Turbine Generator Gas Analyzer Appendix
Teledyne Analytical Instruments 39
Appendix
A.1 Specifications
Range: 0-100% hydrogen in CO
2
0-100% air in CO
2
80-100% hydrogen in air
Accuracy: +5% at constant temp & pres (equilibrium)
+
6% at constant pressure over operating
temp range
Sensor: Thermal conductivity cell
Resolution: 0.1%
Response Time: 0-90% in less than 10 seconds (0.5 SCFH
flowrate)
Drift Rate: <1%/day
Display: 3 1/2” LCD
Flowrate: 0.5-1 SCFH nominal
2.5 SCFH maximum
(0.5-30 psig)
Wetted Parts: Brass, SS, aluminum, glass, Teflon, nylon
AC Power: 9V DC power adapter powered from
a 90-264 47-63 Hz VAC (US type 2 prong
plug)
Enclosure: NEMA 4X PET plastic housing
Dimensions: (10.6 L x 4.8 H x 9.9 W)in
(269.2 x 121.9 x 251.5) mm
Operating Temp: 0-40°C
Voltage output: 0-1VDC (0.8-1 VDC 80-100% range)