Data Sheet

4
SPECIFICATIONS
DC Characteristics Accuracy ± (% of reading + % of range)
1)
Function Range
2)
Test Current
Or Shunt
Voltage
24 Hour
3)
TCAL °C ± 1°C
90 day
TCAL °C ± 5°C
1 Year
TCAL °C ± 5°C
Temperature
coefcient 0°C to
(TCAL °C - 5°C)
(TCAL °C + 5°C) to
50°C
DC Voltage
200.0000 mV 0.0020 + 0.0015 0.0030 + 0.0020 0.0040 + 0.0023 0.0005 + 0.0003
2.000000 V 0.0015 + 0.0004 0.0020 + 0.0004 0.0035 + 0.0006 0.0005 + 0.0001
20.00000 V 0.0020 + 0.0003 0.0030 + 0.0004 0.0040 + 0.0004 0.0005 + 0.0001
200.0000 V 0.0020 + 0.0005 0.0040 + 0.0004 0.0050 + 0.0005 0.0005 + 0.0001
1000.000 V
4)
0.0020 + 0.0005 0.0040 + 0.0008 0.0055 + 0.0008 0.0005 + 0.0001
DC Current
200.0000μA < 0.03V 0.009 + 0.010 0.040 + 0.005 0.050 + 0.005 0.0020 + 0.0026
2.000000 mA < 0.25V 0.007 + 0.001 0.030 + 0.001 0.050 + 0.002 0.0020 + 0.0001
20.00000 mA < 0.07 V 0.006 + 0.008 0.030 + 0.005 0.050 + 0.005 0.0020 + 0.0015
200.0000 mA < 0.7V 0.009 + 0.001 0.030 + 0.001 0.050 + 0.002 0.0020 + 0.0001
2.000000 A < 0.12 V 0.045 + 0.015 0.080 + 0.005 0.100 + 0.012 0.0050 + 0.0008
10.00000 A
5)
< 0.6 V 0.090 + 0.002 0.120 + 0.005 0.150 + 0.005 0.0050 + 0.0018
Resistance
6)
200.0000Ω 1 mA 0.0030 + 0.0031 0.008 + 0.005 0.010 + 0.004 0.0006 + 0.0006
2.000000KΩ 1 mA 0.0020 + 0.0005 0.008 + 0.001 0.010 + 0.001 0.0006 + 0.0001
20.00000KΩ 100μA 0.0020 + 0.0005 0.008 + 0.001 0.010 + 0.001 0.0006 + 0.0001
200.0000KΩ 10μA 0.0020 + 0.0005 0.008 + 0.001 0.010 + 0.001 0.0006 + 0.0001
1.000000MΩ 2μA 0.0020 + 0.0010 0.010 + 0.001 0.012 + 0.001 0.0010 + 0.0002
10.00000MΩ 200 nA 0.015 + 0.001 0.030 + 0.001 0.040 + 0.001 0.0030 + 0.0005
100.0000MΩ 200nA||10MΩ 0.300 + 0.010 0.800 + 0.010 0.800 + 0.010 0.1500 + 0.0002
Diode Test
7)
0 ~ 2V 1 mA 0.002 + 0.009 0.008 + 0.020 0.010 + 0.020 0.0010 + 0.0020
2 ~ 4V 1 mA 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.0010 + 0.0020
Continuity Test 2000.0Ω 1 mA 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.0010 + 0.0020
Remarks:
1)
Specicationsarefor90-minutewarm-upand100NPLCintegrationtime.Forintegrationtime<100NPLC,addtheappropriate“RMSNoiseAdder”listedinthefollowingtable.
2)
10 %overrangeonallrangesexceptDCV1000VandDCI10Arange.
3)
Relative to calibration standards.
4)
For each additional volt over ± 500 V, add 0.03 mV error.
5)
For continuous current > 7A DC or 7A AC RMS, use 30 seconds ON followed by 30 seconds OFF.
6)
Specicationsarefor4-wireresistancemeasurementor2-wireresistancemeasurementusingRELoperation.WithoutRELoperation,add0.2Ωadditionalerrorin2-wire
resistance measurement.
7)
Accuracyspecicationsforthevoltagemeasuredattheinputterminalonly.1mAtestcurrentistypical.Variationinthecurrentsourcewillcreatesomevariationinthe
voltage drop across a diode junction. Adjustable voltage range: 0 ~ 4V.