MAUI Oscilloscopes Remote Control and Automation Manual
MT_PF_COUNTERS?, MTPC?
Description
This query returns the number of passed waveforms out of the total number tested and the percent fail
rate for the last ET-PMT mask test.
Query Syntax
MTPC?
Response Format
MT_PF_COUNTERS PASS,<value>,OF,<value>,FAIL_RATE,<percent value>
Related Commands
CUSTOM_APPLICATION, CUSTOM_OPTIONS?, MT_FAIL_ACTIONS, MT_OPC?, MT_SELECT_TEST, MT_
SYMBOL?