Datasheet

D11000PS DIFFERENTIAL PROBE SYSTEM
www.lecroy.com 239
PROBES AND
PROBE ACCESSORIES
Compatibility
The D11000PS is designed specifically for use with the SDA 11000, SDA 9000, or
SDA 18000. However, it does contain additional calibration data for use with all
of the lower bandwidth WaveMaster, SDA, and DDA 5005A Series oscilloscopes
and analyzers.
SPECIFICATIONS
Bandwidth, System, -3 dB 11 GHz (Typical)*
Rise Time, System < 50 ps (Typical)*
Rise Time, Probe only < 40 ps
Attenuation, Nominal ÷3
LF Attenuation Accuracy 2% (20–30 ˚C)
Output Zero < 15 mV referred to input
Noise, System 5 mV rms (Typical)
1
Differential Mode Range ±1 V
Common Mode Range ±4 V, Solder-In tip
±2 V, SMA cable input
Input Resistance at DC, 40 kΩ differential
(Solder-In lead) 20 kΩ each side to ground
Minimum Input Impedance, > 175 Ω
(Solder-In lead, to 11 GHz) (Refer to graph)
Input Impedance, (SMA cable input) 50 Ω
Input VSWR, < 1.5:1 DC–6 GHz
(Typical, each lead to ground) < 2.0:1 6 GHz–11 GHz
CMRR, > 40 dB DC–1 GHz
(Typical) > 30 dB 1 GHz–4 GHz
> 20 dB 4 GHz–11 GHz
* Measured as a system with SDA 11000, SDA 9000 and SDA 18000.
Can be extended by using DC Blocking Adapters.
PROBES AND PROBE ACCESSORIES