Datasheet
AP015 CP150 CP500
30 A 150 A 500 A
50 MHz 10 MHz 2 MHz
50 A ≤ 10 s 500 A ≤ 30 µs 700 A
≤ 7 ns < 35 ns < 175 ns
10 mA/div 200 mA/div 200 mA/div
– 500 A 1150 A
1% 1% 1%
≤ 25 mA 25 mA
AC, DC, GND AC, DC, GND AC, DC, GND
2 m 2 m 6 m
300 g 500 g 630 g
5 mm 20 mm 20 mm
ProBus, 1 MΩ only
‡
ProBus, 1 MΩ only
‡
ProBus, 1 MΩ only
‡
Indoor Indoor Indoor
0 ˚C to 40 ˚C 0 ˚C to 40 ˚C 0 ˚C to 40 ˚C
80% 80% 80%
2000 m 2000 m 2000 m
300 V CAT I 600 V CAT II, 300 V CAT III 600 V CAT II, 300 V CAT III
CURRENT PROBES
www.lecroy.com 271
PROBES AND
PROBE ACCESSORIES
PROBES AND PROBE ACCESSORIES
0.001
0.01
0.1
1
10
Frequency (Hz)
Insertion
Impedance
()
AP015 – 30 A, 50 MHz
0
50
100
150
200
250
300
350
400
450
500
550
Frequency (Hz)
Current
(A rms)
10 100 1 k 10 k 100 k 1 M 10 M
20
0
40
60
80
100
120
140
160
Frequency (Hz)
Current
(A rms)
CP150 – 150 A, 10 MHz CP500 – 500 A, 2 MHz
0.001
0.010
0.100
Frequency (Hz)
Insertion
Impedance
0.001
0.01
0.1
1
Frequency (Hz)
Insertion
Impedance
CP150 – 150 A, 10 MHz CP500 – 500 A, 2 MHz
* Guaranteed at 23 °C ±3 °C
†
The CP031 and CP030 are compatible with LeCroy X-Stream oscilloscopes running firmware version 4.3.1.1 or greater.
‡
Requires AP-1M for use with 50 Ω input only oscilloscopes.
Insertion Impedance vs. Frequency
(typical)
Maximum Input Current vs. Frequency
Insertion Impedance vs. Frequency
(typical)
Maximum Input Current vs. Frequency
Insertion Impedance vs. Frequency
(typical)