: 122210
With pressure compensation factor set to 0.00 and the vent needle valve
not fully open:
Oxygen: 101.15 Pressure: 123554
Percent change of oxygen:
14.1
01.100
01.10015.100
100
Percent change of pressure:
10.1
122210
122210123554
100
New Pressure compensation factor:
1.14/1.10 = 1.04
8. Input the New Pressure Compensation Factor calculated from
your test data.
Press the System key
Select the ‘PressureAdj’ function
Press the Enter key
Adjust the value with Up/Down arrow keys to value you calculated
(In Eample 1 above, the New Pressure Compensation Factor would
be set to ‘1.04’)
Press the Enter key.
The compensation is now set. The factor can still be tuned further, by
repeating steps 5 through 8. The pressure compensation factor entered at
the end of step 8 is now your starting factor. On the second, and further
iterations, you will find less and less change in the O
2
reading. A change
of 0.02 % O
2
or less would be acceptable. If it is more than that, go back