Data Sheet
8
Temperature Measuring
Measurement Method Support for TC and RTD types of sensor
Trigger and Memory
Samples/Trigger 1 ~ 10000
Trigger Delay 6 ms ~ 10000 ms optional
External Trigger Input
Input Level TTL compatible (High level when left input terminal is
hanging in the air)
Trigger Condition Rising and Falling selectable
InputImpendance ≥20KΩ//400pF,DC-coupled
Min Pulse 500 us
VMC
Level TTL compatible
Output Polarity Positive and negative optional
OutputImpendance 200Ω,typical
History Records
Volatile Memory 10 K reading history records
Nonvolatile Memory 1GbNandFlash,Massstorage,congurationles,anddatales,SupportU-diskexternal
storage
Math Functions
Min/Max/Average, dBm, dB, Pass/Fail, Relative, Standard deviation, Hold, histogram, Trend chart, Bar chart
General Specications
Power Supply
AC 100 V ~ 120 V 50/60 Hz
AC 200 V ~ 240 V 50/60 Hz
Consumption 20 VA max
Mechanism
Dimension 282 mm × 260 mm × 105 mm
Weight 3.33 Kg
Other Characteristics
Display Screen 4.3” TFT-LCD, resolution 480*272
Operating Environment
Fullaccuracyfrom0to50,80 %RHand40°C,noncondensing
StorageTemperature:-20°C–+70°C
ShockandVibration:conformingtoMIL-T-28800E,5level(onlyforsine)
Height above sea level: up to 3000 meters
Safety
Conforming to IEC61010-1:2001. Measure CAT I 1000 V/CAT II 600V
Class of pollution: 2
Remote Interface 10/100 Mbit LAN, USB2.0 Full Speed Device and Host
Programmer Language Standard SCPI
Warm Up Time 30 minutes
MEASURING METHOD AND OTHER CHARACTERISTICS