SPECIFICATIONS
Amplitude Characteristic
T3SA3200 T3SA3100
Amplitude and Level
Measurement range DANL to +10 dBm, 100 kHz ~ 1 MHz, preamplier off DANL to +20 dBm, 1 MHz ~ 3.2 GHz,
preamplier off
Reference level -100 dBm to +30 dBm, 1 dB steps Preamplier
20 dB (nom.), 9 kHz ~ 3.2 GHz Input attenuation
0 ~ 51 dB, 1 dB steps
Maximum input DC voltage ±50 VDC
Maximum average RF power 30 dBm, 3 minutes, fc ≥ 10 MHz, attenuation > 20 dBm, preamp off Maximum
damage level 33 dBm, fc ≥ 10 MHz, attenuation > 20 dBm, preamp off
Displayed Average Noise Level (DANL)
20°C ~ 30°C, attenuation = 0 dB, sample detector, trace average > 50
Preamp off RBW = 10 Hz Normalization to 1 Hz
9 kHz ~ 100 kHz -100 dBm (nom.) -110 dBm (nom.)
100 kHz ~ 1 MHz -97 dBm, -101 dBm (typ.) -107 dBm, -111 dBm (typ.)
1 MHz ~ 10 MHz -122 dBm, -126 dBm (typ.) -132 dBm, -136 dBm (typ.)
10 MHz ~ 200 MHz -127 dBm,-131 dBm (typ.) -137 dBm, -141 dBm (typ.)
200 MHz ~ 2.1 GHz -125 dBm, -129 dBm (typ.) -135 dBm, -139 dBm (typ.)
2.1 GHz ~ 3.2 GHz -116 dBm, -122 dBm (typ.) -126 dBm, -132 dBm (typ.)
Preamp on 9 kHz ~ 100 kHz -107 dBm (nom.) -117 dBm (nom.)
100 kHz ~ 1 MHz -122 dBm, -127 dBm (typ.) -132 dBm, -137 dBm (typ.)
1 MHz ~ 10 MHz -138 dBm, -144 dBm (typ.) -148 dBm, -154 dBm (typ.)
10 MHz ~ 200 MHz -146 dBm, -151 dBm (typ.) -156 dBm, -161 dBm (typ.)
200 MHz ~ 2.1 GHz -145 dBm, -148 dBm (typ.) -155 dBm, -158 dBm (typ.)
2.1 GHz ~ 3.2 GHz -135 dBm, -139 dBm (typ.) -145 dBm, -149 dBm (typ.)
Phase Noise
20°C ~ 30°C, f c= 1 GHz
Phase Noise <-95 dBc/Hz @10 kHz offset, <-98 dBc/Hz (typ.)
<-96 dBc/Hz @100 kHz offset, <-97 dBc/Hz (typ.)
<-115 dBc/Hz @1 MHz offset, <-117 dBc/Hz (typ.)
Level Display
Logarithmic level axis 10 dB to 200 dB
Linear level axis 0 to reference level
Units of level axis dBm, dBmV, dBμV, dBμA, V, W
Number of display points 751
Number of traces 4
Trace detectors Positive-peak, Negative-peak, Sample, Normal, Average (Voltage/RMS/Video),
Quasi-peak (with EMI option)
Trace functions Clear write, Max Hold, Min Hold, View, Blank, Average
Frequency Response
20°C to 30°C, 30 % to 70 % relative humidity, attenuation = 20 dB, reference frequency 50 MHz
Preamp off ±0.8 dB
±0.4 dB, (typ.)
Preamp on ±0.9 dB
±0.5 dB, (typ.)