Data Sheet

6
SPECIFICATIONS
T3VNA1500
Sweep and Trigger
Sweep time 1 ms to 3200 s
Sweep mode RBW = 100 Hz – 1 MHz, Sweep
RBW = 1 Hz – 10 kHz, FFT
Sweep rule Single, Continuous
Trigger source Free, Video, External
External trigger 5 V TTL level, Rising edge/Falling edge
Amplitude Accuracy and Range Specications
Amplitude and Level
Measurement range DANLto+10dBm,100kHz–1MHz,preamplieroff
DANLto+20dBm,1MHz–1.5GHz,preamplieroff
Reference level -200 dBm to +30 dBm, 1 dB steps
Preamplier 20 dB (nom.)
Input attenuation 0 – 30 dB, 1 dB steps
Maximum input DC voltage +/- 50 V
DC
Maximum average power 30dBm,3minutes,fc≥10MHz,attenuation>20dBm,preampoff
Maximum damage level 33dBm,fc≥10MHz,attenuation>20dBm,preampoff
Displayed Average Noise Level (DANL)
20°C to 30°C, attenuation = 0 dB, sample detector, trace average > 50, Normalized to 1 Hz, TG off
Preamp off 100 kHz – 1 MHz -101 dBm, -107 dBm (typ.)
1 MHz – 10 MHz -124 dBm, -130 dBm (typ.)
10 MHz – 200 MHz -128 dBm, -134 dBm (typ.)
200 MHz – 1.5 GHz -121 dBm, -127 dBm (typ.)
Preamp on 100 kHz – 1 MHz -120 dBm, -128 dBm (typ.)
1 MHz – 10 MHz -147 dBm, -152 dBm (typ.)
10 MHz – 200 MHz -150 dBm, -156 dBm (typ.)
200 MHz – 1.5 GHz -142 dBm, -148 dBm (typ.)
Phase Noise
20°C to 30°C, fc = 1 GHz
Phase Noise < -95 dBc/Hz @ 10 kHz offset, < -98 dBc/Hz (typ.)
< -96 dBc/Hz @ 100 kHz offset, < -97 dBc/Hz (typ.)
< -115 dBc/Hz @ 1 MHz offset, < -117 dBc/Hz (typ.)
Level Display
Logarithmic level axis 1 dB to 200 dB
Linear level axis 0 to reference level
Units of level axis dBm,dBmV,dBμV,dBμA,Volt,Watt
Number of display points 751
Number of traces 4
Trace detectors Positive-peak, Negative-peak, Sample, Normal, Average(Voltage/RMS/Video), Quasi-peak
Trace functions Clear write, Max Hold, Min Hold, View, Blank, Average, Math
Frequency Response
20°Cto30°C,30%to70 %relativehumidity,att=20dB,relativetofc=50MHz
Preamp off ±0.8 dB, ±0.4 dB (typ.)
Preamp on ±0.9 dB, ±0.5 dB (typ.)