Owner's manual

Percent OxygenAnalyzer Maintenance 5
5-1
Teledyne Analytical Instruments
5.0 Maintenance
Aside from normal cleaning and checking for leaks at the gas connec-
tions, routine maintenance is limited to replacing filter elements and fuses,
and recalibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
5.1 Replacing the Fuse
Remove Power to Unit before replacing the
fuse.
- Push and turn the Fuse Holder 1/2 turn to the left.
- Remove the old fuse, and
- Assemble in thenew fuse in reverse order
Figure 5-1: Fuse holder, 3AG Type