User manual
WaveSurfer 3000/3000z Oscilloscopes Operator's Manual
Setting Up WaveScan
Set up your source channel and triggers before setting up the scan.
1. Press the front panel Stop button to stop acquisition.
2. Choose Analysis > WaveScan.
3. Check Enable.
4. Choose the Source waveform.
5. Choose the Scan Mode and enter values for any additional settings that appear at the right of the
dialog based on your selection.
6. Select each Scan View in which you wish to display results by checking the box at the bottom of the
dialog. Each view selected is displayed simultaneously.
7. Optionally, choose an Action to trigger when an event that meets your scan criteria is found.
8. Restart acquisition.
Scan Modes
The Scan Mode determines the type of search to be performed. Select the Mode along with the Source
trace to be searched on the main WaveScan dialog. For each mode, different controls appear on the
WaveScan dialog, providing additional inputs to the search criteria. Make the appropriate entries in these
fields before starting the search.
Edge Mode
Edge Mode is used for detecting the occurrence of edges. Events that meet the threshold level are
captured and tabulated. When the acquisition is stopped, scan filters can be applied to the edges to find
specific characteristics. Edge Mode settings are:
l Slope. Choose Pos, Neg, or Both.
l Level is (set in...). Choose Percent or Absolute.
l Percent/Absolute Level. Enter a threshold value as a percentage of Top to Base or voltage level.
86