Data Sheet

19
SPECIFICATIONS
Vertical - Analog Channels
HDO4024A
HDO4024A-MS
HDO4034A
HDO4034A-MS
HDO4054A
HDO4054A-MS
HDO4104A
HDO4104A-MS
Bandwidth @ 50 Ω (-3 dB) 200 MHz 350 MHz 500 MHz 1 GHz
Rise Time (10–90%, 50 Ω) 1.75 ns 1 ns 700 ps 450 ps
Input Channels 4
Vertical Resolution 12-bits; up to 15-bits with enhanced resolution (ERES)
Effective Number of Bits (ENOB) 8.8 bits 8.7 bits 8.6 bits 8.4 bits
Vertical Noise Floor
1 mV/div 70 μVrms 85 μVrms 100 μVrms 145 μVrms
2 mV/div 70 μVrms 85 μVrms 100 μVrms 145 μVrms
5 mV/div 75 μVrms 90 μVrms 105 μVrms 150 μVrms
10 mV/div 80 μVrms 95 μVrms 110 μVrms 155 μVrms
20 mV/div 100 μVrms 110 μVrms 130 μVrms 185 μVrms
50 mV/div 195 μVrms 210 μVrms 265 μVrms 275 μVrms
100 mVdiv 340 μVrms 360 μVrms 450 μVrms 500 μVrms
200 mV/div 1.00 mVrms 1.10 mVrms 1.25 mVrms 1.75 mVrms
500 mV/div 1.90 mVrms 2.10 mVrms 2.60 mVrms 2.75 mVrms
1 V/div 3.40 mVrms 3.70 mVrms 4.50 mVrms 4.90 mVrms
Sensitivity 50 Ω: 1 mV/div–1 V/div, fully variable; 1 MΩ: 1 mV/div–10 V/div, fully variable
DC Vertical Gain Accuracy
(Gain Component of DC Accuracy)
±(0.5%) F.S, offset at 0 V
Channel-Channel Isolation DC-200 MHz:
60 dB (>1000:1),
(For any two input
channels, same V/div
settings, typical)
DC-200 MHz:
60 dB (>1000:1),
200 MHz up to rated BW:
50 dB (>300:1),
(For any two input
channels, same V/div
settings, typical)
DC-200 MHz:
60 dB (>1000:1),
200 MHz up to
rated BW: 50 dB (>300:1),
(For any two input
channels, same V/div
settings, typical)
DC-200 MHz:
60 dB (>1000:1),
200-500 MHz: 50 dB
(>300:1), 500 MHz up to
rated bandwidth:
40 dB (>100:1)
(For any two input
channels, same V/div
settings, typical)
Offset Range 50 Ω: 1 mV - 4.95 mV: ±1.6 V, 5 mV - 9.9 mV: ±4 V, 10 mV - 19.8 mV: ±8 V, 20 mV - 1 V: ±10 V
1 MΩ: 1 mV - 4.95 mV: ±1.6 V, 5 mV - 9.9 mV: ±4 V, 10 mV - 19.8 mV: ±8 V, 20 mV - 100 mV: ±16 V,
  102 mV - 198 mV: ±80V, 200 mV - 1 V: ±160 V, 1.02 V -10 V: ±400 V
DC Vertical Offset Accuracy ±(1.0% of offset setting + 0.5%FS + 0.02% of max offset + 1mV)
Maximum Input Voltage 50 Ω: 5 Vrms, 1 MΩ: 400 V max (DC + Peak AC ≤ 10 KHz)
Input Coupling 50 Ω: DC, GND; 1 MΩ: AC, DC, GND;
Input Impedance 50 Ω ± 2.0%;1 MΩ ± 2.0% || 16 pF,
Bandwidth Limiters 20 MHz, 200 MHz
Horizontal - Analog Channels
Acquisition Modes Real-time, Roll, Random Interleaved Sampling (RIS), Sequence
Time/DIvision Range 200 ps/div - 1.25 ks/div with standard memory (up to 2.5 ks/div with -L memory);
RIS available at ≤ 10 ns/div; Roll Mode available at ≥ 100 ms/div and ≤ 5 MS/s
Clock Accuracy ±2.5 ppm + 1.0ppm/year from calibration
Sample Clock Jitter Up to 10 ms acquired time range: 280 fsrms (internal timebase reference)
Delta Time Measurement Accuracy
+
(Sample Clock Jitter)
2
(RMS, seconds, TIE)
Noise
SlewRate
+
(Sample Clock Jitter)
2
(RMS) + (clock accuracy
*
reading) (seconds)
2
2
*
Noise
SlewRate
2
Jitter Measurement Floor
+
(Sample Clock Jitter)
2
(RMS, seconds, TIE)
Noise
SlewRate
+
(Sample Clock Jitter)
2
(RMS) + (clock accuracy
*
reading) (seconds)
2
2
*
Noise
SlewRate
2
Jitter Between Channels Analog Channels: 2 psrms (TIE, typical); Digital Channels: 350 ps (maximum) between any two channels
Analog-Digital Channels: <5ns (maximum) between any analog and any digital channel
Channel-Channel Deskew Range ±9 x time/div. setting, 100 ms max., each channel
External Timebase Reference (Input) 10 MHz ±25 ppm at 0 to 10 dBm into 50 Ω
External Timebase Reference (Output) 10 MHz, 2.0 dBm ±1.5 dBm, sinewave synchronized to reference being used (internal or external reference)
Acquisition - Analog Channels
Sample Rate (Single-shot) 10 GS/s on all 4 Channels with Enhanced Sample Rate
Sample Rate (Repetitive) 125 GS/s, user selectable for repetitive signals (20 ps/div to 10 ns/div)
Memory Length
(# of Segments in Sequence Mode)
Standard: 12.5 Mpts/ch for all channels, 25 Mpts (interleaved) (10,000 segments)
Option - L: 25 Mpts/ch for all channels, 50 Mpts (interleaved) (10,000 segments)
Intersegment Time 1 μS
Averaging Summed averaging to 1 million sweeps; continuous averaging to 1 million sweeps
Enhanced Resolution (ERES) From 12.5- to 15-bits vertical resolution
Envelope (Extrema) Envelope, floor, or roof for up to 1 million sweeps
Interpolation Linear or Sin x/x (2 pt and 4 pt); 5 or 10 GS/s Enhanced Sample Rate defaults to 2 pt or 4 pt Sin x/x respectively