Data Sheet
PRODUCT OVERVIEW
The T3EL15060P and T3EL50015P Series of are single channel, 300 Watt Electronic Loads ideal for R&D, product
validation and Q&A in a bench or automated environment for low to medium power applications starting from 60mA,
such as electronic components, batteries, portable chargers and power products.
The T3ELXXXXXP has seven operating modes. Among them, the four basic operating modes are constant current,
constant voltage, constant resistance, and constant power. Three other combined operating modes are constant
current + constant voltage, constant resistance + constant voltage, constant power + constant voltage. Users can
select operating modes based upon products' test requirements. For C.C. mode the electronic load will sink a
constant current according to the set current value; for C.V. mode the electronic load will attempt to sink sucient
current to control the source voltage to the programmed value; for C.R. Mode the electronic load will sink a current
linearly proportional to the input voltage according to the set resistance value; for C.P. Mode the electronic load will
initiate load power sinking operation (load voltage x load current) in accordance with the programmed power setting.
To meet the requirements of dierent test conditions, the Static function is to sink a constant current; the Dynamic
function is to periodically switch between two sink conditions, and the Sequence function is to provide tests for more
than two sink conditions. The sequence function can be divided into Normal Sequence and Fast Sequence. Normal
Sequence is the most exible means of generating complex sequences, that can allow users to establish a set of
changing current sink conditions based upon dierent sinking requirements (CC, CR, CV or CP mode) and time
(adjustable range: 1ms to 999h 59min 59s in Normal Mode). Fast sequence allows time resolution of 25us to be set
for the smallest step. Setting parameters for multiple steps can simulate consecutive current changes of various real
load conditions. For instance, while using an electronic load to test a power-driven tool's power supply.
The Soft Start function allows users to determine the rise time of the current sink by measuring the required time to
reach the electronic load's set current, resistance or power value. Setting a proper rise time for Soft Start will counter
output voltage uctuations caused by a DUT's (power supply) transient output current. Generally DC loads do not
have the soft start function that is built in to the T3ELXXXXXP Series.
When conducting high speed current sink operation, the inductance eect on the cables connecting the electronic
load and DUT will lead to transient voltage drop on the electronic load's input terminals. That will result in Voltage
Non-monotonic increase. The T3ELXXXXXP Series soft start function not only allows output the voltage to follow a
Monotonic increase, but also prevents inrush current and surge voltages that could damage the DUT. For instance,
tests using a power supply, LED and a DC load (activate the soft start function) can prevent inrush current and surge
voltages from causing damage to the LEDs.
The built-in BATT Test Automation provides battery discharge applications with more exible discharge stop settings
as well as rise and fall Slew Rates for discharge current settings.
The OCP, OPP test Automation for DUTs (ex. Power Supply), provide users with high resolution measurement values
to verify DUT's activation point. Provide users with measurement results so as to help them determine whether DUT's
actual over protection activation point meets regulations.
The T3ELXXXXXP provides users with an analog control terminal to control the T3ELXXXXXP using an external
voltage, external resistance and switch.
Ordering Information
T3EL15060P 150 V / 60 A / 300 W Programmable Single Channel D.C. Electronic Load
T3EL50015P 500 V / 15 A / 300 W Programmable Single Channel D.C. Electronic Load
Supplied Accessories include Quick Start Guide, Power Cord and Remote Sense Cables (1 x Red, 1 x Black).