Instructions

T3VNA Programming Guide 53
3.6.2.12 Limit X-axis Unit (:CALCulate:LLINe:CONTrol:DOMain)
Command
Format
:CALCulate:LLINe:CONTrol:DOMain FREQuency|TIME
:CALCulate:LLINe:CONTrol:DOMain?
Instruction Toggles the limit X-axis value between frequency and time.
Gets the limit X-axis unit.
Parameter
Type
Enumeration
Parameter
Range
FREQuency|TIME
Return Enumeration
Default FREQuency
Menu Limit > Setup > X Axis
Example :CALCulate:LLINe:CONTrol:DOMain FREQuency
3.6.2.13 Limit Beep State (:CALCulate:LLINe:CONTrol:BEEP)
Command
Format
:CALCulate:LLINe:CONTrol:BEEP OFF|ON|0|1
:CALCulate:LLINe:CONTrol:BEEP?
Instruction Use this command to turn on/off the limit beep status.
Gets limit beep state.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return 0|1
Default ON
Menu Limit > Setup > Buzzer
Example :CALCulate:LLINe:CONTrol:BEEP OFF
3.6.2.14 Query Limits Result (:CALCulate:LLINe:FAIL?)
Command
Format
:CALCulate:LLINe:FAIL?
Instruction This query returns the limits pass/failed result. If the test result fails, this
command will get result FAIL. If the test result passes, it will get result PASS.
Parameter
Type
None
Parameter
Range
None
Return PASS|FAIL
Default None
Menu None
Example :CALCulate:LLINe:FAIL?
3.6.2.15 Limit Fail to Stop (:CALCulate:LLINe:FAIL:STOP)
Command
Format
:CALCulate:LLINe:FAIL:STOP OFF|ON|0|1
:CALCulate:LLINe:FAIL:STOP?
Instruction Sets whether to stop the test if the test fails.
Gets whether to stop the test if the test fails.
Parameter Boolean