The Mixed Signal Setup Dialog
LocatedunderSetup→MixedSignalonthemenubar,thisfunctionshowstheMixedSignalSetupDialogwhere
youcanspecify(innanoseconds)ananalogtriggeralignmentcorrectionvalue.Thissettingcanbeusedtofine
tuneanalogdataacquiredfromtheoscilloscopetothedatafromthedigitalpod(for
exampletocompensateifa
differentlengthofcrosstriggercableisused).Positivevaluesshifttheanalogtriggerpointlaterintimeand
maketheanalogdatamoveearlierintimerelativetothedigitaldataontheWaveformDisplayArea.Negative
valuebehavesinversely.
Help
HelpinformationisaccessedunderHelponthemenubar.
AccesstheUserManualtoconvenientlyretrievethissamedocumentationorclickAboutLogicStudioforthe
softwareversionandLeCroycontactinformation.
Tell LeCroy
UnderHelponthemenubar,theTellLeCroyoptionisalwaysavailableforprovidingyourfeedback.Itcreates
andsendsanemailtoLogicStu diosupportatLeCroy.Youcanevenautomaticallyattachyourcurrentsetupand