User manual

Acquisition
Digital Display Setup
Choose the type and position of the digital traces that appear on screen for each digital group.
1. Choose a Display Mode:
l Lines (default) shows a time-correlated trace indicating high, low, and transitioning points
(relative to the Threshold) for every digital line in the group. The size and placement of the
lines depend on the number of lines, the Vertical Position and Group Height settings.
l Bus collapses the lines in a group into their Hex values. It appears immediately below all the
Line traces when both are selected.
l Line & Bus displays both types of digital trace.
2. In Vertical Position, enter the number of divisions (positive or negative) relative to the zero line of
the grid where the display begins.The top of the first trace appears at this position.
3. In GroupHeight, enter the total number of grid divisions the entire display should occupy. All the
selected traces (Line and Bus) will appear in this much space. Individual traces are resized to fit the
total number of divisions available.
To close digital traces, uncheck the Group box on the Digitaln dialog.
35