Instructions

T3VNA Programming Guide 65
Instruction Sets TG normalize on-off.
Gets TG normalize state.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return 0|1
Default 0
Menu TG > Normalize > Normalize
Example :CALCulate:NTData ON
3.9.5 TG Normalize Reference Level
(:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel)
Command
Format
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel <value>
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel?
Instruction Sets TG normalize reference level.
Gets TG normalize reference level.
Parameter
Type
Float, unit: dB
Parameter
Range
-200 dB ~ 200 dB
Return Float, unit: dB
Default 0 dB
Menu TG > Normalize > Ref Lvl
Example :DISPlay:WINDow:TRACe:Y:NRLevel 10
3.9.6 TG Normalize Reference Position
(:DISPlay:WINDow:TRACe:Y[:SCALe]:NRPosition)
Command
Format
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRPosition <integer>
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRPosition?
Instruction Sets TG normalize reference position.
Gets TG normalize reference position.
Parameter
Type
Integer
Parameter
Range
0 ~ 100%
Return Float
Default 100%
Menu TG > Normalize > Position
Example :DISPlay:WINDow:TRACe:Y:NRPosition 10
3.9.7 TG Normalize Reference Trace on-off
(:DISPlay:WINDow:NTTRace[:STATe])
Command
Format
:DISPlay:WINDow:NTTRace[:STATe] OFF|ON|0|1
:DISPlay:WINDow:NTTRace[:STATe]?
Instruction Sets TG normalize reference trace on-o.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1