2. Password
Set the password for login VNC.
3. View Only
Set whether the analyzer is in view-only mode on VNC.
When the view-only mode is turned on, the analyzer can only be viewed and cannot be operated on
VNC. When the view-only mode is turned o, the analyzer can be viewed and operated simultaneously
on VNC.
After modifying the view-only mode, VNC needs to be restarted.
6.1.4 Calibration
6.1.4.1 Auto Cal
Set whether to turn on the auto calibration or not and turn off by default. When Auto Cal is turned on,
the analyzer will process self-calibration according to the temperature dierence. Half an hour after
startup, the device will query temperature every minute. If the temperature dierence is greater
than 2°, the device will be self-calibrated.
6.1.5 System Info
6.1.5.1 System Info
Product Model, Serial and HOST ID
Software Version and Hardware Version
Option Information