Instructions
30 T3VNA Programming Guide
Format :DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision?
Instruction This command sets the per-division display scaling for the y-axis when scale
type of Y axis is set to Log.
Gets Scale/Div when scale type of Y axis is set to Log.
Parameter
Type
Float
Parameter
Range
1 dB ~ 10 dB
Return Float, unit: dB
Default 10 dB
Menu Amplitude > Scale/Div
Example :DISPlay:WINDow:TRACe:Y:PDIVision 10 dB
3.5.3.9 Correction O ([:SENSe]:CORRection:OFF)
Command
Format
[:SENSe]:CORRection:OFF
Instruction Turn off the amplitude correction function o and all of the correction sets are
off.
Parameter
Type
None
Parameter
Range
None
Return None
Default None
Menu None
Example :SENSe:CORRection:OFF
3.5.3.10 Correction Apply State
([:SENSe]:CORRection:CSET:ALL[:STATe])
Command
Format
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Instruction Turns on or off the amplitude corrections.
When turned on, only the correction sets that were turned on are enabled.
When turned off, all of the correction Sets are disabled. If there is no correction
enabled, state cannot be set to on.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return 0|1
Default OFF
Menu Amplitude > Corrections > Apply Corrections
Example :SENSe:CORRection:CSET:ALL:STATe OFF
3.5.3.11 Sets Correction X State O
([:SENSe]:CORRection:CSET[1]|2|3|4:OFF)
Command
Format
[:SENSe]:CORRection:CSET[1]|2|3|4:OFF
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]?
Instruction Turns the amplitude correction function on/off.