User manual

WaveSurfer 3000/3000z Oscilloscopes Operator's Manual
Trigger
Triggers define the event around which digitized information is displayed on the grid.
Different Trigger Types are used to select different events in the trigger source waveforms: edge voltages,
pulse widths, high/low states, etc. These may be a single channel event or a complex pattern of events
across several channels. On instruments with Mixed Signal capabilities, many triggers can be set on either
analog channels, including the External Trigger input, or digital lines.
In addition to the type, the Trigger Mode determines how the instrument behaves as it encounters trigger
events: take a single acquisition and stop, holding on to the display of the last acquisition, or continuously
take and display acquisitions.
In both cases, when the previous acquisition has completed processing, the oscilloscope is again ready to
acquire and the READY indicator is lit. If, while READY, the trigger circuit detects a signal that matches the
trigger conditions, the oscilloscope triggers on the next matching event, and the TRIG'D indicator is lit.
Unless modified by a pre- or post-trigger Delay, the trigger event appears at time 0 at the horizontal center
of the grid, and a period of time equal to five divisions of the timebase is shown to the left and right of it.
Delay shifts the acquisition "window" on screen, displaying a different portion of the waveform.
An additional condition of Holdoff by time or events is available for Edge and Pattern triggers, including
those that appear within MultiStage triggers. Holdoff arms the trigger on the first matching event, inserts
the holdoff count, then triggers on a subsequent event. Often, especially with repetitive signals, the initial
arming event appears to the left of the trigger in "negative" acquisition time.
Trigger Modes
The Trigger Mode determines how often the instrument acquires. It is equivalent to how analog
oscilloscopes "sweep," or refresh, the display. Trigger Mode can be set from the Trigger menu or from the
front panel Trigger control group.
In Single mode, when you choose Trigger >Single or press the front panel Single button, the oscilloscope
readies, arms, and triggers provided all trigger conditions (including Holdoff) are met. It then stops and
continues to display the last acquisition until a new one is taken. The oscilloscope remains armed unless
manually stopped or triggered, and if a valid trigger does not occur, invoking Single a second time will force
a trigger and display the acquisition.
In Normal mode, operation is the same as in Single, except that the trigger automatically re-arms after the
previous acquisition is complete, and data is continuously refreshed on the touch screen.
Auto operates the same as Normal mode, except that a trigger is forced if the trigger event has not
occurred within a preset timeout period.
Stop ceases acquisition processing until you select one of the other three modes. The arming and Holdoff
counters are cleared, even if there has not yet been a trigger since the previous acquisition.