Instructions

60 T3VNA Programming Guide
Default None
Menu Meas > Occupied BW
Example : MEASure:OBW:CENTroid?
3.7.3.7 Query Transmit Frequency Error
(:MEASure:OBWidth:OBWidth:FERRor?)
Command
Format
:MEASure:OBWidth:OBWidth:FERRor?
Instruction Uses this command to query transmit frequency error.
Parameter
Type
None
Parameter
Range
None
Return Float, unit: Hz
Default None
Menu Meas > Occupied BW
Example :MEASure:OBWidth:OBWidth:FERRor?
3.7.4 SubsectionT-power(T-Power)
3.7.4.1 T-power Center Frequency
([:SENSe]:TPOWer:FREQuency:CENTer)
Command
Format
[:SENSe]:TPOWer:FREQuency:CENTer <freq>
[:SENSe]:TPOWer:FREQuency:CENTer?
Instruction Sets T-power center frequency.
Gets T-power center frequency.
Parameter
Type
Float, unit: Hz, KHz, MHz, GHz
Parameter
Range
50 Hz~1.499999950 GHz
Zero Span: 0~1.5 GHz
Return Float, unit: Hz
Default 750MHz
Menu Meas > T-power > Meas Setup > Center Freq
Example :TPOWer:FREQuency:CENTer 15KHz
3.7.4.2 T-power Start Line ([:SENSe]:TPOWer:LLIMit)
Command
Format
[:SENSe]:TPOWer:LLIMit <time>
[:SENSe]:TPOWer:LLIMit?
Instruction Sets T-power start line.
Gets T-power start line.
Parameter
Type
Float, unit: s
Parameter
Range
0 ~ 1000 s
Return Float, time unit: s
Default 0
Menu Meas > T-power > Meas Setup > Start Line
Example :TPOWer:LLIMit 0.01