Instruction Manual

Page 19 of 26 M2240 R0
both 0-5 V and 4-20 mA, and the percent of hydrogen
being detected.
Note: before proceeding with this function, first
select the appropriate Analog Output Configuration
Mode.
Press the Enter button while “Analog Full Scale” is
displayed.
Specify the hydrogen concentration represented by both
the maximum voltage from the 0-5 V Analog Output and
the maximum current from the 4-20 mA Analog Output
(Default, 2% hydrogen). The minimum voltage and
current value is factory set to represent 0% hydrogen,
and is not configurable.
For example, using the second “0-5 Volt Config”
format of “0.5-4.5 Volt Mode”, the operator chooses to
designate the maximum voltage output, 4.5 V, to
indicate 4% hydrogen in air. This results in 4.5 V =
4% hydrogen, 0.5 V= 0% hydrogen, and 2.5 V = 2%
hydrogen.
For example, using the 4-20 mA Analog Output, the
operator chooses 20 mA to indicate 10% hydrogen in
air. This results in 20 mA = 10% hydrogen, 4 mA =
0% hydrogen, and 12 mA = 5% hydrogen.
Use the Scroll buttons to increase or decrease the percent
value. Use the Mode button to change the digit affected
by the scroll buttons.
Press the Enter key to set the displayed number into
memory and return to the Configuration Mode.
Backlight – Controls how the LCD backlight operates.
Press the Enter button while "Backlight" is displayed.
Select one of three options:
Always Off
Always On
Page 20 of 26 M2240 R0
On With Change - Under this option, the backlight
illuminates for 15 seconds when any button is
pushed, or on any Alarm. It also illuminates
continuously in configuration mode.
Press the Enter button to select the option.
Press the Mode button to exit without making any
change.
Rev X.xx – The controller firmware revision level is
displayed.