Instructions
36 T3VNA Programming Guide
Parameter
Type
Integer
Parameter
Range
1 ~ 999
Return Integer
Default 1
Menu Trace > Avg Times
Example :AVERage:TRACe1:COUNt 10
3.5.7.3 Average Restart
([:SENSe]:AVERage:TRACe[1]|2|3|4:CLEar)
Command
Format
[:SENSe]:AVERage:TRACe[1]|2|3|4:CLEar
Instruction Restarts the trace average. This command is only available when average is
on.
Parameter
Type
None
Parameter
Range
None
Return None
Default None
Menu None
Example :AVERage:TRAC1:CLEar
3.5.8 Sweep Subsection
3.5.8.1 Sweep Mode ([:SENSe]:SWEep:MODE)
Command
Format
[:SENSe]:SWEep:MODE AUTO|FFT|SWEep
[:SENSe]:SWEep: MODE?
Instruction Sets sweep mode.
Gets sweep mode.
Parameter
Type
Enumeration
Parameter
Range
AUTO|FFT|SWEep
Return Enumeration
Default SWEep
Menu Sweep
Example :SWEep:MODE SWEep
3.5.8.2 Sweep Time ([:SENSe]:SWEep:TIME)
Command
Format
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Instruction Specifies the time in which the instrument sweeps the display. A span value of 0
Hz causes the analyzer to enter zero span mode. In zero span the X-axis
represents time rather than frequency.
Parameter
Type
Float, unit: ks, s, ms, us