Data Sheet

9
Function Probe
Type
Type Optimum
Range
1 Year
TCAL °C ± 5°C
Temperature
coefcient
0°C to (TCAL °C - 5°C)
(TCAL °C + 5°C) to 50°C
Temperature
RTD
2)
(R0is49Ωto2.1kΩ)
α=0.00385 -200°C ~ 660°C 0.16°C 0.01°C
Thermocouple
3)
B 0°C ~ 1820°C 0.76°C 0.14°C
E -270°C ~ 1000°C 0.5°C 0.02°C
J -210°C ~ 1200°C 0.5°C 0.02°C
K -270°C ~ 1370°C 0.5°C 0.03°C
N -270°C ~ 1300°C 0.5°C 0.04°C
R -270°C ~ 1760°C 0.5°C 0.09°C
S -270°C ~ 1760°C 0.6°C 0.11°C
T -270°C ~ 400°C 0.5°C 0.03°C
Remarks:
1)
Specicationsarefor90minuteswarm-up.Exclusiveofsensorerror.
2)
Specicationisfor4WRsensormeasurementor2WRmeasurementusingRELoperation.
3)
Relativetocoldjunctiontemperature,accuracyisbasedonITS-90.Built-incoldjunctiontemperaturereferstothetemperatureinsidethebananajack
and its accuracy is ± 2.5 .
Function Setting Integration Readings/s 50 Hz (60 Hz)
DC Voltage
DC Current
2-wire Resistance
4-wire Resistance
0.005 (0.006) NPLC 100 (100) µs 10000 (10000)
0.05 (0.06) NPLC 1 (1) ms 1000 (1000)
0.5 (0.5) NPLC 4 (4) ms 100 (100)
1 NPLC 20 (16.7) ms 50 (60)
10 NPLC 200 (167) ms 5 (6)
100 NPLC 2 (1.67) s 0.5 (0.6)
AC Voltage
AC Current
3 Hz AC Filter 0.5
20 Hz 2
200 Hz 50
Frequency and Period
1)
1sGatetime 1
0.1 s 10
0.01 s 100
0.001 s 500
Capacitance
2)
100 mF Range 0.5
Remarks:
1)
20Vrange,1kHzinput.
2)
The measurement period changes with the capacitance under test.
3)
Auto zero off, auto range off.
Temperature Characteristics Accuracy± (% of Reading)
1)
Measurement Rate Measurement rate
3)