Datasheet
10
SPECIFICATIONS
Acquire System
Sampling Rate 2 GSa/s (single-channel
1)
), 1 GSa/s (dual-channel)
Memory Depth 200 Mpts (single-channel), 100 Mpts (dual-channel)
Peak Detect 1 ns
Average Averages: 4, 16, 32, 64, 128, 256, 512, 1024
Eres Enhance bits: 0.5, 1, 1.5, 2., 2.5, 3 selectable
Interpolation Sinx/x, Linear
Input
Channels 2/4 + EXT
Coupling DC, AC, GND
Impedance DC:(1MΩ±2 %)||(17pF±2pF)
50Ω:50Ω±1 %
Max.Input voltage 1MΩ≤400Vpk(DC+PeakAC),DC–10kHz
50Ω≤5Vrms,±10VPeak
CH to CH Isolation DC–100MHz>40dB,100MHz–BW≥34dB
Probe Attenuation 1X, 10X, 100X, Custom
Horizontal System
Time Scale 1.0 ns/div – 1000 s/div T3DSO2502A 0.5 ns/div – 1000 s/div
Channel Skew < 100 ps
Waveform Capture Rate Up to 120,000 wfm/s (normal mode), 500,000 wfm/s (sequence mode)
Intensity grading 256-level
Display Format Y-T,X-Y,Roll(≥50ms/div)
Time base Accuracy ±1ppminitial;±1ppm1styearageing;±3.5ppm10-yearageing
Roll Mode 50 ms/div – 1000 s/div (1-2-5 Step)
Vertical System
Bandwidth (-3dB) 500 MHz T3DSO2502A (2 Channels), 350 MHz (4 Channels)
350 MHz (T3DSO2354A)
200 MHz (T3DSO2204A)
100 MHz (T3DSO2104A)
VerticalResolution 8 bit
10bitmode≤100MHz
VerticalRange 8 divisions
VerticalScale(Probe1X) 1MΩ500µV/div–10V/div
50Ω:500µV/div–1V/div
Offset Range (Probe 1X) 500µV/div–100mV/div:±2V
102mV/div–1V/div:±20V
1.02V/div–10V/div:±200V
Bandwidth Limit 20MHz-0 %–+20 %,200MHz-0 %–+20 %
Bandwidth Flatness
50Ω
DC–10 %(BW):±0.5dB
10 %–33 %(BW):±0.8dB
33 %–66 %(BW):+1dB,-1.2dB
66 %–BW:+2dB,-2.5dB
Low Frequency Response
(AC Coupling –3 dB)
≤5Hz(typical,atinputBNC)
Noise 80µVat500MHzbandwidth
DC Gain Accuracy ≤3.0 %
Offset Accuracy ±(1 %*offset+1.5 %*8*div+1mV)
Rise Time
1)
50Ω
(Typ.)≤800ps(T3DSO2502A2Channel500MHzmode)≤1ns(T3DSO2502A4Channelmode)
(Typ.)≤1ns(T3DSO2354A)
(Typ.)≤1.7ns(T3DSO2204A)
(Typ.)≤3.5ns(T3DSO2104A)
(Typ.)≤3.3ns(T3DSO2104A,T3DSO2204A,T3DSO2354A,T3DSO2502Ain10bitmode)
Overshoot (150 ps Fast Edge,
50Ωinput)
<12 %
All specications are not guaranteed unless the following conditions are met:
●
The oscilloscope calibration period is valid
●
Theoscilloscopehasbeenworkingcontinuouslyforatleast30minutesatthespeciedtemperature(18 – 28)