Data Sheet
6
SPECIFICATIONS
Amplitude Characteristic
T3SA3200 T3SA3100
Error and Accuracy
Resolution bandwidth switching
uncertainty
10 kHz RBW
Logarithmic resolution ±0.2 dB, liner resolution ±0.01, nominal
Input attenuation switching
uncertainty
20 to 30 , fc = 50 MHz, preamp off, Relative to 20 dB, 1 to 51 dB attenuation
±0.5 dB
Absolute amplitude accuracy 20°C to 30°C, fc = 50 MHz, RBW = 1 kHz, VBW = 1 kHz, peak detector, attenuation = 20 dB,
95th percentile reliability
preamp off ±0.4 dB, input signal -20 dBm
preamp on ±0.5 dB, input signal -40 dBm
Total amplitude accuracy 20°C to 30°C, Fc >100 kHz, input signal -50 dBm ~ 0 dBm, RBW = 1 kHz, VBW = 1 kHz,
peak detector, attenuation = 20 dB, preamp off, 95th percentile reliability
± 0.7 dB
RF input VSWR input attenuation 10 dB, 1 MHz ~ 3.2 GHz
<1.5, nominal
Distortion and Spurious Responses
Second harmonic distortion fc ≥ 50 MHz, mixer level -30 dBm, attenuation = 0 dB, preamp off, 20°C to 30°C, typ.
-65 dBc
Third-order intercept fc ≥ 50 MHz, two -20 dBm tones at input mixer spaced by 100 kHz, attenuation = 0 dB,
preamp off, 20°C to 30°C, typ.
+10 dBm
1dB Gain Compression fc ≥ 50 MHz, attenuation = 0 dB, preamp off, 20°C to 30°C, nom.
>-5 dBm
Residual response input terminated = 50 Ω,attenuation = 0 dB, 20°C to 30°C, typ.
<-90 dBm
Input related spurious Mixer level = -30 dBm, 20°C to 30°C
<-65 dBc
Sweep and Trigger
Sweep time 1 ms to 3000 s
Sweep accuracy Accuracy, Speed
Sweep mode Sweep FFT
RBW = 30 Hz ~ 1 MHz RBW = 1 Hz ~ 10 kHz
Sweep rule Single, Continuous
Trigger source Free, Video, External
External trigger 5 V TTL level, rising edge/falling edge
Tracking Generator (Option T3SA3000-TG)
Frequency range 100 kHz ~ 3.2 GHz 100 kHz~2.1 GHz
RBW 30 Hz ~ 1 MHz, only sweep mode
Output level -20 dBm ~ 0 dBm
Output level resolution 1 dB
Output flatness ±3 dB
Output maximum reverse level Mean power: 30 dBm, DC: ±50 VDC
EMI Receiver Measurement (Option T3SA3000-EMI)
Resolution bandwidth (6 dB) 200 Hz, 9 kHz, 120 kHz
Detector Quasi-peak (following CISPR 16-1-1)
Dwell time 0 µs ~ 10 s
Reflection Measurement (Option T3SA3000-RFM-KIT)
Function VSWR, Return loss, Reflection coefcient
(The T3SA3000-RFM-KIT option REQUIRES the T3SA3000-TG Option)