Data Sheet

TeledyneLeCroyCorporation VoyagerM310PAnalyzer
68 USBProtocolSuiteUserManual
Note:USB2.0LinkLEDsoperateonlywhileUSB2.0RecordingorReal‐Time Statistics(RTS)isrunning.
USB3.1LEDsalwaysoperate,unlessUSB3.1hasbeendisabledintheRecordingOptions
GeneralTab.
WARNING: DonotopentheVoyagerM310Penclosure.Nooperatorserviceablepartsare
inside.ReferservicingtoTeledyneLeCroycustomercare.
2.1.4 Specifications
TheAnalyzerhasthefollowingspecifications.
Power Requirements
24VDC,6.67Amps.
Environmental Conditions
Probing Characteristics
Operating
Temperature
0to50°C(32to122°F)
StorageRange ‐20to80°C(‐4to176°F)
OperatingHumidity 10to90%,non‐condensing
OperatingAltitude Upto6560feet(2000meters)
Connection
SuperSpeedPlusUSBType‐C
TM
connectors:The
VoyagerSuperSpeed(3.1)analyzerandexerciserport
differentialinputimpedanceisbetween80and120
ohms.
USB3.1USBType‐C
TM
connectorcables
HighSpeedUSBConnectors
Standardcables
Note:TheUSB2.0exerciserportisnotterminated
differentially.Thesingleendedterminationis
45 ohmstoground+/‐10%orbetween40.5
and49.5ohms.TheeffectivedifferentialUSB
2.0terminationisbetween80and100ohms
(90ohms+/‐10%).