Percent Oxygen Analyzer Maintenance 5
5-5
Teledyne Analytical Instruments
2. Unlatch and open the outer and inner doors.
3. Leave the cell block installed. Pull up on the probe, with a slight
rocking motion, to release it from the probe receptacle.
4. Do not remove the O-rings unless they are worn and no longer hold the
probe tightly. (If worn, replace them.)
Figure 5-2: Exploded View of Cell Block and Micro-Fuel Cell
5. When it is free, unscrew the cap from the nylon probe. Hold the probe
vertically to prevent dropping the cell out of the probe.