User manual

WaveSurfer 3000/3000z Oscilloscopes Operator's Manual
Measure
Measurement parameters are tools that give you access to a wide range of waveform properties. Use
them to analyze many attributes of your waveform such as rise-time, rms voltage, and peak-to-peak
voltage. Measurements can also be graphed as a trend for statistical analysis.
Create a custom set of parameters drawn from all available measurements.
Measure Table
Measurement readouts appear in a table below the grid. The value row shows the measurements taken
for each parameter on the last cycle of the last acquisition. Other rows show optional statistics.
Measurement Status Indicators
Symbols in the status row of the Measure table indicate the following:
Symbol Description
Problem with the signal or the setup. Touch cell
to see explanation in message bar.
Valid value returned.
Unable to determine top and base; however,
measurement may still be valid.
Underflow condition.
Overflow condition.
Simultaneous underflow and overflow condition.
Navigating with the Table
Close setup dialogs when the Measure table is displayed to maximize the touch screen area available for
viewing waveforms. To quickly return to the Measure dialog when closed, touch anywhere in the table.
62