Instructions

42 T3VNA User Manual
Table 2-25 Reference level under normalization
Parameter Explanation
Default 0 dB
Range -200 dB - 200 dB
Unit dB
Knob Step 1 dB
Direction Key Step 10 dB
2.2.7.8 Norm Ref Pos
Adjust the vertical position of the normalization reference level on the screen by adjusting the reference
position when normalization is enabled.
The function of this menu is similar to that of Norm Ref Level. When it is set to 0%, the
normalization reference level is displayed at the bottom of the screen grid and at the top when it is
set to 100%.
Table 2-26 TG reference position under normalization
Parameter Explanation
Default 100%
Range 0 - 100%
Unit 100%
Knob Step 1%
Direction Key Step 10%
2.2.8 Demod
Press Demod at the front panel to enter the demodulation setting menu. Both AM and FM
demodulations are available on the analyzer.
2.2.8.1 Demod (AM/FM)
Set the demodulation type to AM or FM or o. The default is o.
The analyzer features an earphone jack and the demodulated signal can be output in audio
frequency (AF) mode through the earphone. The frequency and intensity of AF denotes the
frequency and amplitude of the signal respectively.
A marker would be enabled automatically, place it at the desired frequency and set a demod time.
The analyzer would demod at the marker frequency for the demod time, and then performs a
normal sweep.
Zero span and linear scale type amplitude can be used to see the time domain waveform of the
modulated signal.
More demodulation analysis, please refer to AM/FM demodulation in Modulation Analyzer section.