21 Settle time: 30 min
(Y/N)?
Typey
22 Apply 30.000%H2:
Ready (Y/N)?
Typey
Streaming data…
23 Taking Average…
cap=0.00000
res=x.xxxxx
Calibration Gas #2 finished.
Calibration complete.
4) Per TELEDYNE ANALYTICAL INSTRUMENTS’s recommendation, upon completion of
Calibration, conduct the Verification sequence as described in the previous section to
verify that the calibration process was executed correctly.