Data Sheet
5
SPECIFICATIONS
Function Range
2)
Frequency Range 1 Year
23°C ± 5°C
Temperature coefcient
0°C ~ 18°C
28°C ~ 50°C
True-RMS
AC Voltage
3)
200 mV
20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 KHz 0.2 + 0.05 0.01 + 0.005
20 KHz – 50 KHz 1.0 + 0.05 0.01 + 0.005
50 KHz –100 KHz 3.0 + 0.05 0.05 + 0.010
2 V
20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 KHz 0.2 + 0.05 0.01 + 0.005
20 KHz – 50 KHz 1.0 + 0.05 0.01 + 0.005
50 KHz –100 KHz 3.0 + 0.05 0.05 + 0.010
20 V
20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 KHz 0.2 + 0.05 0.01 + 0.005
20 KHz – 50 KHz 1.0 + 0.05 0.01 + 0.005
50 KHz –100 KHz 3.0 + 0.05 0.05 + 0.010
200 V
20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 KHz 0.2 + 0.05 0.01 + 0.005
20 KHz – 50 KHz 1.0 + 0.05 0.01 + 0.005
50 KHz –100 KHz 3.0 + 0.05 0.05 + 0.010
750 V
20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 KHz 0.2 + 0.05 0.01 + 0.005
20 KHz – 50 KHz 1.0 + 0.05 0.01 + 0.005
50 KHz –100 KHz 3.0 + 0.05 0.05 + 0.010
True-RMS
AC Current
4)
20 mA
20 Hz – 45 Hz 1.5 + 0.10 0.015 + 0.015
45 Hz – 2 KHz 0.50 + 0.10 0.015 + 0.006
2 KHz – 10 KHz 2.50 + 0.20 0.015 + 0.006
200 mA
20 Hz – 45 Hz 1.5 + 0.10 0.015 + 0.005
45 Hz – 2 KHz 0.50 + 0.10 0.015 + 0.005
2 KHz – 10 KHz 2.50 + 0.20 0.015 + 0.005
2 A
20 Hz – 45 Hz 1.5 + 0.20 0.015 + 0.005
45 Hz – 2 KHz 0.50 + 0.20 0.015 + 0.005
2 KHz – 10 KHz 2.50 + 0.20 0.015 + 0.005
10 A
5)
20 Hz – 45 Hz 1.5 + 0.15 0.015 + 0.005
45 Hz – 2 KHz 0.50 + 0.15 0.015 + 0.005
2 KHz – 10 KHz 2.50 + 0.20 0.015 + 0.005
AC Characteristic Accuracy ± (% of Reading + % of Range)
1)
Additional wave crest factor error (not Sine)
6)
Wave crest coefcient Error (% Range)
1 – 2 0.05
2 – 3 0.2
Remarks:
1)
Specicationsarefor0.5Hourwarm-up,“Slow”measurementrateandcalibrationtemperature18°C~28°C.
2)
20 %overrangeonallrangesexceptforDCV1000V,ACV750V,DCI10AandACI10A.
3)
Specicationsareforamplitudeofsinewaveinput>5 %ofrange.Forinputsfrom1%to5 %ofrangeand<50kHz,addextra0.1%ofrangeerror.
For50kHzto100kHz,addextra0.1%ofrangeerror.
4)
Specicationsareforsinewaveinput>5%ofrange.0.1%errorswillsbeaddedwhentherangeofinputsinewaveis1 %to5 %.
5)
30 seconds OFF / 30 seconds ON is recommend for the continuous current that higher than DC 7 A or AC RMS 7 A.
6)
For frequency < 100 Hz