Instruction Manual

A-2
Appendix Model 3000TB
Teledyne Analytical Instruments
Accuracy: ±2% of full scale at constant temperature.
±5% of full scale over operating temperature range, on
factory default analysis ranges, once thermal equilibrium
has been achieved.
XL-Option Accuracy: ±1 ppm on 0-10 ppm once thermal equilbrium is
reached.
0-1 ppm range is
+0.2 ppm once thermal equilibrium
has been achieved.
Analog outputs: 0-1 V dc percent-of-range
0-1 V dc range ID.
4-20 mA dc percent-of-range, isolated
4-20 mA dc range ID, isolated
Password Access: Can be user-configured for password protection.
A-2 Recommended 2-Year Spare Parts List
Qty. Part Number Description
1 C62959 Display PCB
1 D65295-A Customer Interface PCB
1 C62368-A Trace Preamplifier Board
1 C73870-B Main PCB (4-20mADC Option)
3 F1295 Fuse, 4A, 250V, 5x20 mm, T (Slow Blow)
1 O38 O-ring
1 C6689-L2C Micro-Fuel Cell
1 C69367 3000T-XL Amplifier
1 C6689-B2CXL B2CXL Micro-Fuel Cell
1 B66378 B2C Adapter for Cell Block
A minimum charge is applicable to spare parts orders.
Note: Orders for replacement parts should include the part number (if
available) and the model and serial number of the instrument for
which the parts are intended.