Instructions

T3VNA Programming Guide 39
3.5.9.3 Display Line (:DISPlay:WINDow:TRACe:Y:DLINe)
Command
Format
:DISPlay:WINDow:TRACe:Y:DLINe <value>
:DISPlay:WINDow:TRACe:Y:DLINe?
Instruction Sets the amplitude value for the display line.
Gets the amplitude value for the display line.
Parameter
Type
Float, unit: dBm
Parameter
Range
Ref Level ~ Ref Level - 100 dBm
Return Float, unit: dBm
Default 0 dBm
Menu Display > Display Line
Example :DISPlay:WINDow:TRACe:Y:DLINe -10
3.6 Calculate Subsystem
3.6.1 Marker Subsection
3.6.1.1 Marker On/O
(:CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe)
Command
Format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe?
Instruction This command toggles the selected marker status between on and off.
Gets marker state.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return 0|1
Default OFF
Menu Marker
Example :CALCulate:MARK1:STATe ON
3.6.1.2 Marker All O (:CALCulate:MARKer:AOFF)
Command
Format
:CALCulate:MARKer:AOFF
Instruction Turn all the markers o.
Parameter
Type
None
Parameter
Range
None
Return None
Default None
Menu None